Document Detail


Comparing three CPR feedback devices and standard BLS in a single rescuer scenario: a randomised simulation study.
MedLine Citation:
PMID:  24215730     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
Background Efficiently performed basic life support (BLS) after cardiac arrest is proven to be effective. However, cardiopulmonary resuscitation (CPR) is strenuous and rescuers' performance declines rapidly over time. Audio-visual feedback devices reporting CPR quality may prevent this decline. We aimed to investigate the effect of various CPR feedback devices on CPR quality. Methods In this open, prospective, randomised, controlled trial we compared three CPR feedback devices (PocketCPR®, CPRmeter®, iPhone app PocketCPR®) with standard BLS without feedback in a simulated scenario. 240 trained medical students performed single rescuer BLS on a manikin for 8minutes. Effective compression (compressions with correct depth, pressure point and sufficient decompression) as well as compression rate, flow time fraction and ventilation parameters were compared between the four groups. Results Study participants using the PocketCPR® performed 17±19% effective compressions compared to 32±28% with CPRmeter®, 25±27% with the iPhone app PocketCPR®, and 35±30% applying standard BLS (PocketCPR® vs. CPRmeter® p=0.007, PocketCPR® vs. standard BLS p=0.001, others: ns). PocketCPR® and CPRmeter® prevented a decline in effective compression over time, but overall performance in the PocketCPR® group was considerably inferior to standard CPR. Compression depth and rate were within the range recommended in the guidelines in all groups. Conclusion While we found differences between the investigated CPR feedback devices, overall BLS quality was suboptimal in all groups. Surprisingly, effective compression was not improved by any CPR feedback device compared to standard BLS. All feedback devices caused substantial delay in starting CPR, which may worsen outcome.
Authors:
Bernhard Zapletal; Robert Greif; Dominik Stumpf; Franz Josef Nierscher; Sophie Frantal; Moritz Haugk; Kurt Ruetzler; Christoph Schlimp; Henrik Fischer
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2013-11-8
Journal Detail:
Title:  Resuscitation     Volume:  -     ISSN:  1873-1570     ISO Abbreviation:  Resuscitation     Publication Date:  2013 Nov 
Date Detail:
Created Date:  2013-11-12     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0332173     Medline TA:  Resuscitation     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2013. Published by Elsevier Ireland Ltd.
Affiliation:
Resident, Department of Anaesthesia, General Intensive Care and Pain Medicine, Medical University Vienna, Austria.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A novel nickel-chelated surfactant for affinity-based aqueous two-phase micellar extraction of histi...
Next Document:  Applying Lessons from Commercial Aviation Safety and Operations to Resuscitation.