Clinical Utility of Combined Placido-Scanning-Slit Midperipheral and Thinnest Point Pachymetry After Corneal Ablation for Myopia.
MedLine Citation:
PMID:  24473224     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
PURPOSE:: The aim of this study was to assess the random error of the thinnest point and midperipheral pachymetry with a combined Placido-scanning-slit system (Orbscan II) in the intermediate-term follow-up of excimer laser keratorefractive surgery for treating myopia.
METHODS:: Sixty-five patients who had undergone aspheric surface ablation for treating myopia with a minimum follow-up of 6 months, and showed no biomicroscopically detectable corneal haze, were subjected to 5 consecutive topographic examinations. All eyes underwent a complete ophthalmic examination before the surgery. The within-subject SD (Sw), repeatability (2.77 × Sw), coefficient of repeatability (2.77 × Sw/mean), and intraclass correlation coefficients of pachymetry at the thinnest point and in each 4 quadrants at a 6-mm diameter were calculated.
RESULTS:: The patients had a mean age of 31.9 ± 6.2 years, and the average follow-up was for 7.5 ± 2.6 months. The mean ablation depth for myopia was 63.6 ± 23.6 μm. The average pachymetry for the thinnest point and midperipheral superior, inferior, nasal, and temporal locations was 456.9, 602.0, 595.5, 609.7, and 566.1 μm, respectively. The repeatability (and coefficient of repeatability) for the thinnest point and midperipheral superior, inferior, nasal, and temporal locations was 26.5 (5.8%), 37.9 (6.3%), 31.0 (5.2%), 30.5 (5.0%), and 35.4 μm (6.2%), respectively. The intraclass correlation coefficients were 0.96, 0.88, 0.89, 0.97, and 0.91 for each location, respectively.
CONCLUSIONS:: This study provides the repeatability of the thinnest and midperipheral pachymetry using the combined Placido-scanning-slit system in transparent corneas after laser keratorefractive surgery for treatment of myopia. The test-retest reliability provided here will help differentiate real corneal thickness change from measurement noise. For this, only pachymetric changes >6% are likely to be real and therefore useful when evaluating postoperative keratectasia suspects.
Authors:
Alberto López-Miguel; Loreto Martínez-Almeída; María B Coco-Martín; María Del Val; María Pichel; Miguel J Maldonado
Related Documents :
18183364 - What type of monitoring has been shown to improve outcomes in acutely ill patients?
19818574 - Continuous positive airway pressure vs. pressure support ventilation in acute cardiogen...
23878784 - Ponseti method compared with soft-tissue release for the management of clubfoot: a meta...
15017544 - Impact of a current treatment protocol on outcome of high-risk congenital diaphragmatic...
20595544 - Fibular collateral ligament anatomical reconstructions: a prospective outcomes study.
6641054 - Thirty-two noninfected total hip arthroplasties revised due to stem loosening.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2014-1-27
Journal Detail:
Title:  Cornea     Volume:  -     ISSN:  1536-4798     ISO Abbreviation:  Cornea     Publication Date:  2014 Jan 
Date Detail:
Created Date:  2014-1-29     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8216186     Medline TA:  Cornea     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Late-Onset Lattice Corneal Dystrophy Without Typical Lattice Lines Caused by a Novel Mutation in the...
Next Document:  Use of Conjoint Analysis to Assess Breast Cancer Patient Preferences for Chemotherapy Side Effects.