Document Detail


Brief review: Supplementing oxygen through an airway exchange catheter: efficacy, complications, and recommendations.
MedLine Citation:
PMID:  21465321     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: An airway exchange catheter (AEC) may be employed as a conduit for endotracheal tube placement and for oxygen insufflation or jet ventilation via its lumen. The recent barotrauma-related death of a young healthy patient receiving oxygen insufflated through an AEC prompted the Chief Coroner of Ontario to seek guidelines regarding their use. A literature search was undertaken using a number of search strategies to investigate both the efficacy and complications associated with supplying oxygen through an AEC.
PRINCIPAL FINDINGS: No studies were found comparing either oxygen insufflation or jet-ventilation through an AEC to any standard forms of oxygen therapy. The only case series found using AEC jet ventilation reported that 11% of patients sustained pulmonary barotrauma. Thirteen case reports documented jet ventilation as being associated with pneumothorax, pneumomediastinum, pneumoperitoneum, cardiovascular collapse, and death. In three case series (totalling 76 adults and 20 children) using only oxygen insufflation, no complications were reported.
CONCLUSIONS: Jet ventilation through an AEC may be associated with a significant risk of barotrauma. Oxygen insufflation appears to be associated with a lower risk, but it is not risk-free. The authors caution against the use of an AEC to administer oxygen failing the proven benefit of its use over the use of standard oxygen therapies. Should a patient decompensate with an AEC in situ, tracheal re-intubation is the key management strategy. Supplemental oxygen can be provided using standard techniques prior to tracheal intubation or between attempts. Under emergency circumstances, oxygen insufflation or manual ventilation through an AEC may be considered provided vigilance for barotrauma is maintained and re-intubation is not delayed.
Authors:
Laura V Duggan; J Adam Law; Michael F Murphy
Publication Detail:
Type:  Journal Article; Review     Date:  2011-04-05
Journal Detail:
Title:  Canadian journal of anaesthesia = Journal canadien d'anesthésie     Volume:  58     ISSN:  1496-8975     ISO Abbreviation:  Can J Anaesth     Publication Date:  2011 Jun 
Date Detail:
Created Date:  2011-05-18     Completed Date:  2011-09-12     Revised Date:  2011-12-23    
Medline Journal Info:
Nlm Unique ID:  8701709     Medline TA:  Can J Anaesth     Country:  United States    
Other Details:
Languages:  eng     Pagination:  560-8     Citation Subset:  IM    
Affiliation:
Department of Anesthesiology, Pharmacology and Therapeutics, University of British Columbia, Royal Columbian Hospital, 330 East Columbia Street, New Westminster, BC, V3L 1W7, Canada. laura.duggan@fraserhealth.ca
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Barotrauma / etiology
Catheters*
Humans
Intubation, Intratracheal / adverse effects,  instrumentation*
Oxygen / administration & dosage*
Chemical
Reg. No./Substance:
7782-44-7/Oxygen
Comments/Corrections
Comment In:
Can J Anaesth. 2011 Dec;58(12):1142-3; author reply 1443   [PMID:  22072059 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A comparison between the Boussignac™ continuous positive airway pressure mask and the venturi mask...
Next Document:  Anatomical right posterior sectionectomy: a further expansion of the ultrasound-guided compression t...