Biomechanical analysis of bicortical versus unicortical locked plating of mid-clavicular fractures.
MedLine Citation:
PMID:  21069362     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
OBJECTIVES: Operative fixation of displaced mid-shaft clavicle fractures has been shown to improve the functional outcomes and decrease the likelihood of non-union; however, little is known about the need for locking screws versus traditional screws. We, therefore, evaluated the strength of unicortical locked plating versus traditional bicortical non-locking fixation methods.
METHODS: Ten matched pairs of fresh, frozen cadaver clavicle specimens were obliquely osteotomized through the mid-shaft to represent the most common fracture pattern. After randomization, the clavicles were repaired using pre-contoured plates with either standard bicortical non-locking screws or unicortical locking screws. The constructs were then potted in cement and tested on a MTS machine using a custom gimble and evaluated for load to failure and axial and rotational stiffness.
RESULTS: There was no significant difference between the constructs in terms of axial stiffness (locking 688.3 ± 306.2 N/mm, non-locking 674.5 ± 613.0 N/mm; p = 0.77) or load to failure (locking 720.1 ± 232.0 N, non-locking 664.8 ± 167.5 N; p = 0.46). However, rotational stiffness varied significantly (locking 1.70 ± 0.91 N-m/mm, non-locking 2.49 ± 0.78 N-m/mm, p = 0.049) with bicortical non-locking constructs exhibiting higher torque values.
CONCLUSIONS: Unicortical fixation using pre-contoured plates and locking screws has a similar biomechanical profile compared to gold standard non-locked bicortical screws in cyclic axial compression and axial load to failure. Non-locking constructs were stiffer under rotational testing. This technique may provide a suitable biomechanical environment for bony healing. This may also improve the safety of clavicle plating by protecting infraclavicular structures from injury during drilling or screw penetration as it obviates the need for bicortical fixation.
Authors:
Daniel Hamman; Derek Lindsey; Jason Dragoo
Publication Detail:
Type:  Journal Article     Date:  2010-11-11
Journal Detail:
Title:  Archives of orthopaedic and trauma surgery     Volume:  131     ISSN:  1434-3916     ISO Abbreviation:  Arch Orthop Trauma Surg     Publication Date:  2011 Jun 
Date Detail:
Created Date:  2011-05-18     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9011043     Medline TA:  Arch Orthop Trauma Surg     Country:  Germany    
Other Details:
Languages:  eng     Pagination:  773-8     Citation Subset:  IM    
Affiliation:
Orthopedic Physicians of Colorado, 799 E. Hampden Ave, #400, Englewood, CO, 80113, USA, danielhamman@gmail.com.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Three kinds of forearm flaps for hand skin defects: experience of 65 cases.
Next Document:  Influence of femoral malrotation on knee joint alignment and intra-articular contract pressures.