Cardiovascular catheters and accessories: ex vivo testing of ferromagnetism, heating, and artifacts associated with MRI.
MedLine Citation:
PMID:  9848749     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The purpose of this study was to evaluate the MR safety of cardiovascular catheters and accessories. Intravascular cardiovascular catheters and accessories were tested for MR safety at 1.5 T using previously described techniques with respect to the evaluation of magnetic field attraction (deflection angle method), heating (temperature measured immediately before and after performing MRI), and artifacts (using a fast spoiled gradient-recalled acquisition in steady state [GRASS] pulse sequence). Two devices were attracted (RV pacing lead and Oximetrix 3 SO2 optical module) by the static magnetic field. Each of the other objects displayed no attraction. Heating was +0.2 degrees C for the sample cardiovascular catheter tested (Opticath). Artifacts varied from moderate to severe, depending on the amount and type of metal present in the device. Despite these ex vivo test results, further safety consideration should be given to the cardiovascular devices that have a conductive wire component (ie, certain types of the cardiovascular catheters) because of the potential for inducing current and excessive heating in these devices during MRI, especially using a high-field-strength MR system. The cardiovascular catheters evaluated in this study or those with a similar design are not recommended for use in patients undergoing MRI procedures.
Authors:
F G Shellock; V J Shellock
Related Documents :
20036589 - On the applications of micror=-1 metamaterial lenses for magnetic resonance imaging.
21256259 - Imaging and surgical repair of a bifurcating aortico-left ventricle tunnel.
17795619 - Environmental applications of magnetic measurements.
632049 - Tomographic imaging with nuclear magnetic resonance.
10221909 - Magnetic lineations in the ancient crust of mars
4001399 - Magnetic resonance imaging performance: a comparison of sodium and hydrogen.
1111009 - Effect of collimator motion on image quality in nuclear medicine.
21570209 - On the benefits and risks of proton therapy in pediatric craniopharyngioma.
20479449 - Osteochondral lesions of the talus: comparison of three-dimensional fat-suppressed fast...
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Journal of magnetic resonance imaging : JMRI     Volume:  8     ISSN:  1053-1807     ISO Abbreviation:  J Magn Reson Imaging     Publication Date:    1998 Nov-Dec
Date Detail:
Created Date:  1999-02-09     Completed Date:  1999-02-09     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9105850     Medline TA:  J Magn Reson Imaging     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  1338-42     Citation Subset:  IM    
Affiliation:
University of Southern California, School of Medicine, Los Angeles, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Artifacts*
Heart Catheterization* / adverse effects,  instrumentation
Heating / adverse effects*
Humans
Magnetic Resonance Imaging / adverse effects,  instrumentation*
Magnetics / adverse effects*
Phantoms, Imaging
Safety

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Black-blood MR angiography with GRASE: measurement of flow-induced signal attenuation.
Next Document:  A compact respiratory-triggering device for routine microimaging of laboratory mice.