Document Detail


Clinical evaluation of morphology discrimination: an algorithm for rhythm discrimination in cardioverter defibrillators.
MedLine Citation:
PMID:  11449597     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The aim of this study was to test the new morphology discrimination diagnostic algorithm for ICDs that differentiates supraventricular tachycardias (SVTs) from VTs by analysis of ventricular depolarization complexes morphology. Twenty-five patients implanted with a St. Jude Ventritex single chamber ICD were studied during electrophysiological evaluation at predischarge and were followed for 7 +/- 4 months. Sensitivity and specificity for VT detection and overall diagnostic accuracy of the morphology discrimination algorithm were calculated on 326 detected events. At electrophysiological evaluation, the algorithm was tested during 67 episodes of right atrial pacing, during 119 episodes of RV pacing (at basal interventricular septum and RV apex) and during 27 episodes of sustained AF: specificity was 98%, sensitivity was 66%, and diagnostic accuracy was 80%. All episodes of AF were correctly diagnosed as SVT. Exclusion of detections related to pacing at the basal interventricular septum, resulted in a specificity of 98%, a sensitivity of 85%, and a diagnostic accuracy of 93%. During follow-up, evaluation of the morphology discrimination algorithm on 113 spontaneous episodes (31 VTs, 31 AF, 7 SVTs, and 44 sinus tachycardias) exhibited a specificity of 89%, a sensitivity of 100%, and a diagnostic accuracy of 92%. In conclusion, the morphology discrimination algorithm exhibits a high specificity in discriminating VTs from SVTs, although with a corresponding reduction in sensitivity. The preliminary experience on spontaneous episodes is promising. To correct for the reduction in sensitivity, it is advisable to use this algorithm in parallel with other algorithms for rhythm discrimination (sudden onset, stability) coupled with extended high rate.
Authors:
G Boriani; M Biffi; L Frabetti; J J Lattuca; A Branzi
Related Documents :
17592407 - Diagnostic tests in wrist instability.
8609687 - The impact of diagnostic testing on therapeutic interventions.
8316417 - Contemporary algorithms for evaluation of the painful total hip replacement.
2781687 - Corticosteroid-induced alkaline phosphatase isoenzyme in the diagnosis of canine hyperc...
3041957 - Evaluation of the specificity and sensitivity of two commercial enzyme-linked immunosor...
25116047 - The impact of a total hip replacement on jaw position, upper body posture and body sway.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  24     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  2001 Jun 
Date Detail:
Created Date:  2001-07-13     Completed Date:  2002-05-14     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  994-1001     Citation Subset:  IM    
Affiliation:
Institute of Cardiology, University of Bologna, Italy. cardio1@almadns.unibo.it
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Algorithms*
Defibrillators, Implantable*
Female
Follow-Up Studies
Humans
Male
Middle Aged
Patient Discharge
Reproducibility of Results
Sensitivity and Specificity
Tachycardia / diagnosis*,  physiopathology,  therapy*