Document Detail


Comparison of I-125 sources used for permanent interstitial implants.
MedLine Citation:
PMID:  11339765     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The increase in the number of manufacturers of 125I sources used in prostate brachytherapy has generated many questions in the radiation oncology community. In this investigation, the physical and dosimetric characteristics were evaluated for the following sources listed by marketing company and source model: Nycomed-Amersham 6711 (OncoSeed), Nycomed-Amersham 6702, Mentor IoGold, UroMed Symmetra, Imagyn IsoSTAR, UroCor, (PSA, Mallincrkrodt) ProstaSeed, Syncor PharmaSeed, SourceTech Medical, (BARD) 125Implant (BrachySource), Med-Tec I-Plant, Best Medical Model 2301, DraxImage BrachySeed, and International Brachytherapy, Inc. (IBT) InterSource125. The investigation examined the differences in design, construction, and the dosimetric characteristics created from each source. The dosimetric characteristics of the new sources were compared to that of the Amersham 6711 source. Parameter studies have led to the development of a simple equation that can be used to clinically convert the standard 6711 source strength to an equivalent strength of a new source.
Authors:
B H Heintz; R E Wallace; J M Hevezi
Related Documents :
522685 - A pilot experiment on the inter-examiner reliability of short essay questions.
12244665 - Diagnostic investigations. part 1. lumbar puncture.
16846925 - Feeling of ear fullness in acute sensorineural hearing loss.
2055455 - Mitochondrial dna heteroplasmy maintained in natural populations of drosophila simulans...
11241955 - Aspects of the ecology of proteocephalid cestodes parasites of sorubim lima (pimelodida...
17690595 - Open-label escitalopram treatment for pathological skin picking.
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Medical physics     Volume:  28     ISSN:  0094-2405     ISO Abbreviation:  Med Phys     Publication Date:  2001 Apr 
Date Detail:
Created Date:  2001-05-07     Completed Date:  2001-09-27     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  0425746     Medline TA:  Med Phys     Country:  United States    
Other Details:
Languages:  eng     Pagination:  671-82     Citation Subset:  IM    
Affiliation:
Department of Radiology, University of Texas Health Science Center at San Antonio, 78229, USA. bhheintz@hotmail.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Absorption
Adsorption
Anisotropy
Brachytherapy / instrumentation*,  methods*
Ceramics
Glass
Iodine Radioisotopes*
Monte Carlo Method
Radiometry*
Resins, Plant
Chemical
Reg. No./Substance:
0/Ceramics; 0/Iodine Radioisotopes; 0/Resins, Plant
Comments/Corrections
Comment In:
Med Phys. 2002 Apr;29(4):641   [PMID:  11991137 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Dosimetry of the I-Plant Model 3500 iodine-125 brachytherapy source.
Next Document:  A theoretical derivation of the nomograms for permanent prostate brachytherapy.