OBJECTIVES: The objectives of this study were to compare numerically the bending and torsional mechanical behavior of 5 endodontic rotary Ni-Ti instruments with equivalent size and various designs for tapers, pitch, and cutting blades.First, the geometries of Hero (20/0.06), HeroShaper (20/0.06), ProFile (20/0.06), Mtwo (20/0.06), and ProTaper F1 were generated by finite element code. Then, the 2 most representative clinical loadings, i.e., bending and torsion, were studied with an ad hoc model for the superelasticity of Ni-Ti. Bending was generated by tip deflection and torsion by a constant twist-angle of the tip.
RESULTS: Mechanical behavior of these 5 endodontic rotary Ni-Ti instruments could be evaluated and compared. Protaper F1 presented the greatest level of bending stress and torque. Hero and HeroShaper were more rigid than ProFile and Mtwo.
CONCLUSIONS: This numerical comparison evaluated the effects of the geometrical parameters on the instrumental mechanical behavior. The 5 endodontic instruments, investigated in the present study, do not have the same bending and torsional mechanical behavior. Each clinician must be aware of these behavior differences so as to use the adequate file according to the clinical situation and to the manufacturer's recommendations.
Authors:
Reza Arbab-Chirani; Valérie Chevalier; Shabnam Arbab-Chirani; Sylvain Calloch
Related Documents :
21986816 - Atoms-to-microns model for small solute transport through sticky nanochannels.
21765496 - Single-etch grating coupler for micrometric silicon rib waveguides.
9712576 - Moho offset across the northern margin of the tibetan plateau
22639226 - Complex network analysis in microbial systems: theory and examples.
15625816 - Deposition of coarse particles in cystic fibrosis: model predictions versus experimenta...
18259576 - Multiangle imaging spectroradiometer: optical characterization of the calibration panels.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics     Volume:  111     ISSN:  1528-395X     ISO Abbreviation:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod     Publication Date:  2011 Jan 
Date Detail:
Created Date:  2010-12-23     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9508562     Medline TA:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod     Country:  United States    
Other Details:
Languages:  eng     Pagination:  115-21     Citation Subset:  D; IM    
Copyright Information:
Copyright © 2011 Mosby, Inc. All rights reserved.
Affiliation:
Centre Hospitalier Universitaire-Service d'Odontologie/Université européenne de Bretagne-Faculté d'Odontologie, Brest, France. arbab@univ-brest.fr
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effect of low-surface-tension EDTA solutions on the wettability of root canal dentin.
Next Document:  Effect of selected accelerants on setting time and biocompatibility of mineral trioxide aggregate (M...