Comparison of three portable instruments to measure compression pressure.
MedLine Citation:
PMID:  20924346     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
AIM: Measurement of interface pressure between the skin and a compression device has gained practical importance not only for characterizing the efficacy of different compression products in physiological and clinical studies but also for the training of medical staff.
METHODS: A newly developed portable pneumatic pressure transducer (Picopress®) was compared with two established systems (Kikuhime® and SIGaT tester®) measuring linearity, variability and accuracy on a cylindrical model using a stepwise inflated sphygmomanometer as the reference. In addition the variation coefficients were measured by applying the transducers repeatedly under a blood pressure cuff on the distal lower leg of a healthy human subject with stepwise inflation.
RESULTS: In the pressure range between 10 and 80 mmHg all three devices showed a linear association compared with the sphygmomanometer values (Pearson r>0.99). The best reproducibility (variation coefficients between 1.05-7.4%) and the highest degree of accuracy demonstrated by Bland-Altman plots was achieved with the Picopress® transducer. Repeated measurements of pressure in a human leg revealed average variation coefficients for the three devices of 4.17% (Kikuhime®), 8.52% (SIGaT®) and 2.79% (Picopress®).
CONCLUSION: The results suggest that the Picopress® transducer, which also allows dynamic pressure tracing in connection with a software program and which may be left under a bandage for several days, is a reliable instrument for measuring the pressure under a compression device.
Authors:
H Partsch; G Mosti
Related Documents :
8180746 - Pressure and volume control for local drug-delivery catheters: development of a new mic...
2135186 - Effect of subcutaneous implantation of anti-siphon devices on csf shunt function.
2372676 - Validation of a non-invasive device to measure intracavernous pressure as an index of p...
10212376 - Validation of the omron f3 blood pressure measuring device for use in the finger throug...
1761726 - Human autonomic responses to actual and simulated weightlessness.
17396026 - The initial response to inhaled nitric oxide treatment for intensive care unit patients...
Publication Detail:
Type:  Comparative Study; Evaluation Studies; Journal Article    
Journal Detail:
Title:  International angiology : a journal of the International Union of Angiology     Volume:  29     ISSN:  1827-1839     ISO Abbreviation:  Int Angiol     Publication Date:  2010 Oct 
Date Detail:
Created Date:  2010-10-06     Completed Date:  2011-01-28     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8402693     Medline TA:  Int Angiol     Country:  Italy    
Other Details:
Languages:  eng     Pagination:  426-30     Citation Subset:  IM    
Affiliation:
hugo.partsch@meduniwien.ac.at
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Compression Bandages*
Equipment Design
Hemodynamics*
Humans
Leg / blood supply*
Manometry / instrumentation*
Posture
Pressure
Reproducibility of Results
Sphygmomanometers
Time Factors
Transducers, Pressure*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Measuring venous pumping function by strain-gauge plethysmography.
Next Document:  Pascal's law and the dynamics of compression therapy: a study on healthy volunteers.