Color and conventional image-directed Doppler ultrasonography: accuracy and sources of error in quantitative blood flow measurements.
MedLine Citation:
PMID:  1313832     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Accuracy of two systems--conventional (DRF 400, Diasonics) and color-coded (Angiodynograph, Quantum/Phillips) image-directed Doppler ultrasonography--was investigated using an in vitro model that generated both monophasic and triphasic pulsatile flow patterns. Estimated and actual blood volume flow rates showed good correlations, but the sampling with a hand-held transducer led to wide variations in measurement error for the conventional (-69.2% to 50%) and the color-coded (-79.3% to 265.7%) systems. By performing multiple measurements, one could improve accuracy considering only the maximal values of a series instead of the mean values. Accuracy was impaired by interposed muscular or fatty tissue due to false low time-average velocity measurements caused by a loss of Doppler signal. Comparison of both systems revealed significant differences between pulsatility index values (p less than 0.001), blood flow velocities (p less than 0.001), and blood volume flow rates (p less than 0.05 for program flow, p less than 0.001 for manual and automatic flow program of the color-coded system).
Authors:
C Ranke; P Hendrickx; U Roth; F Brassel; A Creutzig; K Alexander
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Journal of clinical ultrasound : JCU     Volume:  20     ISSN:  0091-2751     ISO Abbreviation:  J Clin Ultrasound     Publication Date:    1992 Mar-Apr
Date Detail:
Created Date:  1992-05-14     Completed Date:  1992-05-14     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  0401663     Medline TA:  J Clin Ultrasound     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  187-93     Citation Subset:  IM    
Affiliation:
Department of Angiology, Medical School, Hannover, West Germany.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Flow Velocity*
Humans
Models, Anatomic
Pulsatile Flow
Reproducibility of Results
Ultrasonography / methods*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  An assessment of hysterosalpingosonography (HSSG) as a diagnostic tool for uterine cavity defects an...
Next Document:  Efficiency of cleaning agents for an inorganic membrane after milk ultrafiltration.