Abstract/OtherAbstract:
The sidestream capnogram is delayed behind real time by transport delay (TD; time to aspirate gas through the sampling tubing) and by the dynamic response (DR) of the measurement cuvette. In six capnometers, we measured TD and DR by plunging the end of the sample tubing into a flask containing CO2 and then digitally analyzing the capnogram. TD ranged from 0.6 to 5.0 s and accounted for 89% or more of the total response time (TD + DR) of the capnometer. TD was generally not reported in the manufacturers' specifications. TD was further prolonged by low aspiration rates or by sampling tube extensions. During a series of quick breaths after endotracheal intubation, long TD can delay the appearance of CO2 and result in a false diagnosis of esophageal intubation. Also, long TD can prolong DR, which can result in underestimation of end-tidal PCO2 during rapid ventilation in pediatric anesthesia.
Authors:
P H Breen; B Mazumdar; S C Skinner
Related Documents :
6443089 - Does geographical location influence inflow requirements of the bain breathing system?
3924099 - Evidence for hco3- conductance pathways in nutrient membrane of bullfrog antrum.
9741659 - Effect of exercise on acid-base status and ventilatory kinetics.
9008979 - Contribution of peripheral chemoreceptor drive in exercise hyperpnea in humans.
22648139 - Lactate response to different volume patterns of power clean.
3103969 - Research utility of noninvasive methods for measurement of cardiac output.
8520769 - The impact of exercise reconditioning on breathlessness in severe chronic airflow limit...
3646619 - Predictors of dyspnea intensity in asthma.
1753729 - Former athletes physical fitness.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't; Research Support, U.S. Gov't, P.H.S.    
Journal Detail:
Title:  Anesthesia and analgesia     Volume:  78     ISSN:  0003-2999     ISO Abbreviation:  Anesth. Analg.     Publication Date:  1994 Mar 
Date Detail:
Created Date:  1994-03-24     Completed Date:  1994-03-24     Revised Date:  2007-11-14    
Medline Journal Info:
Nlm Unique ID:  1310650     Medline TA:  Anesth Analg     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  584-6     Citation Subset:  AIM; IM    
Affiliation:
Department of Anesthesia and Critical Care, University of Chicago, IL 60637.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Gas Monitoring, Transcutaneous / instrumentation*
Humans
Time Factors
Grant Support
ID/Acronym/Agency:
HL-42637/HL/NHLBI NIH HHS