Comparison of tissue Doppler velocities obtained by different types of echocardiography systems: are they compatible?
MedLine Citation:
PMID:  20486956     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Both systolic and diastolic tissue Doppler (TD) velocities have an important diagnostic and prognostic role in cardiology. We aimed to compare TD velocities between two different echocardiography systems. Patients: Thirty-one consecutive patients (mean age: 65.2 +/- 17.5 years; 12 males) were enrolled. METHODS: Systolic (Sa), early (Ea), and late (Aa) diastolic velocities were measured by TD at the lateral mitral annulus by a Sonos 2000 (Hewlett-Packard, Andover, MA, USA) and a Philips iE33 system. The E/Ea ratio was calculated. RESULTS: Ea, Aa, and Sa velocities were higher when measured by the Sonos system (Ea: 13.2 +/- 4.1 cm/s vs. 8.3 +/- 3.6 cm/s; Aa: 14.8 +/- 3.8 cm/s vs. 9.3 +/- 2.3 cm/s; Sa: 15.2 +/- 3.6 cm/s vs. 8.4 +/- 2.0 cm/s; P < 0.0001 all). A significant correlation was found in Ea and in Ea/Aa (r = 0.84 and r = 0.85 resp; P < 0.0001 for both), and a weaker in Aa (r = 0.43; P = 0.02) between the machines. The Bland-Altman analysis showed broad limits of agreement between the measurements for Ea, Aa, and Sa (mean difference: 4.95 cm/s; 5.52 cm/s; 6.73 cm/s, respectively; limits: 0.64-9.25 cm/s; -1.39-12.39 cm/s; -0.37-13.83 cm/s, respectively). An E/Ea ratio >5.6 by the Sonos system showed 75% sensitivity and 79% specificity for elevated left ventricular filling pressure, defined as E/Ea >10 by the reference Philips system. CONCLUSIONS: Although diastolic TD velocities had excellent correlations between the two machines, there was a systematic overestimation by the Sonos system. Since the limits of agreement do not allow replacing the measurements, we suggest using the same echocardiographic equipment at patient follow-up.
Authors:
Mónika Dénes; Katalin Farkas; Tamás Erdei; Mária Lengyel
Related Documents :
8854436 - Reappraisal of cardiac murmurs related to aortic regurgitation.
507456 - Echocardiographic and hemodynamic correlates in prinzmetal angina pectoris: a case report.
20828746 - Continuous-flow cardiac assistance: effects on aortic valve function in a mock loop.
10424506 - Noninvasive estimation of pulmonary artery diastolic pressure in patients with tricuspi...
16794216 - Lack of evidence for an association between hemodynamic variables and hematoma growth i...
10596886 - How easily do topical antibiotics pass through tympanostomy tubes?--an in vitro study.
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Echocardiography (Mount Kisco, N.Y.)     Volume:  27     ISSN:  1540-8175     ISO Abbreviation:  Echocardiography     Publication Date:  2010 Mar 
Date Detail:
Created Date:  2010-05-21     Completed Date:  2010-09-20     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8511187     Medline TA:  Echocardiography     Country:  United States    
Other Details:
Languages:  eng     Pagination:  230-5     Citation Subset:  IM    
Affiliation:
Gottsegen György Hungarian Institute of Cardiology, H-1096 Budapest, Hungary. denesmoni@gmail.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Aged, 80 and over
Blood Flow Velocity
Echocardiography, Doppler / instrumentation*,  methods*
Female
Humans
Male
Middle Aged
ROC Curve

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Noninvasive evaluation of membranous subaortic stenosis: complimentary roles of echocardiography and...
Next Document:  Subclinical left ventricular dysfunction in asymptomatic severe aortic regurgitation patients with n...