Comparison of ankle-brachial pressure index measurements using an automated oscillometric device with the standard Doppler ultrasound technique.
MedLine Citation:
PMID:  12608969     Owner:  NLM     Status:  MEDLINE    
BACKGROUND: The aim of the present study was to assess the accuracy of an automated oscillometric device (DINAMAP) in measuring ankle-brachial pressure index (ABPI) in the clinical setting. Results were then compared to those obtained using the standard Doppler technique. METHODS: A prospective comparative study of 50 healthy volunteers was carried out using a standard ward DINAMAP machine, hand-held Doppler and a 14 cm sphygmomanometer cuff. Three consecutive pressure readings were taken from all four limbs of the volunteers using both Doppler and DINAMAP techniques. The two techniques were carried out by different investigators who were blind to the findings of the other. The ABPI was calculated for each technique and the two techniques were compared. RESULTS: There was no significant agreement between the ABPI readings obtained by DINAMAP and those obtained by the standard Doppler technique. There was also a substantial variation in pressure readings produced by the DINAMAP for a given limb as measured by intraclass correlation of the three readings taken. CONCLUSIONS: There is no role for standard automated oscillometric devices in the calculation of ABPI in the vascular clinic.
Anantha Ramanathan; Philip J Conaghan; Andrew D Jenkinson; Christopher R Bishop
Related Documents :
21148199 - Ambulatory blood pressure monitoring and structural changes in carotid arteries in norm...
682389 - Segmental blood pressure of the leg and its clinical use.
1877599 - The hypertension prevention trial: assessment of the quality of blood pressure measurem...
7776359 - Comparison of oscillometric and intra-arterial blood pressure and pulse measurement.
19480489 - In-use comparison of blood pressure measurements from an automated blood pressure instr...
15802289 - Non-invasive pulsatile arterial pressure and stroke volume changes from the human finger.
8883639 - The effect of arteriovenous malformations on the distribution of intracerebral arterial...
10774869 - Blood distribution adaptations in paraplegics during posture changes: peripheral and ce...
19949639 - Metalloproteinase-3 genotype as a predictor of cardiovascular risk in hypertensive adol...
Publication Detail:
Type:  Comparative Study; Journal Article; Validation Studies    
Journal Detail:
Title:  ANZ journal of surgery     Volume:  73     ISSN:  1445-1433     ISO Abbreviation:  ANZ J Surg     Publication Date:  2003 Mar 
Date Detail:
Created Date:  2003-02-28     Completed Date:  2003-08-01     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  101086634     Medline TA:  ANZ J Surg     Country:  Australia    
Other Details:
Languages:  eng     Pagination:  105-8     Citation Subset:  IM    
Department of Surgery, Whittington Hospital, London, United Kingdom.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Ankle / blood supply,  physiology*,  ultrasonography*
Arm / blood supply,  physiology,  ultrasonography
Blood Pressure / physiology*
Blood Pressure Determination*
Brachial Artery / physiology*,  ultrasonography*
Leg / blood supply,  physiology,  ultrasonography
Middle Aged
Prospective Studies
Reference Values
Reproducibility of Results
Ultrasonography, Doppler*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Previous Document:  Endoluminal stent grafts for aortic diseases: experience at a major teaching hospital in Hong Kong.
Next Document:  Laparoscopic placement of peritoneal dialysis catheters: 7 years experience.