Document Detail


Automatic implantable cardioverter/defibrillator: inadvertent discharges during permanent pacemaker magnet tests.
MedLine Citation:
PMID:  2440009     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
A patient with an automatic implantable cardioverter defibrillator (AICD) received two inadvertent shocks when a magnet was placed over the pacer during a routine permanent pacer check. Analysis of the rhythm strip suggested that both patients' QRS complexes (133 beats/minute) and asynchronous pacer artifacts (70 beats/minute) were counted by the AICD sensing system and exceeded the rate criteria of 153 beats/minute. This resulted in shocks from the AICD during sinus rhythm at 133 beats/minute. To avoid possible inadvertent shocks, an AICD should be deactivated while a magnet is placed over the pacemaker during a permanent pacer check.
Authors:
S G Kim; S Furman; J A Matos; L E Waspe; R Brodman; J D Fisher
Related Documents :
19494419 - Back from irreversibility: use of percutaneous cardiopulmonary bypass for treatment of ...
15013709 - Ischemia, reperfusion, and the role of surgery in the treatment of cardiogenic shock se...
2026029 - Oxygen consumption in sepsis and septic shock.
1378609 - High and low strength nonsynchronized shocks given during canine ventricular tachycardia.
3130879 - Spectrum of double outlet right ventricle induced by electrical shock at the conotruncu...
17572329 - Pericardiotomy and right ventricular failure: a case report.
10822039 - Cost-effective management of acute atrial fibrillation: role of rate control, spontaneo...
1389719 - Emergency coronary artery surgery after refractory cardiac arrest: a single centre expe...
3500629 - Beneficial effect of nitrates on myocardial glucose utilization in unstable angina pect...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  10     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  1987 May 
Date Detail:
Created Date:  1987-08-05     Completed Date:  1987-08-05     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  579-82     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Electric Countershock / adverse effects*,  instrumentation
Electric Injuries / etiology*
Heart Ventricles
Humans
Magnetics
Male
Middle Aged
Pacemaker, Artificial / adverse effects*
Tachycardia / therapy
Ventricular Fibrillation / therapy

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine