Recently, a new oxygenator (Dideco 903 [D903], Dideco, Mirandola, Italy) has been introduced to the perfusion community, and we set about testing its oxygen transfer performance and then comparing it to two other models. This evaluation was based on the comparison between oxygen transfer slope, gas phase arterial oxygen gradients, degree of blood shunting, maximum oxygen transfer, and diffusing capacity calculated for each membrane. Sixty patients were randomized into three groups of oxygenators (Dideco 703 [D703], Dideco; D903; and Quadrox, Jostra Medizintechnik AG, Hirrlingen, Germany) including 40/20 M/F of 68.6 +/- 11.3 years old, with a body weight of 71.5 +/- 12.1 kg, a body surface area (BSA) of 1.84 +/- 0.3 m(2), and a theoretical blood flow rate (index 2.4 times BSA) of 4.4 +/- 0.7 L/min. The maximum oxygen transfer (VO(2)) values were 313 mL O(2)/min (D703), 579 mL O(2)/min (D903), and 400 mL O(2)/min (Quadrox), with the D903 being the most superior (P < 0.05). Oxygen (O(2)) gradients were 320 mm Hg (D703), 235 mm Hg (D903), and 247 mm Hg (Quadrox), meaning D903 and Quadrox are more efficient versus the D703 (P < 0.05). Shunt fraction (Qs/Qt) and diffusing capacity (DmO(2)) were comparable (P = ns). Diffusing capacity values indexed to BSA (DmO(2)/m(2)) were 0.15 mL O(2)/min/mm Hg/m(2) (D703), 0.2 mL O(2)/min/mm Hg/m(2) (D903), and 0.18 mL O(2)/min/mm Hg/m(2) (Quadrox) with D903 outperforming D703 (P < 0.0005). During hypothermia (32.0 +/- 0.3 degrees C), there was a lower absolute and relative VO(2 )for all three oxygenators (P = ns). The O(2) gradients, DmO(2) and DmO(2)/m(2), were significantly lower for all oxygenators (P < 0.01). Also, Qs/Qt significantly rose for all oxygenators (P < 0.01). The oxygen transfer curve is characteristic to each oxygenator type and represents a tool to quantify oxygenator performance. Using this parameter, we demonstrated significant differences among commercially available oxygenators. However, all three oxygenators are considered to meet the oxygen needs of the patients.
Authors:
David Jegger; Hendrik T Tevaearai; Iker Mallabiabarrena; Judith Horisberger; Isabelle Seigneul; Ludwig K von Segesser
Related Documents :
7812538 - Effect of cooling on oxygen consumption in febrile critically ill patients.
19378658 - Effect of oxygen partial pressure and chemical oxygen demand loading on the biofilm pro...
16899838 - Peritoneal perfusion with oxygenated perfluorocarbon augments systemic oxygenation.
8710798 - Effects of variation in oxygen tension on responses of the human fetoplacental vasculat...
18405898 - Central cardiovascular actions of angiotensin ii in trout.
25278638 - Analytical study of cavitation surge in a hydraulic system.
Publication Detail:
Type:  Comparative Study; Journal Article; Randomized Controlled Trial    
Journal Detail:
Title:  Artificial organs     Volume:  31     ISSN:  0160-564X     ISO Abbreviation:  Artif Organs     Publication Date:  2007 Apr 
Date Detail:
Created Date:  2007-04-17     Completed Date:  2007-05-31     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7802778     Medline TA:  Artif Organs     Country:  United States    
Other Details:
Languages:  eng     Pagination:  290-300     Citation Subset:  IM    
Affiliation:
Department of Cardiovascular Surgery, Centre Hospitalier Universitaire Vaudois (CHUV), Lausanne, Switzerland. David.Jegger@chuv.ch
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Cardiopulmonary Bypass / instrumentation*,  methods
Coronary Artery Bypass / instrumentation*,  methods
Extracorporeal Membrane Oxygenation / instrumentation*,  methods
Female
Humans
Hypothermia, Induced*
Male
Oxygen / metabolism*
Oxygen Consumption
Oxygenators, Membrane*
Temperature
Chemical
Reg. No./Substance:
7782-44-7/Oxygen

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Pulse pressure respiratory variation as an early marker of cardiac output fall in experimental hemor...
Next Document:  Dynamic characteristics of a magnetically levitated impeller in a centrifugal blood pump.