Clinical oxygen transfer comparison of the Terumo Capiox SX18 and SX25 membrane oxygenators.
MedLine Citation:
PMID:  9533118     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The purpose of this current study was to compare the Terumo Capiox SX18 (1.8 m2) with the recently released Capiox SX25 (2.5 m2). Specifically, their oxygen transfer slopes, degree of blood shunting, extrapolated maximum oxygen transfer, blood side pressure drop and oxygen transfer consistency were compared. The lower intercept value (0.209 vs 0.236) coupled with the flatter slope (0.00171 vs 0.00225) of the oxygen transfer line for the SX25 is consistent with improved oxygen transfer performance. A lower FiO2 value would be predicted for the SX25, to achieve a specific PaO2 value, when the oxygen transfer requirement is equal for both devices. Extrapolated maximum oxygen transfer for the SX25 (462.6 ml O2/min) was 36.2% higher than that for the SX18 (339.6 ml O2/min). When indexed to membrane surface area, the SX18 (188.7 ml O2/m2/min performed 2.0% better than the SX25 (185.0 ml O2/m2/min). Both the slope (3.110 vs 3.744) and the intercept (4.595 vs 6.223) of the shunt fraction line were lower for the SX25, indicating that a lower shunt fraction would be predicted for all clinical blood flow rates. The slope (23.934 vs 22.443) and intercept values (-28.388 vs -22.650) of the pressure drop lines for the two devices indicate that the blood side pressure drop, over the range of clinical blood flows, were within 2% of each other. Oxygen transfer consistency, when expressed as the standard deviation of the oxygenator performance index and the percentage of predicted shunt, was not statistically different for the two devices.
Authors:
D W Fried; T L Zombolas; J J Leo; H Mohamed; G J Mattioni
Related Documents :
7713798 - Excess oxygen delivery during muscle contractions in spontaneously hypertensive rats.
10315398 - Nonmedical hazards of long-term oxygen therapy.
16384838 - Oxygen release from arterioles with normal flow and no-flow conditions.
11292898 - Effects of hot shot on recovery after hypothermic ischemia in neonatal lamb heart.
8419708 - Warm versus cold blood cardioplegia--is there a difference?
7556298 - Estimation of glomerular filtration rate using chromium-51 ethylene diamine tetra-aceti...
Publication Detail:
Type:  Clinical Trial; Comparative Study; Controlled Clinical Trial; Journal Article    
Journal Detail:
Title:  Perfusion     Volume:  13     ISSN:  0267-6591     ISO Abbreviation:  Perfusion     Publication Date:  1998 Mar 
Date Detail:
Created Date:  1998-05-27     Completed Date:  1998-05-27     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  8700166     Medline TA:  Perfusion     Country:  ENGLAND    
Other Details:
Languages:  eng     Pagination:  119-27     Citation Subset:  IM    
Affiliation:
Albert Einstein Medical Center, Philadelphia, Pennsylvania, USA. DWF1@erols.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Blood Flow Velocity
Humans
Linear Models
Oxygen / blood*
Oxygenators, Membrane*
Pressure
Reproducibility of Results
Chemical
Reg. No./Substance:
7782-44-7/Oxygen

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Foetal bypass: concepts and controversies.
Next Document:  Effects of flow types in cardiopulmonary bypass on gastric intramucosal pH.