Comparing the accuracy of ES-BC, EIS-GS, and ES Oxi on body composition, autonomic nervous system activity, and cardiac output to standardized assessments.
Jump to Full Text
MedLine Citation:
PMID:  22915943     Owner:  NLM     Status:  PubMed-not-MEDLINE    
Abstract/OtherAbstract:
BACKGROUND AND PURPOSE: THE ELECTRO SENSOR COMPLEX (ESC) IS SOFTWARE THAT COMBINES THREE DEVICES USING BIOELECTRICAL IMPEDANCE, GALVANIC SKIN RESPONSE, AND SPECTROPHOTOMETRY: (1) ES-BC (Electro Sensor-Body Composition; LD Technology, Miami, FL) to assess body composition, (2) EIS-GS (Electro Interstitial Scan-Galvanic Skin; LD Technology) to predict autonomic nervous system activity, and (3) ES Oxi (Electro Sensor Oxi; LD Technology) to assess cardiac output. The objective of this study was to compare each to a standardized assessment: ES-BC to dual-energy X-ray absorptiometry (DXA), EIS-GS to heart rate variability, and ES Oxi to BioZ Dx Diagnostic System (BioZ Dx; SonoSite Inc, Bothell, WA).
PATIENTS AND METHODS: The study was conducted in two waves. Fifty subjects were assessed for body composition and autonomic nervous system activity. Fifty-one subjects were assessed for cardiac output.
RESULTS: We found adequate relative and absolute agreement between ES-BC and DXA for fat mass (r = 0.97, P < 0.001) with ES-BC overestimating fat mass by 0.1 kg and for body fat percentage (r = 0.92, P < 0.001) with overestimation of fat percentage by 0.4%. For autonomic nervous system activity, we found marginal relative agreement between EIS-GS and heart rate variability by using EIS-GS as the predictor in a linear regression equation (adjusted R(2) = 0.56, P = 0.03). For cardiac output, adequate relative and absolute agreement was found between ES Oxi and BioZ Dx at baseline (r = 0.60, P < 0.001), after the first exercise stage (r = 0.79, P < 0.001), and after the second exercise stage (r = 0.86, P < 0.001). Absolute agreement was found at baseline and after both bouts of exercise; ES Oxi overestimated baseline and stage 1 exercise cardiac output by 0.3 L/minute and 0.1 L/minute, respectively, but exactly estimated stage 2 exercise cardiac output.
CONCLUSION: ES-BC and ES Oxi accurately assessed body composition and cardiac output compared to standardized instruments, whereas EIS-GS showed marginal predictive ability for autonomic nervous system activity. The ESC software managing the three devices would be useful to help detect complications related to metabolic syndrome, diabetes, and cardiovascular disease and to noninvasively and rapidly manage treatment follow-up.
Authors:
John E Lewis; Stacey L Tannenbaum; Jinrun Gao; Angelica B Melillo; Evan G Long; Yaima Alonso; Janet Konefal; Judi M Woolger; Susanna Leonard; Prabjot K Singh; Lawrence Chen; Eduard Tiozzo
Related Documents :
22581743 - Gastric emptying and intestinal absorption of electrolytes,and exercise performance in ...
656233 - Long-term haemodynamic evaluation of aortic pericardial xenograft.
21996773 - Physical and mental features of elderly persons who experienced group exercise for care...
Publication Detail:
Type:  Journal Article     Date:  2011-09-16
Journal Detail:
Title:  Medical devices (Auckland, N.Z.)     Volume:  4     ISSN:  1179-1470     ISO Abbreviation:  Med Devices (Auckl)     Publication Date:  2011  
Date Detail:
Created Date:  2012-08-23     Completed Date:  2012-10-02     Revised Date:  2013-05-30    
Medline Journal Info:
Nlm Unique ID:  101566041     Medline TA:  Med Devices (Auckl)     Country:  New Zealand    
Other Details:
Languages:  eng     Pagination:  169-77     Citation Subset:  -    
Affiliation:
Department of Psychiatry and Behavioral Sciences, University of Miami Miller School of Medicine, Miami, FL.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Med Devices (Auckl)
Journal ID (iso-abbrev): Med Devices (Auckl)
Journal ID (publisher-id): Medical Devices: Evidence and Research
ISSN: 1179-1470
Publisher: Dove Medical Press
Article Information
Download PDF
© 2011 Lewis et al, publisher and licensee Dove Medical Press Ltd.
License:
collection publication date: Year: 2011
Electronic publication date: Day: 16 Month: 9 Year: 2011
Volume: 4First Page: 169 Last Page: 177
ID: 3417887
PubMed Id: 22915943
DOI: 10.2147/MDER.S24291
Publisher Id: mder-4-169

Comparing the accuracy of ES-BC, EIS-GS, and ES Oxi on body composition, autonomic nervous system activity, and cardiac output to standardized assessments
John E Lewis1
Stacey L Tannenbaum1
Jinrun Gao3
Angelica B Melillo1
Evan G Long1
Yaima Alonso2
Janet Konefal1
Judi M Woolger2
Susanna Leonard1
Prabjot K Singh1
Lawrence Chen1
Eduard Tiozzo1
1Department of Psychiatry and Behavioral Sciences, University of Miami Miller School of Medicine, Miami, FL
2Department of Medicine, University of Miami Miller School of Medicine, Miami, FL
3State Farm Insurance, Bloomington, IL, USA
Correspondence: Correspondence: John E Lewis, University of Miami Miller School of Medicine, Department of Psychiatry and Behavioral Sciences, 1120 NW 14th Street, Suite #1474 (D21), Miami, FL 33136, USA, Tel +1 305 243 6227, Fax +1 305 243 3648, Email jelewis@miami.edu

Introduction

During the past twenty years, the US has succumbed to a pervasive obesity epidemic.1,2 In 1980, fewer than 47% of Americans were overweight (body mass index [BMI] >25 kg/m2) and less than 15% were obese (BMI >30 kg/m2). Today, approximately 67% of Americans are overweight and over 27% are obese.2,3 Unfortunately, a concurrent and parallel rise in metabolic syndrome exists and is accompanied by complications of diabetes, hypertension, and cardiovascular disease (CVD). The prevalence of metabolic syndrome appears to be roughly 33% of the American population according to standard guidelines.4,5

An interaction of factors contributes to the complications of metabolic syndrome on multiple organ system levels. For example, central fat is linked to increased inflammation, risk of diabetes, and metabolic syndrome.68 Autonomic nervous system imbalance, including vasoconstriction, increases heart rate and levels of epinephrine and norepinephrine, and hemodynamic disorders (eg, systemic vascular resistance or cardiac output deficits) are also pathological factors leading to hypertension, diabetes, and CVD.9,10

A software complex (Electro Sensor Complex [ESC]) managing three devices, ES-BC (Electro Sensor-Body Composition; LD Technology, Miami, FL), EIS-GS (Electro Interstitial Scan-Galvanic Skin; LD Technology), and ES Oxi (Electro Sensor Oxi; LD Technology), can rapidly measure all of the aforementioned factors related to metabolic syndrome: (1) body composition, (2) autonomic nervous system activity, and (3) cardiac output, respectively. The ES-BC device uses bioelectrical impedance analysis to estimate percent body fat, but it has never been compared to a gold standard for body composition, such as dual-energy X-ray absorptiometry (DXA).11 The EIS-GS response device has been shown to reflect autonomic nervous system activity, however it has never been used to predict a standardized assessment, such as heart rate variability (HRV).12 The ES Oxi device, using a digital pulse oximeter with photoelectric plethysmography, can estimate arterial stiffness,13 but it has never been compared to a standardized assessment of cardiac output, such as with the BioZ Dx Diagnostic System (BioZ Dx; SonoSite Inc, Bothell, WA) that uses impedance cardiography. Thus, if the assessments from the three devices (ES-BC, EIS-GS, and ES Oxi) are not different when compared to standardized assessments, then the ESC software would be particularly useful for detecting complications related to metabolic syndrome, diabetes, and CVD and to potentially assist in managing treatment follow-up. In addition, the ESC software managing the three devices can assess patients within 5 minutes, providing a rapid and convenient clinical tool.

Given the enormous public health burden of metabolic syndrome and the associated morbidities of diabetes, hypertension, and CVD, an accurate, easy-to-use, low-cost, and rapid assessment of body composition, autonomic nervous system activity, and cardiac output would greatly improve the assessment and treatment of at-risk individuals, especially if it was performed by trained nonclinical personnel, ie, individuals assisting physicians. Therefore, the objectives of this study were to compare the: (1) ES-BC to DXA, the gold standard of body composition assessment; (2) EIS-GS response device to a standardized assessment of autonomic nervous system activity (ie, HRV); and (3) ES Oxi device to a standardized assessment of cardiac output (BioZ Dx). The results of the study will be applicable for clinicians interested in simplifying the screening and management plan of metabolic syndrome, diabetes, hypertension, and CVD.


Materials and methods

This study was performed in two waves; the first to assess body composition and autonomic nervous system activity and the second to assess cardiac output. Both waves of the study were conducted with the approval of the University of Miami Institutional Review Board for human subjects’ research, participants signed informed consent before commencing in the study, and confidentiality was maintained for all participants’ data.

Wave 1
Study participants

All participants (n = 50) were recruited by referrals at the University of Miami Miller School of Medicine campus throughout 2009. The sample comprised of 40% males (n = 20) and 60% females (n = 30) with a mean age of 32.3 years (standard deviation [SD] = 10.2; range 20–62).

Inclusion and exclusion criteria

Potential participants were ≥18 years of age and had expressed an interest in having their body composition and autonomic nervous system activity assessed. Subjects were not enrolled in the study if they: (1) were unable to consent to the study; (2) wore an automatic external defibrillator device; (3) had erratic, accelerated, or mechanically-controlled irregular heart rhythms; (4) had arterial fibrillation/flutter; (5) had atrioventricular block; (6) had dermatological lesions or excessive hair that would be in contact with the placement of the electrodes on the devices; (7) had any implanted electronic device; (8) were during or within ± 2 days of menstruation; (9) had a fever above 37°C during the assessment; (10) were taking diuretics; (11) had a history of renal or heart failure; (12) had excessively used alcohol or stimulants (amphetamines) 12 hours before the examination; (13) had diarrhea during the assessment; (14) had engaged in intense physical activity or the use of a sauna 8 hours before the examination; (15) had edema; and/or (16) were pregnant.

Procedure

Each participant completed a basic demographics and medical history questionnaire on the day of the assessment. Height, to the nearest 0.5 cm, and weight, to the nearest 0.5 kg, were measured with a standard scale. Subjects were assessed with the ES-BC and EIS-GS, including HRV, at the University of Miami Center for Complementary and Integrative Medicine and then taken to the Pediatric Clinical Research Laboratory for the DXA scan. The entire assessment took less than 1 hour for each participant.

Body composition: ES-BC assessment

ES-BC uses a single frequency electrical bioimpedance analyzer in tetrapolar mode to assess body fat percentage and fat mass. The device measures current, voltage, and phase angle and calculates impedance, resistance, and reactance. The software computes these values using accepted peer-reviewed published algorithms tailored to the general population and persons who are obese.1417 The software automatically selects the algorithm based on the participant’s values entered into the database (ie, gender, height, weight, age, and activity level).16

Body composition: DXA assessment

DXA was conducted with a Lunar Prodigy Pro™ scanner with enCORE software 7.5 (GE Healthcare, Madison, WI), which has previously demonstrated adequate reliability for body composition assessment.18 DXA is based on the three-compartment model of body composition and uses dual-energy X-ray to measure fat, lean soft tissue, and bone mineral in approximately 10 minutes.11

Autonomic nervous system activity: EIS-GS assessment

The EIS-GS utilizes a universal serial bus (USB) plug and play hardware including an interface box, disposable electrodes, and reusable plates and cables with software installed on a computer. The system performs bioimpedance in bipolar mode with direct current and measures the electrical conductivity of 11 pathways of the body, each recorded twice from anode to cathode (SDC+) and then from cathode to anode (SDC−). Electrode polarization does not affect the bioimpedance measurement,19 and the transmission of the current from the electrode to the hardware is performed by chronoamperometry. 20 With direct current, the plasma membrane acts as an insulator, and the current is not able to penetrate the cell, so most of the current flows around the cell and remains in the interstitial fluid.19 Analysis of the direct current in electrolytic solution is performed at both the anode and the cathode. The electrochemical reaction at the cathode is: 2H2O + 2e− = H2 (gas) + 2OH-(base), and the electrochemical reaction for water at the anode is: 2H2O = O2(gas) + 4H+ + 4e−(acid). Finally, spectral analysis by EIS-GS is an application of the fast Fourier transform of the entire signal composed of the conductivity measurements. The fast Fourier transform components of EIS-GS are high frequency (HF), low frequency (LF), and very low frequency (VLF).

Autonomic nervous system activity: HRV analysis

Standardized assessment of the autonomic nervous system involves electrocardiogram rhythm analysis of the NN and RR intervals both in the full domain of activity and in the spectral analysis.21 The HRV assessment records three main spectral components in 2–5 minutes: VLF, LF, and HF.22 The HF percent represents parasympathetic (vagal) activity, and the LF percent represents sympathetic system activity.21,22

Wave 2
Study participants

All participants (n = 51) were recruited by referrals at the University of Miami Miller School of Medicine campus throughout 2010. The sample was comprised of 49% males (n = 25) and 51% females (n = 26) with a mean age of 31.1 years (SD = 10.8; range 18–65).

Inclusion and exclusion criteria

In addition to the same inclusion/exclusion criteria as those in wave 1, potential participants for wave 2 could not have: (1) had dyes recently introduced into the bloodstream, such as methylene blue, indocyanine green, indigo carmine, and fluorescein; (2) had significant levels of dysfunctional hemoglobins, such as carboxyhemoglobin or methemoglobin; (3) had any condition restricting blood flow, such as severe systemic vascular resistance; and/or (4) worn fingernail polish or false fingernails during the testing. Any of these could affect the accuracy of peripheral oxygen saturation of arterial hemoglobin (SpO2%) measurement from the oximeter, so subjects meeting any of these criteria were excluded.

Procedure

Each participant completed a basic demographics and medical history questionnaire at the assessment. Subjects were assessed with ES Oxi and BioZ Dx devices in the University of Miami Center for Complementary and Integrative Medicine and Medical Wellness Center. Each subject began the assessment by sitting at rest for 5 minutes and had cardiac output measured simultaneously by ES Oxi and BioZ Dx devices. Then, the subject walked for 6 minutes on a treadmill at low-to-moderate intensity (less than 70% of maximal age-adjusted predicted heart rate [220-age]), and cardiac output was measured simultaneously immediately afterward by ES Oxi and BioZ Dx devices. Finally, each subject walked for a second 6-minute bout of exercise on the treadmill at the same intensity, and cardiac output was measured simultaneously immediately afterward by ES Oxi and BioZ Dx devices. Thus, three measurements were completed: one at rest and two following 6-minute bouts of low-to-moderate intensity exercise on a treadmill. The entire assessment took less than 1 hour for each participant.

Cardiac output: ES Oxi

The ES Oxi module utilizes a USB plug and play hardware device including an electronic box and reusable SpO2 probe and software installed on a computer. It is intended to: (1) spot check or monitor SpO2% and pulse rate and (2) analyze the pulse waveform (photoelectrical plethysmography) provided by the oximeter. The analysis provides indicators of arterial stiffness.23 The ES Oxi uses the same algorithms as BioZ Dx to derive mean arterial pressure, systemic vascular resistance, and stroke volume.24

Cardiac output: BioZ Dx impedance cardiography assessment

BioZ Dx is a noninvasive heart monitor based on technology known as impedance cardiography. BioZ Dx provides the clinician with information about the heart’s ability to deliver blood to the body, the amount of force the heart exerts at each beat, and the amount of fluid in the chest. Specially-designed bioimpedance sensors placed on the neck and chest monitor 12 different parameters, including cardiac output, contractility, systemic vascular resistance, and thoracic fluid content. These sensors monitor the electrical conductivity of the thoracic part of the body.24

Data analysis for study waves 1 and 2

Data were analyzed using SPSS for Windows (v 18; SPSS Inc, Chicago, IL). Frequency and descriptive statistics were calculated on all variables. We used chi-square to compare subjects by body fat percentage categories on the ES-BC and DXA assessments according to criteria of the American Council on Exercise for women and men. For women, essential fat is 10%–13%, athlete is 14%–20%, fitness is 21%–24%, average is 25%–31%, and obesity is 32% or greater. For men, essential fat is 2%–5%, athlete is 6%–13%, fitness is 14%–17%, average is 18%–24%, and obesity is 25% or greater. We used independent sample t-tests to compare differences between ES-BC and DXA for body fat percentage and fat mass and ES Oxi and BioZ Dx for cardiac output. We used Pearson product-moment correlation to determine the strength of the relationships between: (1) ES-BC and DXA for body fat percentage and fat mass and (2) ES Oxi and BioZ Dx at baseline and after both bouts of exercise for cardiac output. To evaluate the presence of systematic error between each pair of variables, Pearson product-moment correlation was utilized to compare the average of the methods and the mean difference score for each method. Bland–Altman analysis was used to determine absolute limits of agreement for each variable.25,26 For autonomic nervous system activity, we used linear regression to determine the predictability of EIS-GS HF (normal range 22%–34%) on HRV LF (normal range 22%–46%). We evaluated the adjusted R2 of the model and the 95% confidence intervals of the intercept and slope to assess the level of systematic error between the two variables. We used α = 0.05 as the criterion for statistical significance.


Results

Overall, subjects completed the assessments without reporting any adverse events, and the assessments were completed in a timely fashion.

Clinical measurements

For wave 1, the participants’ (n = 50) mean height was 169.5 cm (SD = 11.0, range 150–190.5), mean weight was 72.0 kg (SD = 19.0, range 45–117), and mean BMI was 25.1 kg/m2 (SD = 6.2, range 18.4–48.7). For wave 2, the participants’ (n = 51) mean height was 172.0 cm (SD = 11.0, range 152–193), mean weight was 78.5 kg (SD = 19.0, range 54–136), and mean BMI was 26.4 kg/m2 (SD = 6.6, range 19.6–56.7).

Body composition

The chi-square analysis between ES-BC and DXA resulted in a disparate proportion of classifications (χ2[9] = 46.3, P < 0.001). According to ES-BC body fat percentage, subjects were classified as: (1) athlete (n = 4, 8%), fitness (n = 9, 18%), average (n = 20, 40%), and obese (n = 17, 34%). According to DXA body fat percentage, subjects were classified as: (1) athlete (n = 8, 16%), fitness (n = 11, 22%), average (n = 15, 30%), and obese (n = 16, 32%). For example, DXA classified eight subjects in the athlete category, whereas ES-BC only classified four subjects in the athlete category, and only two subjects were classified as an athlete according to both devices.

The mean values according to ES-BC for body fat percentage and fat mass were statistically similar to DXA. We found adequate relative agreement between ES-BC and DXA, as the correlation values of these two methods for fat mass was r = 0.97, P < 0.001 and for body fat percentage was r = 0.92, P < 0.001. Compared to DXA in absolute terms, ES-BC overestimated fat mass by 0.1 kg and body fat percentage by 0.4% with wide limits of agreement for each variable (fat mass −7.0 to 7.2 kg and body fat percentage −8.8% to 9.5%) (Table 1, Figures 1 and 2). For fat mass, systematic error was observed between ES-BC and DXA, as indicated by a significant correlation of the difference value and the average score (r = 0.60, P < 0.001). Systematic error was also observed for body fat percentage with a significant correlation found between the difference value and the average score (r = 0.61, P < 0.001). At higher levels of fat mass and body fat percentage, ES-BC underestimated both values compared to DXA.

Autonomic nervous system activity

Utilizing EIS-GS HF as the predictor variable and the HRV LF as the dependent variable in a linear regression, the model was statistically significant (F[1, 49] = 63.8, P < 0.001). The adjusted R2 value was 0.56. The 95% confidence interval of the intercept was 1.02 and 14.7 (intercept does not include 0), and the 95% confidence interval of the slope was 0.78 and 1.30 (slope does include 1), thus demonstrating some level of systematic error.

Cardiac output

Adequate relative agreement was found between ES Oxi and BioZ Dx based on strong correlation values of r = 0.60, P < 0.001 at baseline, r = 0.79, P < 0.001 after the first exercise stage, and r = 0.86, P < 0.001 after the second exercise stage. The mean cardiac output values according to ES Oxi for baseline, stage 1 exercise, and stage 2 exercise were statistically similar to BioZ Dx. Compared with BioZ Dx in absolute terms, ES Oxi overestimated baseline and stage 1 exercise cardiac output by 0.3 L/minute and 0.1 L/minute, respectively, but exactly estimated stage 2 exercise cardiac output with narrow limits of agreement for each variable (baseline cardiac output −1.9 to 2.4 L/minute; stage 1 exercise cardiac output −1.5 to 1.6 L/minute; and stage 2 exercise cardiac output −1.3 to 1.4 L/minute) (Table 1, Figures 35). No systematic bias was demonstrated for baseline and stage 2 exercise cardiac output (r = 0.18, P = 0.20 and r = 0.28, P = 0.05, respectively). For stage 1 exercise cardiac output, systematic bias was observed between the two methods (r = 0.40, P < 0.01).


Discussion

The results of the study demonstrate relative and absolute agreement of ES-BC, a measure of body composition, and ES Oxi, a measure of cardiac output, when compared with the standard assessments of DXA and BioZ Dx, respectively. For ES-BC, the limits of agreement were wide and systematic error was present. For ES Oxi, all three measures had narrow limits of agreement and systematic error was only present in stage 1 exercise, but not present for baseline or stage 2 exercise. Regarding autonomic nervous system activity, the results indicate marginal predictive ability of EIS-GS when regressed on HRV.

Given the enormous epidemics of metabolic syndrome and obesity, the ES-BC device makes it simple and easy for clinicians to provide their patients with information about body fat composition. Most clinicians do not have access to DXA equipment and are unfamiliar with administering a body fat assessment with a skinfold caliper. The ES-BC device provides the ability to accurately assess patients’ body fat composition. The ES-BC assessment is potentially more accurate than assessing adiposity with BMI, as BMI is not a direct measure of body fat.27,28 The ES-BC assessment may also be more accurate than other bioimpedance devices that have been previously compared to DXA and demonstrated less significant relationships among body composition variables.29

The marginal predictive capacity of EIS-GS on HRV raises a question about what EIS-GS is actually measuring. HRV is related to sympathetic system activity estimated at the level of the HRV, and EIS-GS is related to sympathetic system activity estimated at the level of the interstitial fluid. The interstitial fluid represents the main component of living tissue. The mechanisms of action of the autonomic nervous system activity on living tissue are: (1) vasoconstriction of capillaries to decrease tissue blood flow,30 (2) increased Na-K ATP pump activity,31 and (3) shifting capillary fluid to regulate blood pressure and increase the interstitial fluid volume.32 Therefore, EIS-GS could be a marker of vasoconstriction and blood flow, Na-K ATP pump activity, and interstitial fluid volume.

Additional studies have been performed to further explicate these relationships, and EIS-GS has been used to screen for prostate cancer as an adjunct to prostate specific antigen33 and as a marker of selective serotonin reuptake inhibitor treatment response.34 In these two studies, some conclusions were drawn to show that EIS-GS could detect prostate cancer and could be a marker of the response to selective serotonin reuptake inhibitor treatment through the electrochemical detection of acidic tissue or a change in the impedance in abnormal tissue. However, the detection of autonomic nervous system activity was not determined in these two studies.

Given the severity and magnitude of CVD as the number one global killer35 and hypertension as a major risk factor in the development of metabolic syndrome, ischemic heart disease, heart failure, kidney disease, and cerebrovascular disease leading to stroke,36 the ability to accurately measure the cardiac output with ES Oxi could prove to be invaluable. For example, because hypertension is a hemodynamic disorder, measuring a patient’s cardiac output before and after medical intervention may assist in the decision-making and evaluation of ongoing antihypertensive therapy,36,37 especially given that hypertension is only properly controlled in about one-third of all cases.32 Different medication classes used in the management of hypertension make it challenging at best for the clinician to determine the optimal therapy for each patient, and physician perceptions and patient symptoms negatively affect the management and control of hypertension.38,39

Cardiac output, defined as the blood volume in liters ejected by the heart per minute, is regarded as the ultimate expression of cardiovascular performance. Resting cardiac output should be 4–8 L/minute to maintain normal tissue perfusion and normal delivery of oxygen and nutrients. Cardiac output values outside this range could be a sign of CVD, hypertension, stroke, or heart failure. Hence, regular cardiac output monitoring plays an essential role in the evaluation, treatment, and follow-up of CVD. Accordingly, a noninvasive, inexpensive, safe, and fast device that can assess cardiac output and other hemodynamic parameters would be a suitable alternative to other techniques that are invasive, expensive, and risky in patients with CVD.

Ideally, a technology that measures cardiac output should be noninvasive, accurate, and reliable. At present, no single device meets these criteria. Intermittent thermodilution is widely accepted as the clinical gold standard. This method requires the insertion of a pulmonary artery catheter to obtain one measurement every 3–4 minutes.40 However, this procedure is clearly too invasive for regular monitoring and general assessment of cardiac output. Two existing, less-invasive, and continuous methods are Esophageal Doppler monitoring and carbon dioxide re-breathing, but both of these require skilled operators and expensive measurement devices.40 Among the currently-used methods, impedance cardiography is probably the only noninvasive and automatic technique. However, the impedance device is big and expensive, and its accuracy is often influenced by the change of electrode positions.41 Due to these aforementioned disadvantages, these methods are all unquestionably limited to bedside use. They are not portable or wearable, making them difficult to incorporate into home health care monitoring systems. Thus, a digital pulse oximeter with photoelectric plethysmography, such as the ES Oxi device is a preferable method to assess cardiac output.4244 The demonstrated relationship between ES Oxi and BioZ Dx on cardiac output combined with evaluating arterial stiffness, SpO2%, and blood pressure could be useful for the practice management of metabolic syndrome, hypertension, and CVD for many physicians.


Conclusion

Given that many Americans are suffering from metabolic syndrome and the related maladies of CVD, hypertension, and diabetes, the ability to effectively manage these conditions with accuracy, speed, and reliability is necessary. A noninvasive and very low-risk device that can accurately and quickly portray a patient’s current condition especially with an emphasis on body composition, autonomic nervous system activity, and cardiac output will enable the clinician to assist in planning initial and monitoring follow-up treatment. The ES-BC, EIS-GS, and ES Oxi devices managed with the ESC software demonstrated overall accuracy with a rapid assessment (measurements can be taken in 5 minutes) compared to accepted standard methods for body composition, autonomic nervous system activity, and cardiac output without reports of any adverse events. The results of the study suggest that ES-BC and ES Oxi demonstrate the ability to accurately assess body composition and cardiac output compared to standardized instruments, whereas EIS-GS showed marginal predictive ability for autonomic nervous system activity as measured by HRV and requires further testing. Thus, according to other studies,33,34 it seems that EIS-GS measurements are dependent on more than just autonomic nervous system activity.


Notes

fn1-mder-4-169Disclosure

The authors report no conflicts of interest in this work.

References
1. Lewis JE,Schneiderman N. Nutrition, physical activity, weight management, and healthRev Colomb PsiquiatrYear: 200635Suppl 1157S175S
2. Flegal KM,Carroll MD,Ogden CL,Johnson CL. Prevalence and trends in obesity among US adults, 1999–2000JAMAYear: 2002288141723172712365955
3. Mokdad AH,Bowman BA,Ford ES,Vinicor F,Marks JS,Koplan JP. The continuing epidemics of obesity and diabetes in the United StatesJAMAYear: 2001286101195120011559264
4. Ford ES,Giles WH,Dietz WH. Prevalence of the metabolic syndrome among US adults: findings from the third National Health and Nutrition Examination SurveyJAMAYear: 2002287335635911790215
5. Ford ES. Prevalence of the metabolic syndrome defined by the International Diabetes Federation among adults in the USDiabetes CareYear: 200528112745274916249550
6. Duman BS,Turkoglu C,Gunay D,Cagatay P,Demiroglu C,Buyukdevrim AS. The interrelationship between insulin secretion and action in type 2 diabetes mellitus with different degrees of obesity: evidence supporting central obesityDiabetes Nutr MetabYear: 200316424325014768774
7. Janssen I,Katzmarzyk PT,Ross R. Waist circumference and not body mass index explains obesity-related health riskAm J Clin NutrYear: 200479337938414985210
8. Romero-Corral A,Somers VK,Sierra-Johnson J,et al. Normal weight obesity: a risk factor for cardiometabolic dysregulation and cardiovascular mortalityEur Heart JYear: 201031673774619933515
9. Carnethon MR,Prineas RJ,Temprosa M,Zhang ZM,Uwaifo G,Molitch ME. The association among autonomic nervous system function, incident diabetes, and intervention arm in the Diabetes Prevention ProgramDiabetes CareYear: 200629491491916567837
10. Sanford T,Treister N,Peters C. Use of noninvasive hemodynamics in hypertension managementAm J HypertensYear: 2005182 Pt 287S91S15752938
11. Roche AF,Heymsfield SB,Lohman TGHuman body compositionChampaign, ILHuman KineticsYear: 1996
12. Macleod AF,Smith SA,Cowell T,Richardson PR,Sonksen PH. Non- cardiac autonomic tests in diabetes: use of the galvanic skin responseDiabet MedYear: 19918Spec NoS67701825962
13. Allen J. Photoplethysmography and its application in clinical physiological measurementPhysiol MeasYear: 2007283R13917322588
14. Brodie D,Moscrip V,Hutcheon R. Body composition measurement: a review of hydrodensitometry, anthropometry, and impedance methodsNutritionYear: 19981432963109583375
15. Chumlea WC,Guo SS,Kuczmarski RJ,et al. Body composition estimates from NHANES III bioelectrical impedance dataInt J Obes Relat Metab DisordYear: 200226121596160912461676
16. Rigaud B,Morucci JP,Chauveau N. Bioelectrical impedance techniques in medicine. Part I: Bioimpedance measurement. Second section: impedance spectrometryCrit Rev Biomed EngYear: 1996244–62573519196884
17. Schoeller DA. Bioelectrical impedance analysis. What does it measure?Ann N Y Acad SciYear: 200090415916210865730
18. Guo Y,Franks PW,Brookshire T,Antonio Tataranni P. The intra-and inter-instrument reliability of DXA based on ex vivo soft tissue measurementsObes ResYear: 200412121925192915687392
19. Grimnes S,Martinsen ØG. Bioimpedance and bioelectricity basicsSan Diego, CAAcademic PressYear: 2000
20. Cottrell FG. Application to the Cottrell equation to chronoamperometryZ Phys ChemYear: 190242385
21. Task Force of the European Society of Cardiology and the North American Society of Pacing and ElectrophysiologyHeart rate variability. Standards of measurement, physiological interpretation, and clinical useEur Heart JYear: 19961733543818737210
22. Sayers BM. Analysis of heart rate variabilityErgonomicsYear: 197316117324702060
23. Millasseau SC,Ritter JM,Takazawa K,Chowienczyk PJ. Contour analysis of the photoplethysmographic pulse measured at the fingerJ HypertensYear: 20062481449145616877944
24. Woltjer HH,Bogaard HJ,de Vries PM. The technique of impedance cardiographyEur Heart JYear: 1997189139614039458444
