Document Detail


Benchtop evaluation of pressure barrier insufflator and standard insufflator systems.
MedLine Citation:
PMID:  22833262     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Previous experimental research has reported minimal differences in pressure maintenance between different versions of standard insufflators (SI). However, a recent report identified potential clinical benefits with a valveless pressure barrier insufflator (PBI). We sought to perform a benchtop objective evaluation of SI and PBI systems.
METHODS: A rigid box system with continuous pressure manometry was used to evaluate a PBI (Surgiquest Airseal) and two SIs (SI1 = Stryker PneumoSure High Flow Insufflator and SI2 = Storz SCB Thermoflator). Pressure maintenance of 15 mmHg was evaluated during experimental conditions of leakage from a 5 mm port site, leakage from a 12 mm port site, and continuous suction.
RESULTS: With leakage from the 5 mm port site, the PBI maintained pressure of >13 mmHg whereas the pressures dropped moderately with the SI1 (7-13 mmHg) and SI2 insufflators (3-7 mmHg) and did not regain goal pressure until leakage was stopped. With leakage from 12 mm port site, the PBI pressure decreased to 9-11 mmHg, whereas the SI1 and SI2 lost insufflation pressures completely. The PBI maintained pressure of >11 mmHg during continuous suction while the SI1 and SI2 lost pressure entirely, and actually showed negative pressure from air suction into the rigid box system. When evaluated statistically with the mixed model repeated measures ANOVA, the SI1 and SI2 performed similarly while the PBI maintained increased pressure.
CONCLUSIONS: In the experimental rigid box system, the PBI more successfully maintained pressure in response to leakage and suction than SIs.
Authors:
Kenneth G Nepple; Dorina Kallogjeri; Sam B Bhayani
Publication Detail:
Type:  Evaluation Studies; Journal Article     Date:  2012-07-26
Journal Detail:
Title:  Surgical endoscopy     Volume:  27     ISSN:  1432-2218     ISO Abbreviation:  Surg Endosc     Publication Date:  2013 Jan 
Date Detail:
Created Date:  2012-12-31     Completed Date:  2013-05-24     Revised Date:  2013-11-22    
Medline Journal Info:
Nlm Unique ID:  8806653     Medline TA:  Surg Endosc     Country:  Germany    
Other Details:
Languages:  eng     Pagination:  333-8     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Analysis of Variance
Equipment Design
Equipment Failure
Insufflation / instrumentation*
Manometry / instrumentation
Pneumoperitoneum, Artificial / instrumentation*
Pressure
Grant Support
ID/Acronym/Agency:
KM1 CA156708/CA/NCI NIH HHS

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A case-controlled matched-pair cohort study of single-incision and conventional laparoscopic gastric...
Next Document:  Integration of transanal specimen extraction into laparoscopic anterior resection with total mesorec...