Blood pressure monitoring: automated oscillometric devices.
MedLine Citation:
PMID:  1999699     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
An understanding of the principles of blood pressure measurement with automated oscillometric devices (e.g., Dinamap) allows a user to realize the purpose and level of accuracy for which the equipment is designed. Good technique cells for various actions: (1) using a cuff of proper size (too large a cuff will generally work acceptably, but too small a cuff will yield erroneously high readings); (2) squeezing all of the residual air out of the cuff before applying it to the arm or leg; (3) wrapping the cuff snugly around the arm or leg; (4) instructing the patient to refrain from talking or moving and letting nothing press against the cuff during measurement; (5) keeping the cuff and heart at the same horizontal level. (6) When instrument malfunction is suspected, first checking the status of the patient (not the status of the instrument) to be sure that a clinical emergency is not causing the suspected "malfunction"; (7) repeating a reading several times when it is suspicious and, if necessary, modifying the measurement conditions by using, for example, different limbs or different cuffs until the question is resolved; and (8) remembering that a leak in the cuff, hoses, or connection can cause an unexpected malfunction.
Authors:
M Ramsey
Related Documents :
19480489 - In-use comparison of blood pressure measurements from an automated blood pressure instr...
6844129 - Noninvasive determination of elastic properties and diameter of human limb arteries.
23832349 - Regulation of cytochrome c- and quinol oxidases, and piezotolerance of their activities...
1618549 - Pseudohypertension and the measurement of blood pressure.
285709 - The experimental assessment of techniques of measuring biliary pressure.
21393409 - Serous retinal detachment in hypertensive posterior reversible encephalopathy syndrome.
21686209 - Therapeutic hypothermia: what's hot about cold.
10586259 - Variability of presentation in hereditary neuropathy with liability to pressure palsy r...
12717129 - The effects of isoflurane and desflurane on intracranial pressure, cerebral perfusion p...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of clinical monitoring     Volume:  7     ISSN:  0748-1977     ISO Abbreviation:  J Clin Monit     Publication Date:  1991 Jan 
Date Detail:
Created Date:  1991-04-11     Completed Date:  1991-04-11     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  8506707     Medline TA:  J Clin Monit     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  56-67     Citation Subset:  IM    
Affiliation:
Critikon, Inc., Tampa, FL 33631-3800.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Pressure Monitors*
Equipment Failure
Humans
Microcomputers
Oscillometry / instrumentation,  methods

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A target feedback device for ventilatory muscle training.
Next Document:  Skin reflectance pulse oximetry: in vivo measurements from the forearm and calf.