| Clinical application of UF-1000i in combination with AX-4280 for the screening test ability of urinary formed elements. | |
| | |
MedLine Citation:
|
PMID: 23135347 Owner: NLM Status: Publisher |
Abstract/OtherAbstract:
|
AIMS: To evaluate the screening test ability of UF-1000i combination with AX-4280 for major pathological urinary formed elements and to establish screening rules suitable for our laboratory. METHODS: A total of 2220 cases of urine samples were collected from out-patients and inpatients. The sample tests were finished within 4 h after collection. The results were verified and corroborated with microscopic tests. RESULTS: In 'both-negative' group, the negative predictive values for red blood cell, white blood cell and cast were 99.12%, 99.56% and 100%, respectively, and the false negative rates were 0.87%, 0.44% and 0%, respectively. In 'both-positive' group, the positive predictive values for red blood cell, white blood cell and cast were 69.18%, 74.41% and 33.33%, respectively, and the false positive rates were 30.82%, 25.59% and 66.67%, respectively. In 'U positive/A negative' group, the positive predictive values for red blood cell, white blood cell and cast were 32.47%, 23.67% and 8.57%, respectively, and the false positive rates were 67.53%, 76.36% and 91.43%, respectively. In 'U negative/A positive' group, the positive predictive values for red blood cell, white blood cell and cast were 8.18%, 19.50% and 4.67%, respectively, and the false positive rates were 91.82%, 50.50% and 95.24%, respectively. CONCLUSIONS: When the results of the red blood cell, white blood cell and cast tested by UF-1000i were all negative, and the results of the red blood cell, white blood cell and protein tested by AX-4280 were negative, an automatic report stating 'normal microscopic appearance' was generated, and then the microscopic test rate was reduced by 37.6%. However, if any automated parameter measurement was out of the reference range, corroboration by manual microscopy was advisable. |
| | |
Authors:
|
Feng Zhao; Yi Jin; Xiaoyan Chen; Xinyou Xie |
Related Documents
:
|
16325447 - Abo incompatible transfusions--experience from the uk serious hazards of transfusion (s... 10972917 - Determination of red cell volume in infants needing blood transfusion. 8751307 - Post-transfusion hyperhaemolysis in a patient with sickle cell disease: use of steroids... 5460077 - Plasticizers from plastic devices extraction, metabolism, and accumulation by biologica... 7032277 - Distribution of blood group substances in normal human colon. use of the unlabeled anti... 19964497 - Digital thermal monitoring (dtm) of vascular reactivity closely correlates with doppler... |
Publication Detail:
|
Type: JOURNAL ARTICLE Date: 2012-11-7 |
Journal Detail:
|
Title: Journal of clinical pathology Volume: - ISSN: 1472-4146 ISO Abbreviation: J. Clin. Pathol. Publication Date: 2012 Nov |
Date Detail:
|
Created Date: 2012-11-8 Completed Date: - Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 0376601 Medline TA: J Clin Pathol Country: - |
Other Details:
|
Languages: ENG Pagination: - Citation Subset: - |
Affiliation:
|
Department of Clinical Laboratory, Sir Run Run Shaw Hospital, College of Medicine, Zhejiang University, Hangzhou, Zhejiang, People's Republic of China. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
|
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: An efficient organocatalytic enantioselective synthesis of spironitrocyclopropanes.
Next Document: Genetic and epigenetic markers in the evaluation of pancreatic masses.