Catheter obstruction effect on pulsatile flow rate--pressure drop during coronary angioplasty.
MedLine Citation:
PMID:  10396693     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The coupling of computational hemodynamics to measured translesional mean pressure gradients with an angioplasty catheter in human coronary stenoses was evaluated. A narrowed flow cross section with the catheter present effectively introduced a tighter stenosis than the enlarged residual stenoses after balloon angioplasty; thus elevating the pressure gradient and reducing blood flow during the measurements. For resting conditions with the catheter present, flow was believed to be about 40 percent of normal basal flow in the absence of the catheter, and for hyperemia, about 20 percent of elevated flow in the patient group. The computations indicated that the velocity field was viscous dominated and quasi-steady with negligible phase lag in the delta p(t)-u(t) relation during the cardiac cycle at the lower hydraulic Reynolds numbers and frequency parameter. Hemodynamic interactions with smaller catheter-based pressure sensors evolving in clinical use require subsequent study since artifactually elevated translesional pressure gradients can occur during measurements with current angioplasty catheters.
Authors:
R K Banerjee; L H Back; M R Back; Y I Cho
Related Documents :
1175183 - Myocardial function in general and regional left ventricular ischaemia in dogs at contr...
3717043 - Acute effect of systemic versus intracoronary dipyridamole on coronary circulation.
1175183 - Myocardial function in general and regional left ventricular ischaemia in dogs at contr...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of biomechanical engineering     Volume:  121     ISSN:  0148-0731     ISO Abbreviation:  J Biomech Eng     Publication Date:  1999 Jun 
Date Detail:
Created Date:  1999-09-28     Completed Date:  1999-09-28     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  7909584     Medline TA:  J Biomech Eng     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  281-9     Citation Subset:  IM    
Affiliation:
Fluent, Inc., Lebanon, NH 03766, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Algorithms
Angioplasty, Transluminal, Percutaneous Coronary / adverse effects*
Blood Flow Velocity*
Blood Pressure / physiology*
Humans
Models, Cardiovascular
Pulsatile Flow / physiology*
Stress, Mechanical
Time Factors

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A numerical study of blood flow patterns in anatomically realistic and simplified end-to-side anasto...
Next Document:  An in vitro study of implant--tooth-supported connections using a robot test system.