Clinical testing of cellular phone ringing interference with automated external defibrillators.
MedLine Citation:
PMID:  16987586     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: This study examined cellular phone ringing interference with automated external defibrillators (AED). METHODS: The phone systems tested were two single band handheld telephones: (1) a Global System for Mobile Communication (GSM) receiver; and (2) a Personal Communication Services (PCS) receiver. The ringing phase of a digital cellular phone includes a brief burst of peak-emitted power. The GSM had a maximum power output of 2 W, operating on a 900 MHz carrier frequency, and the PCS had a maximum output of 1 W, operating on a 1800 MHz carrier frequency. During AED monitoring, the digital cellular telephone was placed successively in three positions: (1) on the AED; (2) on the patient's chest between the electrodes; and (3) on the connector between the electrodes and the AED cable. After positioning the cellular phone, calls were placed during the AED analyzing phase. RESULTS: Three AED models were tested using their original electrodes: (1) LifePak 20 monitor/defibrillator device; (2) Lifepak 20 P monitor/defibrillator/stimulator (Medtronic Emergency Response Systems, Redmond, WA, USA); and (3) HeartStart XL M4735A monitor/defibrillator (Philips Medical Systems, Andover, MA, USA). The first two devices had Quik-Combo electrodes and the third device had Adults Plus multifunction electrodes. Ninety-one tests were performed on 13 patients. The only disturbance provoked by testing was noise emitted by the AED speaker when the receiver was close to the device. The noise began 2-4 s before the first audible ringing tone and persisted throughout the ringing phase. The distance at which this effect could be prevented was 15 cm. CONCLUSION: Clinical testing during ECG monitoring by an AED during call from a cellular phone did not show any analysis dysfunction during unshockable rhythms and provoked only transient dysfunction of the speaker device.
Authors:
Alexandre Trigano; Olivier Blandeau; Christian Dale; Man-Faï Wong; Joe Wiart
Related Documents :
8720056 - A canine case of discoid lupus erythematosus with circulating autoantibody.
17241836 - Copper t 380a iud and magnetic resonance imaging.
6610586 - Simple test system for interleukin-2.
20543276 - Accessible virtual reality therapy using portable media devices.
20828856 - Illusions of scientific legitimacy: misrepresented science in the direct-to-consumer ge...
20740266 - Self-reported symptoms and neuropsychological function among tunnel workers previously ...
Publication Detail:
Type:  Comparative Study; Journal Article     Date:  2006-09-20
Journal Detail:
Title:  Resuscitation     Volume:  71     ISSN:  0300-9572     ISO Abbreviation:  Resuscitation     Publication Date:  2006 Dec 
Date Detail:
Created Date:  2006-11-29     Completed Date:  2007-02-23     Revised Date:  2009-08-25    
Medline Journal Info:
Nlm Unique ID:  0332173     Medline TA:  Resuscitation     Country:  Ireland    
Other Details:
Languages:  eng     Pagination:  391-4     Citation Subset:  IM    
Affiliation:
Department of Cardiology, Assistance Publique-Hôpitaux de Marseille, Hôpital Nord, 13915 Marseille, France. alexandre.trigano@mail.ap-hm.fr
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Aged, 80 and over
Cellular Phone*
Defibrillators*
Electrocardiography
Electromagnetic Fields / adverse effects*
Equipment Design
Equipment Safety
Female
Humans
Male
Middle Aged
Time Factors

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Outcome of cardiopulmonary resuscitation in a 2300-bed hospital in a developing country.
Next Document:  Comparison of three simulation-based training methods for management of medical emergencies.