The Hewlett-Packard 47201A ear oximeter and the Radiometer TCM1 transcutaneous oxygen monitor were evaluated for use as noninvasive monitors of arterial oxygenation during inhalational anesthesia in man. Thirty-four healthy adult volunteers were anesthetized to steady states with halothane, enflurane, or isoflurane, and were studied either before or during surgery. Oxygen levels were varied over ranges that included hypoxemia, by manipulating FIO2. Oximeter estimates of SaO2 values and transcutaneous estimates of PaO2 values were compared with conventional measurements of each. Oximeter readings responded rapidly to changes of inspired oxygen concentration and were acceptably accurate estimates of SaO2, except at lower SaO2 levels (less than 80%) during anesthesia without surgery. Transcutaneous oxygen tension readings responded relatively slowly to changes of FIO2 and were frequently inaccurate reflections of PaO2 values. We consider this oximeter suitable as a monitor of arterial oxygenation during anesthesia, but find the transcutaneous electrode unsatisfactory.
Authors:
R L Knill; J L Clement; H T Kieraszewicz; B G Dodgson
Related Documents :
15278576 - The neuromuscular effects of sevoflurane and isoflurane alone and in combination with v...
9951756 - Influence of pulse oximetry and capnography on time to diagnosis of critical incidents ...
7910526 - Preliminary investigations of the clinical pharmacology of three short-acting non-depol...
24417616 - Long-term outcome and use of 6-minute walk test in west highland white terriers with id...
24434816 - Changes in the electrical properties of the electrode-skin-underlying tissue composite ...
1415746 - Calculation of oxygen diffusion across the surface of isolated perfused hearts.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Anesthesia and analgesia     Volume:  61     ISSN:  0003-2999     ISO Abbreviation:  Anesth. Analg.     Publication Date:  1982 Jul 
Date Detail:
Created Date:  1982-08-07     Completed Date:  1982-08-07     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  1310650     Medline TA:  Anesth Analg     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  582-6     Citation Subset:  AIM; IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Arteries
Electrodes
Evaluation Studies as Topic
Female
Humans
Male
Monitoring, Physiologic / instrumentation*
Oximetry / instrumentation*
Oxygen / blood*
Skin
Chemical
Reg. No./Substance:
7782-44-7/Oxygen

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Oxygen transfer from mother to fetus during cesarean section under epidural anesthesia.
Next Document:  Pharmacokinetics of inhalation anesthetics: a three-compartment linear model.