25. Bland JM,Altman DG. Statistical methods for assessing agreement between two methods of clinical measurementLancetYear: 1986184763073102868172
26. Bland JM,Altman DG. Applying the right statistics: analyses of measurement studiesUltrasound Obstet GynecolYear: 2003221859312858311
27. Romero-Corral A,Lopez-Jimenez F,Sierra-Johnson J,Somers VK. Differentiating between body fat and lean mass-how should we measure obesity?Nat Clin Pract Endocrinol MetabYear: 20084632232318382423
28. Romero-Corral A,Somers VK,Sierra-Johnson J,et al. Accuracy of body mass index in diagnosing obesity in the adult general populationInt J Obes (Lond)Year: 200832695996618283284
29. Pateyjohns IR,Brinkworth GD,Buckley JD,Noakes M,Clifton PM. Comparison of three bioelectrical impedance methods with DXA in overweight and obese menObesity (Silver Spring)Year: 200614112064207017135624
30. Valensi P,Smagghue O,Paries J,Velayoudon P,Nguyen TN,Attali JR. Peripheral vasoconstrictor responses to sympathetic activation in diabetic patients: relationship with rheological disordersMetabolismYear: 19974632352419054462
31. Bers DM,Despa S. Na/K-ATPase – an integral player in the adrenergic fight-or-flight responseTrends Cardiovasc MedYear: 200919411111819818946
32. Brown CM,Stemper B,Welsch G,Brys M,Axelrod FB,Hilz MJ. Orthostatic challenge reveals impaired vascular resistance control, but normal venous pooling and capillary filtration in familial dysautonomiaClin Sci (Lond)Year: 2003104216316912546638
33. de Abreu DS. Bioimpedance and chronoamperometry as an adjunct to prostate-specific antigen screening for prostate cancerCancer Manag ResYear: 2011310911621629833
34. Alexeev VG,Kuznecova LV. Bioimpedance in monitoring of effects of selective serotonin reuptake inhibitor treatmentPsychol Res Behav ManagYear: 20114818622114538
35. World Health OrganizationCardiovascular Diseases, 2008GenevaWorld Health OrganizationYear: 2008
36. Chobanian AV,Bakris GL,Black HR,et al. The Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure: the JNC 7 reportJAMAYear: 2003289192560257212748199
37. Lund-Johansen P. Hemodynamic effects of antihypertensive agentsDoyle AEClinical pharmacology of antihypertensive drugs (handbook of hypertension, volume II)Amsterdam; the NetherlandsElsevierYear: 19884172
38. Berlowitz DR,Ash AS,Hickey EC,et al. Inadequate management of blood pressure in a hypertensive populationN Engl J MedYear: 199833927195719639869666
39. Paramore LC,Halpern MT,Lapuerta P,et al. Impact of poorly controlled hypertension on healthcare resource utilization and costAm J Manag CareYear: 20017438939811310193
40. Mathews L,Singh RK. Cardiac output monitoringAnn Card AnaesthYear: 2008111566818182765
41. Richard R,Lonsdorfer-Wolf E,Charloux A,et al. Non-invasive cardiac output evaluation during a maximal progressive exercise test, using a new impedance cardiograph deviceEur J Appl PhysiolYear: 2001853–420220711560071
42. Chowienczyk PJ,Kelly RP,MacCallum H,et al. Photoplethysmographic assessment of pulse wave reflection: blunted response to endothelium-dependent beta2-adrenergic vasodilation in type II diabetes mellitusJ Am Coll CardiolYear: 19993472007201410588217
43. McCombie D,Asada H,Reisner A. Identification of vascular dynamics and estimation of the cardiac output waveform from wearable PPG sensorsConf Proc IEEE Eng Med Biol SocYear: 200543490349317280976
44. Wang L,Zhang YT. A novel photoplethysmogram index for total peripheral resistance after bicycle exerciseProceedings of the 5th International Conference on Ubiquitous. Healthcare2008 Oct 29–31Busan, Korea

Article Categories:
  • Original Research

Keywords: fat mass, autonomic nervous system activity, Electro Sensor Complex, dual-energy X-ray absorptiometry, heart rate variability, and bioimpedance cardiography.

Previous Document:  Critical appraisal of cardiac implantable electronic devices: complications and management.
Next Document:  Tobramycin administered by the TOBI(®) Podhaler(®) for persons with cystic fibrosis: a review.