Document Detail


Chronic measurement of epidural pressure with an induction-powered oscillator transducer.
MedLine Citation:
PMID:  1255235     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
An induction-powered oscillator transducer (IPOT) was designed for the chronic measurement of epidural pressure. The transducer was completely implantable so all pressure measurements were made through the intact skin. The IPOT had a linear pressure range from -50 to +200 cm H2O, was sensitive to 1 mm H2O and had a zero drift of less than 1 mm H2O/day under full load. Zero drift was minimized by using a hermetically-sealed metal bellows transducing element which was chemically treated to prevent corrosion and creep. The correlation between epidural pressure and intraventricular pressure was determined during the first 24 hours after implantation in six dogs. Epidural pressure was found to be a linear function of intraventricular fluid pressure. Epidural pressure and intraventricular pressure were essential equal provided the epidural wedge pressure was minimized by proper insertion of the transducer. The correlation between epidural pressure and intraventricular pressure was determined after chronic implantation in five dogs. Epidural pressure was a linear function of intraventricular pressure in the chronically implanted dogs, but epidural pressure was not equal to intraventricular pressure. After chronic implantation, the epidural pressure transducer was not responsive to changes in intraventricular pressure because of mechanical changes in the dura. The dura became stiff and non-compliant. Maximum correlation between epidural pressure and intraventricular fluid pressure in chronic implantations will depend on judicious material selection and mechanical design at the transducer-dura interface.
Authors:
H G Rylander; H L Taylor; J P Wissinger; J L Story
Related Documents :
7726445 - Histamine concentrations and hemodynamic responses after remifentanil.
2884895 - The effect of sufentanil on the cardiovascular responses to tracheal intubation.
1642945 - Esmolol hydrochloride for management of the cardiovascular stress responses to laryngos...
7798415 - Oral transmucosal fentanyl citrate premedication in patients undergoing outpatient derm...
17303985 - Validation of heine gamma g7 (g5) and xxl-lf aneroid devices for blood pressure measure...
10379495 - Venous doppler velocimetry in relationship to central venous pressure and heart rate du...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of neurosurgery     Volume:  44     ISSN:  0022-3085     ISO Abbreviation:  J. Neurosurg.     Publication Date:  1976 Apr 
Date Detail:
Created Date:  1976-05-25     Completed Date:  1976-05-25     Revised Date:  2003-11-14    
Medline Journal Info:
Nlm Unique ID:  0253357     Medline TA:  J Neurosurg     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  465-78     Citation Subset:  AIM; IM; S    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Cerebral Ventricles / physiology
Dogs
Electrodes, Implanted
Electronics, Medical*
Intracranial Pressure*
Monitoring, Physiologic / instrumentation*
Subarachnoid Space / physiology
Transducers

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Classification of cervical spondylosis or disc protrusion by preoperative evoked spinal electrogram....
Next Document:  Use of epsilon aminocaproic acid (EACA) in the preoperative management of ruptured intracranial aneu...