Clinical assessment of the DINAMAP ProCare monitor in an adult population according to the British Hypertension Society Protocol.
MedLine Citation:
PMID:  17303988     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: Clinical recommendation of an automated blood pressure device is dependent on assessment according to a recognized protocol. The DINAMAP ProCare monitor is an oscillometric device intended for hospital, physician office or clinic use. The device algorithm was developed and tested against an auscultatory standard to increase comparability, in contrast to previous DINAMAP products that were designed to be accurate compared with an invasive central aortic reference, as specified in the Association for the Advancement of Medical Instrumentation-SP10 Standard (AAMI SP10 Standard: 1992, 2002). This device was validated in an adult population according to the protocol of the British Hypertension Society. METHODS: Local ethics committee approval was obtained and all patients were required to give written informed consent. Nine sequential same-arm measurements, alternating between observer and device, were taken from each patient. Data analysis was performed according to the guidelines of the British Hypertension Society protocol on 85 patients who fulfilled the blood pressure categories specified by the protocol. RESULTS: The DINAMAP ProCare monitor achieved an A grade for both systolic and diastolic pressures. It also achieved the passing criteria of the ANSI/AAMI SP10 with a mean difference (standard deviation) for systolic and diastolic pressures, respectively, of -0.3 (6.9) and -4.0 (5.9) mmHg. CONCLUSION: The DINAMAP ProCare monitor can be recommended for clinical use in an adult population according to the criteria of the British Hypertension Society.
Authors:
Annemarie de Greeff; Fayrouge Reggiori; Andrew H Shennan
Related Documents :
20113128 - A new streaming potential chamber for zeta potential measurements of particulates.
2589678 - Experimental analysis of catheter-manometer systems in vitro and in vivo.
21091358 - Validation of the fm-800 ambulatory blood pressure monitor according to the association...
20961548 - Novel reduced pressure-balance syringe for chromatographic analysis.
1893928 - Cardiovascular profile of the new dihydropyridine derivative s12968.
3822598 - The differential effects of leukotriene c4 and d4 on the pulmonary and systemic circula...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't; Validation Studies    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  12     ISSN:  1359-5237     ISO Abbreviation:  Blood Press Monit     Publication Date:  2007 Feb 
Date Detail:
Created Date:  2007-02-16     Completed Date:  2007-04-27     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  51-5     Citation Subset:  IM    
Affiliation:
Maternal and Fetal Research Unit, King's College London School of Medicine, St Thomas' Hospital, London, UK.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Aged, 80 and over
Automation / instrumentation,  standards
Blood Pressure Determination / instrumentation*,  standards
Diastole
Female
Humans
Hypertension / diagnosis
Male
Middle Aged
Oscillometry / instrumentation*,  standards
Outpatient Clinics, Hospital
Reproducibility of Results
Systole

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Validation of the Braun BP 3550 wrist blood pressure measuring device with a position sensor and an ...
Next Document:  Giant cell arteritis: diagnosis, management, and cardiovascular implications.