Document Detail


Comparison of Transpulmonary Thermodilution and Ultrasound Dilution Technique: Novel Insights into Volumetric Parameters from an Animal Model.
MedLine Citation:
PMID:  22349665     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
Especially in critically ill children with cardiac diseases, fluid management and monitoring of cardiovascular function are essential. Ultrasound dilution technique (UDT) was recently introduced to measure cardiac output (CO) and volumetric parameters, such as intrathoracic and end-diastolic blood volume. We compared UDT with the well-established transpulmonary thermodilution (TPTD) method (PiCCO) for determining CO measurements and derived volumes in a juvenile animal model. Experiments were performed in 18 ventilated, anesthetized piglets during normovolemia and after isovolemic hemodilution. At baseline and 20 min after each step of isovolemic hemodilution, 3 independent measurements of CO and volumetric parameters were conducted with TPTD and UDT, consecutively, under hemodynamically stable conditions. We observed comparable results for CO measurements with both methods (mean 1.98 l/min; range 1.12-2.87) with a percentage error of 17.3% (r = 0.92, mean bias = 0.28 l/min). Global end-diastolic volume (GEDV) and intrathoracic blood volume (ITBV) by TPTD were almost two times greater than analogous volumes [central blood volume (CBV); total end-diastolic volume (TEDV)] quantified by UDT (CBV = 0.58 × ITBV + 27.1 ml; TEDV = 0.48 × GEDV + 23.1 ml). CO measurements by UDT were found to be equivalent and hence interchangeable with TPTD. Discrepancies in volumetric parameters could either be due to the underlying algorithm or different types of indicators (diffusible vs. nondiffusible). Compared with the anatomically defined heart volume, TPTD seems to overestimate end-diastolic volumes. Future studies will be necessary to assign these results to critically ill children and to validate volumetric parameters with reference techniques.
Authors:
Martin Boehne; Florian Schmidt; Lars Witt; Harald Köditz; Michael Sasse; Robert Sümpelmann; Harald Bertram; Armin Wessel; Wilhelm Alexander Osthaus
Related Documents :
2239665 - Are ioxaglate and iopamidol equally safe and well tolerated in cardiac angiography? a r...
22796715 - Diet and blood pressure in 18-74-year-old adults: the french nutrition and health surve...
22331415 - Cardiac output and sympathetic vasoconstrictor responses during upright tilt to presync...
22709435 - Noninvasive cpap with face mask: comparison among new venturi devices and the boussigna...
12955175 - Assessment of cardiac preload and left ventricular function under increasing levels of ...
17785625 - Change in salt intake affects blood pressure of chimpanzees: implications for human pop...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-2-14
Journal Detail:
Title:  Pediatric cardiology     Volume:  -     ISSN:  1432-1971     ISO Abbreviation:  -     Publication Date:  2012 Feb 
Date Detail:
Created Date:  2012-2-21     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8003849     Medline TA:  Pediatr Cardiol     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Department of Pediatric Cardiology and Intensive Care Medicine, Hannover Medical School, Hannover, Germany, boehne.martin@mh-hannover.de.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Percutaneous Transluminal Mitral Commissurotomy for Rheumatic Mitral Stenosis in a 5-Year-Old Child.
Next Document:  Comparison of maximum vasoactive inotropic score and low cardiac output syndrome as markers of early...