PURPOSE: Some patients fitted with an AMS 800™ type artificial urinary sphincter report urinary leaks in the sitting position whose mechanism remains unclear. In this experimental study we explain these residual leaks by characterizing fluid transfer and pressure variations in the device during cuff compression.
MATERIALS AND METHODS: Two complete sets of AMS 800 devices were tested on an original test bench. A qualitative study using methylene blue dye was performed to assess fluid transfer within the device during manual cuff compression. After estimating the pressures exerted on the perineal floor in the sitting position, pressure variations in the AMS 800 components were measured in real time using the pressure sensors of a urodynamic workstation.
RESULTS: When the AMS 800 cuff was compressed between 2 fingers, reflux of the blue dye into the balloon through the pump resistor was observed. After the cuff was compressed for 5, 10, 15 and 20 seconds at a pressure of 150 cm H2O, the minimal pressure measurements in the cuff were 37, 25, 14 and 12 cm H2O, respectively. Furthermore, the time required for cuff re-pressurization was 68, 87, 101 and 117 seconds, respectively.
CONCLUSIONS: Urinary leaks reported in sitting position by patients fitted with an AMS 800 may be explained by compression of the urethral cuff between the pelvic floor and the chair, which causes emptying of the cuff into the balloon. Patients should be reassured that there is no implant dysfunction and may adopt simple lifestyle strategies to circumvent these leaks.
Authors:
Aurélien Beaugerie; Véronique Phé; Gavishsing Munbauhal; Emmanuel Chartier-Kastler; Pierre Mozer
Related Documents :
7678667 - Improvement of long-term preservation of isolated arrested rat heart: beneficial effect...
11016307 - Reducing postischemic reperfusion damage in neonates using a terminal warm substrate-en...
8096207 - A cd18 antibody prevents lung injury but not hypotension after intestinal ischemia-repe...
9563607 - Cardiac function and glycogen content after twenty-four-hour preservation with various ...
12381797 - The effects of range-of-motion therapy on the plantar pressures of patients with diabet...
17082717 - Predictive factors for masked hypertension within a population of controlled hypertensi...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2014-03-03
Journal Detail:
Title:  The Journal of urology     Volume:  192     ISSN:  1527-3792     ISO Abbreviation:  J. Urol.     Publication Date:  2014 Aug 
Date Detail:
Created Date:  2014-07-18     Completed Date:  2014-09-18     Revised Date:  2014-11-03    
Medline Journal Info:
Nlm Unique ID:  0376374     Medline TA:  J Urol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  483-7     Citation Subset:  AIM; IM    
Copyright Information:
Copyright © 2014 American Urological Association Education and Research, Inc. Published by Elsevier Inc. All rights reserved.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Humans
Male
Postoperative Complications / etiology*
Posture*
Pressure
Prosthesis Design
Urinary Incontinence / etiology*
Urinary Sphincter, Artificial*
Comments/Corrections
Comment In:
Nat Rev Urol. 2014 May;11(5):245   [PMID:  24662728 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Predicting Risk of Erectile Dysfunction Following Pelvic Fracture Urethral Injury in Children.
Next Document:  Evaluation of Allergic Conjunctivitis by Thermography.