| Comparative performance of two point-of-care analysers for lipid testing. | |
| | |
MedLine Citation:
|
PMID: 18257461 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
The aim of this study was to compare the analytical performance of the Cholestech LDX and CardioChek PA lipid point-of-care devices to a CDC-certified laboratory. Inter-assay imprecision (n=10) for blood samples from 2 patients with different lipid profiles was 3.0% for total cholesterol, 2.6% for triglyceride, 5.2% for HDL cholesterol and 6.2% for calculated LDL cholesterol on the Cholestech, and 4.4% for total cholesterol, 4.8% for triglyceride, 7.0% for HDL cholesterol and 7.4% for calculated LDL cholesterol on the Cardiochek. In a patient comparison study (n=100), correlation coefficients (r) between the POCT and laboratory methods were greater than 0,90 for all tests for the Cholestech and greater than 0.84 for all tests for the Cardiochek. The mean difference (bias) between the results obtained on the Cholestech LDX and the laboratory method was not statistically significant; however the mean difference between the CardioChek and the laboratory method was statistically significant for total, HDL and LDL cholesterol (one way analysis of variance with Scheffe post-hoc test). The Cholestech LDX met the NCEP goal for total error for all analytes except LDL cholesterol. The CardioChek PA system met the NCEP total error goal for triglyceride but not the other lipid analytes. We conclude that the Cholestech LDX device is a suitable POCT device for cardiovascular risk assessment in the primary care setting, while the CardioChek device requires more study and refinement. |
| | |
Authors:
|
Mark D S Shephard; Beryl C Mazzachi; Anne K Shephard |
Publication Detail:
|
Type: Comparative Study; Evaluation Studies; Journal Article |
Journal Detail:
|
Title: Clinical laboratory Volume: 53 ISSN: 1433-6510 ISO Abbreviation: Clin. Lab. Publication Date: 2007 |
Date Detail:
|
Created Date: 2008-02-08 Completed Date: 2008-03-04 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 9705611 Medline TA: Clin Lab Country: Germany |
Other Details:
|
Languages: eng Pagination: 561-6 Citation Subset: IM |
Affiliation:
|
Community Point-of-Care Services, Flinders University Rural Clinical School, Flinders University, Bedford Park, Adelaide, South Australia, Australia. Mark.Shephard@flinders.edu.au |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Blood Chemical Analysis
/
instrumentation* Cholesterol, HDL / blood* Cholesterol, LDL / analysis*, blood Humans Point-of-Care Systems* Reproducibility of Results Sensitivity and Specificity Triglycerides / analysis*, blood |
| Chemical | |
Reg. No./Substance:
|
0/Cholesterol, HDL; 0/Cholesterol, LDL; 0/Triglycerides |