Comparison of anti-scatter grids for digital imaging with use of a direct-conversion flat-panel detector.
MedLine Citation:
PMID:  21971637     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
Our purpose in this study was to establish a selection standard for anti-scatter grids for a direct conversion flat-panel detector (FPD) system. As indices for grid evaluation, we calculated the selectivity, Bucky factor, and the signal-to-noise ratio improvement factor (SIF) by measuring rates of scatter transmission, primary transmission, and total transmission (based on the digitally displayed measurement values of the FPD system), using 4 acrylic phantoms of different thicknesses. The results showed that the SIF was less than 1.0 when the phantom thickness was 5 cm. When the phantom thickness was 25 cm and the grid ratio was 16:1, the SIF was 1.505 and 1.518 (maximum value) at 90 and 120 kV, respectively. Compared with the grid ratio of 12:1, the SIF at the grid ratio 16:1 was improved by 6.1% at 90 kV, and by 7.0% at 120 kV. In a direct-conversion FPD system, the grid ratio of 16:1 is considered adequate for eliminating the scattered-radiation effect when much scattered radiation is present, such as with a thick imaged object or a high X-ray tube voltage.
Authors:
Masayoshi Mizuta; Shigeru Sanada; Hiroyuki Akazawa; Toshifumi Kasai; Shuji Abe; Yasuhiro Ikeno; Shigeki Mitou
Related Documents :
7747227 - A controlled comparison of myelography, computed tomography, and magnetic resonance ima...
18537397 - A magnetic resonance imaging-based articulatory and acoustic study of "retroflex" and "...
21559477 - A mturquoise-based camp sensor for both flim and ratiometric read-out has improved dyna...
16079927 - Three-dimensional gadolinium-enhanced magnetic resonance angiography used as a "one-sto...
18425187 - Wide-range sounding of free-tropospheric water vapor with a differential-absorption lid...
17055097 - Clinical assessment of the use of the sonarray system for daily prostate localization.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-10-5
Journal Detail:
Title:  Radiological physics and technology     Volume:  -     ISSN:  1865-0341     ISO Abbreviation:  -     Publication Date:  2011 Oct 
Date Detail:
Created Date:  2011-10-5     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101467995     Medline TA:  Radiol Phys Technol     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Department of Radiology, University Hospital, Kyoto Prefectural University of Medicine, 465 Kajii-cho, Kawaramachi-Hirokoji, Kamigyo-ku, Kyoto, 602-8566, Japan, mizuta3@koto.kpu-m.ac.jp.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Insulin resistance and its association with metabolic syndrome components.
Next Document:  TWIST expression in hypopharyngeal cancer and the mechanism of TWIST-induced promotion o...