Abstract/OtherAbstract:
OBJECTIVE: The purpose of the present study was to compare the bursting pressure of large arteries sealed with hand made bipolar vessel sealer (BVS) using flexible electrosurgical system (HemoSaccab) and the BiClamp. MATERIAL AND METHOD: Sixteen common carotid arteries from a fresh cadaver were sealed using BVS and BiClamp 4 times sequentially for burst testing. A catheter was placed into the open end of the specimen and secured with a purse-string suture. The catheter was connected to a pressure monitor and saline was infused until there was leakage from the sealed end defining the bursting pressure in mmHg. Both devices were used to seal an additional 2 vessels in each group, which were sent for histologic examination. Student's t-test was performed. RESULTS: The BVS's mean burst pressure was statistically higher than that of the BiClamp (573 vs 442 mmHg) (p < 0.001). CONCLUSION: The flexible system of BVS is superior to the BiClamp device for sealing the common carotid artery because the BiClamp automatically finished the sealing process before the sealing area was completely dessicated.
Authors:
Tanvaa Tansatit; Youthana Kulvitit; Depicha Jindatip
Related Documents :
9752714 - The pressure hold/drop integrity test; its correlation to diffusive flow.
8263904 - A new device producing ambulatory intermittent pneumatic compression suitable for the t...
9601324 - Construction and calibration of a low-cost bandage pressure monitor.
25479984 - A vertical mouse and ergonomic mouse pads alter wrist position but do not reduce carpal...
7944814 - Experimental evaluation of different chordal preservation methods during mitral valve r...
376174 - Increased sheep lung vascular permeability caused by escherichia coli endotoxin.
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of the Medical Association of Thailand = Chotmaihet thangphaet     Volume:  89 Suppl 3     ISSN:  0125-2208     ISO Abbreviation:  J Med Assoc Thai     Publication Date:  2006 Sep 
Date Detail:
Created Date:  2007-08-27     Completed Date:  2007-09-27     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7507216     Medline TA:  J Med Assoc Thai     Country:  Thailand    
Other Details:
Languages:  eng     Pagination:  S169-73     Citation Subset:  IM    
Affiliation:
Department of Anatomy, Faculty of Medicine, Chulalongkorn University, King Rama IV Rd, Pratumwan District, Bangkok 10330, Thailand. tansatit@yahoo.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Cadaver
Carotid Artery, Common / surgery*
Electrosurgery / instrumentation*
Hemostasis, Surgical / instrumentation*
Humans
Surgical Instruments / economics

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Anatomic variations of the hepatic arteries in 200 patients done by angiography.