Comparison of FloTrac cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements.
MedLine Citation:
PMID:  17666154     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Arterial pulse waveform analysis has been proposed for cardiac output (CO) determination and monitoring without calibration or thermodilution (FloTrac/Vigileo; Edwards Lifesciences, Irvine, CA, USA). The accuracy and clinical applicability of this technology has not been fully evaluated. We designed this prospective study to compare the accuracy of the FloTrac system (CO(FT)) vs. pulmonary artery catheter standard bolus thermodilution (CO(PAC) ) in patients undergoing coronary artery bypass grafting. METHODS: We studied 11 patients referred for coronary artery bypass grafting. CO(FT) and CO(PAC) were determined at six time points in the operating room including before and 5 min after volume expansion (500 mL 6% hetastarch). Measurements were performed on arrival in the intensive care unit and every 4 h afterwards. Bland-Altman analysis was used to assess the agreement between CO(FT) and CO(PAC). RESULTS: CO(PAC) ranged from 2.0 to 7.6 L min-1 and CO(FT) ranged from 1.9 to 8.2 L min-1. There was a significant relationship between CO(PAC) and CO(FT) (r = 0.662; P < 0.001). Agreement between CO(PAC) and CO(FT) was -0.26 +/- 0.87 L min-1. Volume expansion induced a significant increase in both CO(PAC) and CO(FT) (from 3.4 +/- 0.8 to 4.4 +/- 1.0 L min-1; P < 0.001 and from 3.9 +/- 1.2 to 5.0 +/- 1.1 L min-1; P < 0.001, respectively) and there was a significant relationship between percent change in CO(PAC) and CO(FT) following volume expansion (r = 0.722; P = 0.01). CONCLUSION: We found clinically acceptable agreement between CO(FT) and CO(PAC) in this setting. This new device has potential clinical applications.
Authors:
M Cannesson; Y Attof; P Rosamel; P Joseph; O Bastien; J-J Lehot
Related Documents :
21788114 - The transition from custom-made to standardized multibranched thoracoabdominal aortic s...
21954744 - Aorto-iliac stent graft infection complicated by endotension and consequent rupture of ...
21741854 - Short-term results of retrograde pulmonary embolectomy in massive and submassive pulmon...
6970304 - Complement activation and neutropenia occurring during cardiopulmonary bypass.
6261204 - Surgical palatomaxilloschisis: an avenue to the pterygomaxillary space.
14755814 - Ulnar artery catheterization with occlusion of corresponding radial artery.
Publication Detail:
Type:  Clinical Trial; Comparative Study; Journal Article     Date:  2007-08-01
Journal Detail:
Title:  European journal of anaesthesiology     Volume:  24     ISSN:  0265-0215     ISO Abbreviation:  Eur J Anaesthesiol     Publication Date:  2007 Oct 
Date Detail:
Created Date:  2007-09-07     Completed Date:  2008-01-10     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8411711     Medline TA:  Eur J Anaesthesiol     Country:  England    
Other Details:
Languages:  eng     Pagination:  832-9     Citation Subset:  IM    
Affiliation:
Hospives Civils de Lyon, Hôpital Louis Pradel, Department of Anesthesiology, Bron, France. maxime_cannesson@hotmail.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Aged, 80 and over
Blood Pressure Determination / methods
Calibration
Cardiac Output*
Catheterization, Swan-Ganz
Coronary Artery Bypass*
Female
Humans
Intensive Care
Male
Middle Aged
Monitoring, Intraoperative / instrumentation,  methods*
Monitoring, Physiologic
Postoperative Period
Prospective Studies
Pulmonary Artery
Pulsatile Flow*
Signal Processing, Computer-Assisted
Thermodilution / methods
Time Factors