Document Detail


Cameco anaesthetic ventilator with a modified Mapleson D circuit.
MedLine Citation:
PMID:  532539     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
A modified Mapleson D circuit has been used in connection with a Cameco anaesthetic ventilator during neuroradiological procedures in general anaesthesia. In order to increase mobility of the patient, two or three lengths of corrugated rubber tubing were used to connect the patient to the ventilator. Blood gas analysis was carried out in 20 patients after ventilation to steady state with both circuits. The respiratory minute volume and fresh gas flow were preset in Bain's (Bain & Spoerel 1975) predictions. No significant difference could be detected in respect to Paco2 or Po2, whether 2 or 3 lengths of tubing were used. Mean values of Paco2 were higher compared with the results of Bain (0.37 kPa s.d. 0.50). It is concluded that this system gives maximum mobility of the patient during the radiological procedure and offers reliable adjustment of Paco2, even in patients with apparent increase of intracranial pressure.
Authors:
K N Christensen; J K Jensen
Related Documents :
19770169 - Body mass index is associated with the development of acute respiratory distress syndrome.
16899839 - Management of postintubation tracheobronchial ruptures.
9824019 - Comparing two heat and moisture exchangers, one hydrophobic and one hygroscopic, on hum...
20351029 - Repaired oesophageal atresia: respiratory morbidity and pulmonary function in adults.
8155449 - Risk of aspiration with the laryngeal mask.
22706739 - Investigation into bacteremia and spontaneous bacterial peritonitis in patients with li...
11867409 - The effects of abdominal opening on respiratory mechanics during general anesthesia in ...
22073819 - Prothrombin fragment 1 + 2 (f1 + 2) in effluent is a useful marker for peritoneal perme...
2801159 - Depression among patients with stroke.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Acta anaesthesiologica Scandinavica     Volume:  23     ISSN:  0001-5172     ISO Abbreviation:  Acta Anaesthesiol Scand     Publication Date:  1979 Oct 
Date Detail:
Created Date:  1980-04-17     Completed Date:  1980-04-17     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  0370270     Medline TA:  Acta Anaesthesiol Scand     Country:  DENMARK    
Other Details:
Languages:  eng     Pagination:  419-21     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Aged
Anesthesia, General / instrumentation*
Brain / surgery
Humans
Intracranial Pressure
Middle Aged
Neurosurgical Procedures
Partial Pressure
Respiration, Artificial / instrumentation*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Prediction of postoperative hypoxemia in smokers and non-smokers.
Next Document:  Renal function and fluoride formation and excretion during enflurane anaesthesia.