Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation.
MedLine Citation:
PMID:  20862452     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
PURPOSE: During volume-controlled ventilation, part of the volume delivered is compressed into the circuit. To correct for this phenomenon, modern ventilators use compensation algorithms. Humidity and temperature also influence the delivered volume.
METHODS: In a bench study at a research laboratory in a university hospital, we compared nine ICU ventilators equipped with compensation algorithms, one with a proximal pneumotachograph and one without compensation. Each ventilator was evaluated under normal, obstructive, and restrictive conditions of respiratory mechanics. For each condition, three tidal volumes (V (T)) were set (300, 500, and 800 ml), with and without an inspiratory pause. The insufflated volume and the volume delivered at the Y-piece were measured independently, without a humidification device, under ambient temperature and pressure and dry gas conditions. We computed the actually delivered V (T) to the lung under body temperature and pressure and saturated water vapour conditions (BTPS).
RESULTS: For target V (T) values of 300, 500, and 800 ml, actually delivered V (T) under BTPS conditions ranged from 261 to 396 ml (-13 to +32%), from 437 to 622 ml (-13 to +24%), and from 681 to 953 ml (-15 to +19%), respectively (p < 0.01). Respiratory system mechanics and application of an inspiratory pause significantly affected actually delivered V (T). Assuming a set V (T) of 6 ml/kg of predicted body weight, a difference of 1-2 ml/kg with actually delivered V (T) would be commonly observed.
CONCLUSION: The difference between preset V (T) and actually delivered V (T) is clinically meaningful and differs across modern ICU ventilators.
Authors:
Aissam Lyazidi; Arnaud W Thille; Guillaume Carteaux; Fabrice Galia; Laurent Brochard; Jean-Christophe M Richard
Publication Detail:
Type:  Journal Article     Date:  2010-09-23
Journal Detail:
Title:  Intensive care medicine     Volume:  36     ISSN:  1432-1238     ISO Abbreviation:  Intensive Care Med     Publication Date:  2010 Dec 
Date Detail:
Created Date:  2010-11-16     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7704851     Medline TA:  Intensive Care Med     Country:  United States    
Other Details:
Languages:  eng     Pagination:  2074-80     Citation Subset:  IM    
Affiliation:
Medical Intensive Care Unit, AP-HP, Centre Hospitalier Albert Chenevier, Henri Mondor, Créteil, France.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Vascular hyporesponsiveness to vasopressors in septic shock: from bench to bedside.
Next Document:  Determinants of drug absorption in different ECMO circuits.