| Comparison of three portable instruments to measure compression pressure. | |
| | |
MedLine Citation:
|
PMID: 20924346 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
AIM: Measurement of interface pressure between the skin and a compression device has gained practical importance not only for characterizing the efficacy of different compression products in physiological and clinical studies but also for the training of medical staff. METHODS: A newly developed portable pneumatic pressure transducer (Picopress®) was compared with two established systems (Kikuhime® and SIGaT tester®) measuring linearity, variability and accuracy on a cylindrical model using a stepwise inflated sphygmomanometer as the reference. In addition the variation coefficients were measured by applying the transducers repeatedly under a blood pressure cuff on the distal lower leg of a healthy human subject with stepwise inflation. RESULTS: In the pressure range between 10 and 80 mmHg all three devices showed a linear association compared with the sphygmomanometer values (Pearson r>0.99). The best reproducibility (variation coefficients between 1.05-7.4%) and the highest degree of accuracy demonstrated by Bland-Altman plots was achieved with the Picopress® transducer. Repeated measurements of pressure in a human leg revealed average variation coefficients for the three devices of 4.17% (Kikuhime®), 8.52% (SIGaT®) and 2.79% (Picopress®). CONCLUSION: The results suggest that the Picopress® transducer, which also allows dynamic pressure tracing in connection with a software program and which may be left under a bandage for several days, is a reliable instrument for measuring the pressure under a compression device. |
| | |
Authors:
|
H Partsch; G Mosti |
Related Documents
:
|
8180746 - Pressure and volume control for local drug-delivery catheters: development of a new mic... 2135186 - Effect of subcutaneous implantation of anti-siphon devices on csf shunt function. 2372676 - Validation of a non-invasive device to measure intracavernous pressure as an index of p... 10212376 - Validation of the omron f3 blood pressure measuring device for use in the finger throug... 1761726 - Human autonomic responses to actual and simulated weightlessness. 17396026 - The initial response to inhaled nitric oxide treatment for intensive care unit patients... |
Publication Detail:
|
Type: Comparative Study; Evaluation Studies; Journal Article |
Journal Detail:
|
Title: International angiology : a journal of the International Union of Angiology Volume: 29 ISSN: 1827-1839 ISO Abbreviation: Int Angiol Publication Date: 2010 Oct |
Date Detail:
|
Created Date: 2010-10-06 Completed Date: 2011-01-28 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 8402693 Medline TA: Int Angiol Country: Italy |
Other Details:
|
Languages: eng Pagination: 426-30 Citation Subset: IM |
Affiliation:
|
hugo.partsch@meduniwien.ac.at |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Compression Bandages* Equipment Design Hemodynamics* Humans Leg / blood supply* Manometry / instrumentation* Posture Pressure Reproducibility of Results Sphygmomanometers Time Factors Transducers, Pressure* |