Document Detail


Bench model to simulate upper airway obstruction for analyzing automatic continuous positive airway pressure devices.
MedLine Citation:
PMID:  16899832     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Automatic positive airway pressure (APAP) devices are increasingly being used in patients with obstructive sleep apnea. Some APAP devices present an unstable behavior when subjected to some events or artifacts. The aims were to develop a bench model capable of reproducing real flow, snoring, and obstructive patterns and to compare the response of APAP devices based on flow and snoring with other devices using, in addition, the forced oscillation technique (FOT). METHODS: The bench model subjected APAP devices to apneas with and without obstruction, obstructive hypopneas with and without snoring, periods of flow limitation, and artifacts such as leaks and mouth expiration. RESULTS: Almost all the devices increased the pressure when subjected to apneas with obstruction, but at different rates. The time required by each device to reach 10 cm H(2)O ranged from 2.5 to 13 min. In the presence of apneas without obstruction, all the devices based on flow and snoring increased the pressure at the same rate as during apneas with obstruction. However, the devices using FOT did not modify the pressure. Four devices did not modify the pressure in the presence of obstructive hypopneas, and all but one device increased the pressure in the presence of snoring. Mask leaks had little effect on the response of the devices, but four devices increased the pressure during mouth expiration artifacts. CONCLUSIONS: When, in addition to the flow and snoring signals, the measurement of the upper airway resistance is included, the accuracy of the event detection algorithms is improved.
Authors:
Jordi Rigau; Josep M Montserrat; Holger Wöhrle; Diana Plattner; Matthias Schwaibold; Daniel Navajas; Ramon Farré
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Chest     Volume:  130     ISSN:  0012-3692     ISO Abbreviation:  Chest     Publication Date:  2006 Aug 
Date Detail:
Created Date:  2006-08-10     Completed Date:  2006-09-12     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  0231335     Medline TA:  Chest     Country:  United States    
Other Details:
Languages:  eng     Pagination:  350-61     Citation Subset:  AIM; IM    
Affiliation:
Unitat de Biofísica i Bioenginyeria, Facultat de Medicina, Universitat de Barcelona, Spain.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Airway Obstruction / therapy*
Artifacts
Automation*
Benchmarking / methods*
Computer Simulation
Continuous Positive Airway Pressure / instrumentation*
Equipment Design
Humans
Models, Biological
Reproducibility of Results
Comments/Corrections
Comment In:
Chest. 2006 Aug;130(2):312-4   [PMID:  16899825 ]