Document Detail


Cardiac output by PulseCO is not interchangeable with thermodilution in patients undergoing OPCAB.
MedLine Citation:
PMID:  15872133     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: To investigate the reliability of cardiac output assessed by arterial pressure waveform (PulseCO) in comparison with bolus thermodilution measurements in patients undergoing off-pump coronary artery bypass grafting (OPCAB). METHODS: 23 patients who underwent OPCAB were enrolled in this study. After premedication with oral diazepam 10 mg, anesthesia was induced with midazolam, fentanyl and vecuronium. After induction, radial artery and pulmonary artery catheters were inserted. Cardiac output was measured simultaneously by the PulseCO and the bolus thermodilution method using the Vigilance monitor: 1) after sternotomy, 2) after opening the mediastinum, and 3) at the end of surgery. The PulseCO was calibrated initially with cardiac output determined by the thermodilution method after induction of anesthesia. RESULTS: The correlation coefficients between the two techniques at the three measurement periods were: 1) R2 = 0.49, 2) R2 = 0.52, 3) R2 = 0.55. The limits of agreement (bias +/- 2 SD of bias) were: 1) 0.71 +/- 2.66, 2) 0.30 +/- 1.97, 3) 0.76 +/- 3.85 L.min(-1). CONCLUSIONS: Cardiac output by PulseCO is not interchangeable with cardiac output measured by thermodilution in patients undergoing OPCAB.
Authors:
Koichi Yamashita; Tomoki Nishiyama; Takeshi Yokoyama; Hidehiro Abe; Masanobu Manabe
Related Documents :
3612293 - First-pass radionuclide determination of cardiac output: an improved gamma camera method.
17499663 - Utility of impedance cardiography for dyspneic patients in the ed.
20833143 - Myocardial regeneration potential of adipose tissue-derived stem cells.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Canadian journal of anaesthesia = Journal canadien d'anesthésie     Volume:  52     ISSN:  0832-610X     ISO Abbreviation:  Can J Anaesth     Publication Date:  2005 May 
Date Detail:
Created Date:  2005-05-05     Completed Date:  2005-09-08     Revised Date:  2006-12-01    
Medline Journal Info:
Nlm Unique ID:  8701709     Medline TA:  Can J Anaesth     Country:  Canada    
Other Details:
Languages:  eng     Pagination:  530-4     Citation Subset:  IM    
Affiliation:
Department of Anesthesiology and Critical Care Medicine, Kochi Medical School, Kohasu, Oko-cho, Nankoku-shi, Kochi, 783-8505, Japan. koichiya@kochi-ms.ac.jp
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Blood Pressure
Cardiac Output*
Coronary Artery Bypass*
Female
Humans
Male
Middle Aged
Thermodilution*
Comments/Corrections
Comment In:
Can J Anaesth. 2006 Aug;53(8):850-1   [PMID:  16873359 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine