BACKGROUND: It has been suggested that limited active ROM of reverse shoulder prostheses relates to lack of strength. However, the postoperative strength has not been quantified.
QUESTIONS/PURPOSES: We therefore measured joint torques in patients with reverse shoulder prostheses and correlated torques with functional scores.
METHODS: We recruited 33 patients (age, 72 ± 8 years) with a reverse prosthesis (37 shoulders, 21 primary and 16 revisions). We obtained Constant-Murley, DASH, and Simple Shoulder Test ([D]SST) scores, and performed two isokinetic protocols (abduction/adduction and external/internal rotation) at 60° per second. Minimum followup was 4 months (average, 23 months; range, 4-63 months).
RESULTS: Twenty-three patients (24 shoulders; 13 primaries, 11 revisions) were able to perform at least one of the defined tasks. Mean abduction and adduction torques were 15 Nm ± 7 Nm and 16 Nm ± 10 Nm (19%-78% of normal shoulders). External and internal rotation tasks could be performed by only 13 patients (14 shoulders; nine primary, five revisions) generating 9 Nm ± 4 Nm and 8 Nm ± 3 Nm, respectively (13%-71% of normal shoulders). We found moderate correlations between Constant-Murley, DASH and (D)SST (D = Dutch translation) scores and abduction and external rotation.
CONCLUSIONS: Patients with a reverse prosthesis had reduced strength when compared with normal values reported in the literature (only 65% of patients could perform the protocol). This effect was greatest for external rotation and might explain clinical outcomes with which a moderately strong relationship was observed. Our observations suggest limited strength is a major factor in reduced ROM.
Authors:
Tjarco D W Alta; H E J Veeger; Thomas W J Janssen; W Jaap Willems
Related Documents :
2009399 - The tobacco withdrawal syndrome in unaided quitters.
19717249 - Long-term outcomes of office-based buprenorphine/naloxone maintenance therapy.
22933569 - Failing stentless aortic valves: redo aortic root replacement or valve in a valve?
9872709 - Treatment goals for panic disorder.
6323909 - Cyanamide-induced liver injury. a predictable lesion.
1666649 - Treatment outcomes of impaired physicians in oklahoma.
17218999 - The effectiveness of transvaginal anterior colporrhaphy reinforced with polypropylene m...
8572039 - A one-year multicenter study of efficacy and safety of a continuous, low-dose, estradio...
22945849 - Do endobronchial valves improve outcomes in patients with emphysema?
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2012-02-11
Journal Detail:
Title:  Clinical orthopaedics and related research     Volume:  470     ISSN:  1528-1132     ISO Abbreviation:  Clin. Orthop. Relat. Res.     Publication Date:  2012 Aug 
Date Detail:
Created Date:  2012-07-10     Completed Date:  2012-09-20     Revised Date:  2013-06-26    
Medline Journal Info:
Nlm Unique ID:  0075674     Medline TA:  Clin Orthop Relat Res     Country:  United States    
Other Details:
Languages:  eng     Pagination:  2185-92     Citation Subset:  AIM; IM    
Affiliation:
Department of Orthopaedic Surgery and Traumatology, Onze Lieve Vrouwe Gasthuis, Amsterdam, The Netherlands. tdw.alta@yahoo.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Arthroplasty, Replacement / adverse effects,  methods*
Disability Evaluation
Female
Humans
Joint Prosthesis*
Male
Muscle Strength / physiology*
Muscle Strength Dynamometer
Pilot Projects
Postoperative Complications*
Prosthesis Failure
Range of Motion, Articular / physiology
Recovery of Function
Shoulder / physiopathology*,  surgery
Shoulder Joint / physiopathology*,  surgery
Torque
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Clin Orthop Relat Res
Journal ID (iso-abbrev): Clin. Orthop. Relat. Res
ISSN: 0009-921X
ISSN: 1528-1132
Publisher: Springer-Verlag, New York
Article Information
Download PDF
© The Author(s) 2012
Received Day: 21 Month: 8 Year: 2011
Accepted Day: 30 Month: 1 Year: 2012
Electronic publication date: Day: 11 Month: 2 Year: 2012
pmc-release publication date: Day: 11 Month: 2 Year: 2012
Print publication date: Month: 8 Year: 2012
Volume: 470 Issue: 8
First Page: 2185 Last Page: 2192
ID: 3392387
PubMed Id: 22328239
Publisher Id: 2277
DOI: 10.1007/s11999-012-2277-8

Are Shoulders with A Reverse Shoulder Prosthesis Strong Enough? A Pilot Study
Tjarco D. W. Alta, MD1 Address: tdw.alta@yahoo.com
H. E. J. Veeger, PhD23
Thomas W. J. Janssen, PhD24
W. Jaap Willems, MD, PhD1
1Department of Orthopaedic Surgery and Traumatology, Onze Lieve Vrouwe Gasthuis, P.O. Box 95500, 1090 HM Amsterdam, The Netherlands
2Research Institute MOVE, Faculty of Human Movement Sciences, VU University, Amsterdam, The Netherlands
3Section Biomechatronics and Biorobotics, Faculty of Mechanical, Maritime and Materials Engineering, Delft University of Technology, Delft, The Netherlands
4Duyvensz-Nagel Research Laboratory, Reade, Centre for Rehabilitation and Rheumatology, Amsterdam, The Netherlands

Introduction

The reverse shoulder prosthesis is one surgical option for treatment of cuff tear arthropathy and shoulder pseudoparalysis resulting from a massive cuff tear, severe fractures, prosthetic revision, and tumor surgery [5]. Owing to the mechanical advantage of a medialized center of rotation, the reverse shoulder prosthesis offers a potentially valuable surgical option and has become an alternative in situations in which the rotator cuff and/or the proximal humerus are destroyed or absent [5], with a satisfying reduction of pain after surgery [12, 31].

However, because of the wide variation in published values for active elevation after reverse shoulder replacement (ranging from 88° to 138° [3, 36]), the degree to which this prosthesis restores arm strength is not fully defined. Bergmann et al. previously found a better passive than active ROM [3] and presumed that the limited glenohumeral motion of the reverse prosthesis resulted from a lack of joint torque generation rather than a structural limitation caused by the prosthetic design.

Shoulder strength mostly has been evaluated using subjective methods such as traditional manual muscle testing and handheld dynamometry, which mainly focus on isometric muscle strength. The strength measure in the Constant-Murley score [8, 9] is also isometric. Because most functional activities are dynamic, evaluating isokinetic shoulder strength may be more appropriate when relating strength to functional performance and clinical outcome. However, data for isokinetic strength measurements around the shoulder are available only for normal healthy subjects [7], patients after open fixation of glenoid rim fractures [33], open [1, 10] and arthroscopic anterior stabilization [15, 21], rotator cuff surgery [4, 11, 14, 34, 43], with adhesive capsulitis [26, 27, 41], subacromial impingement [16, 24, 30], and pectoralis major muscle rupture [17], but not for patients with a reverse shoulder prosthesis.

Our objective was to perform a pilot study to measure isometric shoulder strength in patients who underwent either a primary or revision reverse shoulder replacement. We asked the following questions: (1) what joint torques can patients with a reverse shoulder prosthesis produce isokinetically, and (2) does this force-generating capacity correlate with functional scores?


Materials and Methods

Between May 2000 and September 2007, we treated 45 patients (49 shoulders) with a reverse shoulder prosthesis (Tornier®; Edina, MN, USA). Of these, 33 patients (19 women and 14 men), volunteered to participate in this study. Ten patients had surgery on the left side, 19 on the right side, and four on both sides (total of 37 shoulders). In 21 patients, the indication for the reverse prosthesis was cuff tear arthropathy, and in 16 patients the indication was revision after a failed primary placed hemiprosthesis or total shoulder prosthesis. The average time between surgery and measurement was 23 ± 14 months (range, 4–63 months). Mean age of the patients was 72 ± 8 years (range, 58–85 years). The minimum clinical followup was 4 months (mean, 23 months; range, 4–63 months). The institutional ethics committee approved the research protocol and all patients gave their written informed consent before the experiment.

All patients underwent surgery under general anesthesia with an interscalene nerve block in the beach-chair position. We used a standard deltopectoral approach in all patients. All glenoid components had been placed inferior on the glenoid surface with no inferior or superior inclination. Thirty patients had a 36-mm component implanted and seven had a 42-mm sphere diameter. The humeral components had all been placed in 10° to 20° retroversion, and in 32 patients, the teres minor and subscapularis muscles were still intact.

Postoperative management was the same for all patients, consisting of a sling and passive ROM exercises for 6 weeks. After 6 weeks, active assisted ROM exercises were started and at 3 months, strengthening exercises were added to the rehabilitation program.

Shoulder strength was measured with an isokinetic dynamometer, which provides constant velocity with accommodating resistance throughout a joint’s ROM. This resistance is provided using an electric or hydraulic servo-controlled mechanism at a user-defined constant velocity [13]. Two isokinetic protocols were performed to measure the strength of the subjects’ shoulder muscles on the surgically treated side using the Biodex System 3 Pro dynamometer (Biodex Medical Systems, New York, NY, USA). These protocols consisted of an abduction and adduction task and an external and internal rotation task with the arm in 60° abduction in a sitting position with securing bands around the subject’s chest and pelvis. For the abduction and adduction task, the chair was rotated 75° around the vertical axis with the dynamometer in neutral position and 10° tilted. For the external and internal rotation task, the chair was in the neutral position with the dynamometer rotated 20° and 50° tilted. After one session of the abduction and adduction or the external and internal rotation task, the subject had a 60-second recovery time after which the same task was repeated. All tasks were repeated five times at 60° per second with a minimum standard threshold of 15° per second to start the measurements defined by the Biodex System. For each motion, the average maximal torque (Nm) was determined over the two sessions. The subjects were instructed and encouraged to reach the highest possible force level during these tasks. Negative axial rotation was defined as external rotation and positive axial rotation as internal rotation.

Codine et al. [7] reported a systematic review of the literature on isokinetic strength of the shoulder until 2005. We used PubMed to identify other articles providing data for normal shoulder torque values from 2005 and onward [2, 20, 37, 42]. From those studies we took the abduction-adduction and/or external-internal rotation torque values for 60° per second or less and combined those to make comparison possible with our obtained data (Table 1).

For clinical evaluation, we obtained preoperative and postoperative (absolute and relative) Constant-Murley scores [8, 9], postoperative DASH score [23], and the (D)SST [23, 40]. The absolute Constant-Murley score assesses the overall shoulder function and has a maximum score of 100 points. The relative Constant-Murley score is corrected for the age- and sex-related decline in force-generating capacity [46]. It is expressed as a percentage of the respective reference values. The DASH is a 30-item questionnaire that evaluates functional disability in everyday activities, work, and sports. It includes symptoms, physical, social, and psychological function. A DASH score of 0 indicates good shoulder function, or no disability, and the maximum score of 100 indicates no function. The (D)SST is a questionnaire consisting of 13 yes or no questions including subjective items and items that require patients to complete a physical exercise. It evaluates shoulder function in daily activities and a maximum score of 13 indicates good shoulder function.

We used a t-test to determine the difference in mean maximum generated torque at 60° per second between primary and revision cases. For this same group, the effect size was determined by calculating the Cohen’s d. The relationship between the clinical outcome scores (Constant-Murley, DASH, and [D]SST) and strength data was evaluated on the basis of a Pearson product-moment correlation.


Results

Only 23 patients (24 shoulders; 13 primary and 11 revisions) were able to generate sufficient velocity to perform the test, resulting in a mean abduction torque of 15.2 Nm ± 6.6 Nm for the whole group with no substantially better value for the primary prostheses compared with the revisions (Table 2). For the external and internal rotation torques, these values varied between 13% and 71%. We found similar torque values for adduction also with no major difference between primary and revision cases. The external and internal rotation tasks could be performed by only 13 patients (14 shoulders; nine primaries, five revisions). Mean external rotation torque was 9.3 Nm ± 4.4 Nm for the whole group with no major differences between the primary and revision groups. Internal rotation force tended to be higher (p = 0.07) for primary prostheses with a torque of 8.2 Nm ± 2.6 Nm for the whole group (Table 2). Compared with normal healthy subjects (Table 1), patients with a reverse prosthesis who could generate sufficient force to perform the tasks had abduction and adduction torques of 19% to 78% of those of a normal shoulder at a velocity of 60° per second.

We found a correlation between the postoperative Constant-Murley score and the abduction and external rotation torques (Fig. 1). Similar correlations were found for the DASH score and (D)SST (Table 3), with the maximum torque values at 60° per second. There was no major correlation for the adduction and internal rotation motions. An overview of all the clinical outcome scores of the whole group, the primary and the revision cases is presented (Table 4).


Discussion

The reverse shoulder prosthesis provides a surgical option for conditions such as cuff tear arthropathy, shoulder pseudoparalysis resulting from massive cuff tear, severe fractures, prosthetic revision, and tumor surgery [5] with generally satisfying postoperative results [12, 31]. However, the contribution of this prosthesis to restoration of arm function is less clear. Previous research [3] suggests the limited glenohumeral motion of the reverse prosthesis seems to be the result of a lack of joint torque generation rather than a structural limitation caused by the prosthetic design. Therefore, the evaluation of isokinetic shoulder strength after reverse shoulder replacement may be of interest in modeling dynamic upper extremity function, particularly where comparative data are not currently available for this clinical scenario. We therefore (1) determined joint torques in patients with a reverse shoulder prosthesis and (2) determined whether force-generating capacity correlates with functional scores.

We note limitations to our study, one of which is the absence of proper control data. First, ideally a comparison would be made with an age-matched control group without cuff disorders. However, with a prevalence of 31% of asymptomatic (ie, unrecognized) cuff tears in individuals between 70 and 79 years old [38] and a prevalence of 51% in individuals older than 80 years [38], this is not feasible without extensive screening. Another possibility would be to compare the outcomes with those of the contralateral side in the same patient. However, cuff disorders in the contralateral shoulder are not uncommon, as reported by Yamaguchi et al. [45] in their demographic and morphologic study of rotator cuff disease. The average age for patients with a bilateral cuff tear in their group was 67.8 years and logistic regression analysis indicated a 50% likelihood of a bilateral tear after the age of 66 years. Furthermore, patients with a full-thickness symptomatic tear had a 35.5% prevalence of a full-thickness tear on the contralateral side. In our patient population, 12% already had a reverse prosthesis on both sides, showing that our patient group was not suitable to use the contralateral side as a comparison. As a consequence data had to be compared with norm data from the literature. Second, we had a broad range of followup times for the force measurements and clinical outcome scores. Ideally the measurements should have been performed at the same time postoperatively for every patient with a minimum followup of 1 year. This is also true for the clinical outcome scores, because they require time to stabilize. In the scope of this study, it was not possible to include patients with the same followup period, as this would have required an inclusion period of several years. We therefore chose to include all patients available from our pool of treated patients, which inevitably led to a large range in followup times. It is not certain what effect the followup time will have had, which especially applies for the elderly population for whom recovery might be counteracted by ageing effects. Given the number of available patients, controlling for age and followup will be virtually impossible whereas including larger groups and testing for those factors also do not seem to be realistic options. Third, we limited our measurements to 60° per second. Isokinetic strength measurements have been performed at different velocities, mostly from 60° per second to 300° per second [7] with some exceptions at 30° per second (Table 1). In these measurements, the applied torque needs to increase above a threshold value to successfully perform a certain task at higher speeds. Because the physiologic changes at older age lead to a decline in force-generating capacity and the reverse shoulder prosthesis is implanted mainly in patients with a mean age of 72 years [18, 36], similar to the average age of the participants in our study, we decided to apply a relatively low velocity of 60° per second. Considering our data and the number of patients unable to perform the tests (Table 2), it appeared that even 60° per second was too high for most of the patients with a reverse prosthesis. Future research investigating force production of this patient population should incorporate velocities less than 60° per second. Whether a lower velocity would lead to substantially more successful tests however is unknown; in our protocol a standard threshold of 15° per second was used to start measurements, which even proved to be too much for some of our patients.

Trying to place our obtained torque values (Table 2) in perspective, we compared our data with those of normal healthy subjects (Table 1). From this comparison we can conclude that patients with a reverse prosthesis who can generate sufficient force to perform the tasks have abduction and adduction torques of 19% to 78% of those of a normal shoulder at a velocity of 60° per second. For the external and internal rotation torques, these values vary between 13% and 71%. However, those normal values were based on younger subjects than our group of patients and in most series they include groups of athletes. If we compare our data with the only age-related series of Verney et al. [42], our patients have an abduction torque of 33% of that of 10 male elderly volunteers. It is not clear what causes this relatively low abduction torque. From the total of 37 shoulders, only 23 patients (24 shoulders) could generate enough force to perform the abduction and adduction tasks and for the external and internal rotation, the number of patients was even less (Table 2). The difference between those two tasks can be explained by the changed biomechanics caused by the reverse shoulder prosthesis. By displacing the center of rotation medially, more fibers of the anterior and posterior parts of the deltoid muscle are recruited for anteflexion or abduction of the arm and therefore fewer fibers are available to internally or externally rotate the arm [5]. Our study group included patients with primary and revision implantations. In revision surgery with a reverse prosthesis, the improvement of function is reportedly only to approximately 70° of active elevation [25], with a higher complication rate [44] than with primary surgery. Therefore, we expected to find a difference in force-generating capacity between the two groups (Table 2) in favor of the primary prosthesis. However, this could not be confirmed for the abduction and adduction tasks because 62% of the primary and 69% of the revision cases were able to generate enough force. For the internal and external rotation tasks, it was 43% and 31% respectively, confirming our expectation and explained by the previously mentioned change of biomechanics after a reverse prosthesis.

Impaired shoulder strength is likely one of the causes of active ROM limitations. The correlations we found between clinical outcome scores (Constant-Murley, DASH, and [D]SST) and the abduction and external rotation torque values (Table 3) support this contention. Functional outcome probably is not determined by simple ROM ranges alone, but also by the actual capacity for material handling in elevated and axially rotated arm positions. For example, it can be expected that patients who have good anteflexion or abduction with limited external rotation strength define their functional outcome as poor. Therefore, it seems logical that greater external rotation torque provides a better functional outcome. Although our findings support this notion, only 13 of a total of 37 shoulders actually were able to generate enough force to perform the tasks at 60° per second. Testing under lighter conditions (30° per second) could have provided more data but probably would not have led to another observation.

Patients with a reverse prosthesis were moderately to strongly limited in strength, which was the case for abduction and adduction and even more for external and internal rotation. However, future isokinetic data collection in these patients should be performed at a lower velocity than 60° per second. Results for strength correlated with clinical outcome scores (Constant-Murley, DASH, and [D]SST) indicating moderately strong relationships and a moderate predictive value of the outcome scores. Although it is likely that lower isokinetic shoulder strength in patients with joint arthroplasties is a major factor in reduced ROM, the actual causes of loss of strength would need to be identified in future studies.


Notes

One of the authors (WJW) received funding for this research from Tornier SA, France.

All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research editors and board members are on file with the publication and can be viewed on request.

Each author certifies that his or her institution approved the human protocol for this investigation, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.

Patients were operated on at the Onze Lieve Vrouwe Gasthuis and Medisch Centrum Alkmaar, The Netherlands. The measurements were performed at Duyvensz-Nagel Research Laboratory, Reade, Centre for Rehabilitation and Rheumatology where the Biodex system is situated.

We thank Jeroen H. M. Bergmann PhD, Department of Cancer & Surgery, Imperial College London, UK, for assistance with the measurements and critical comments while writing the draft. We also thank Bart J. Burger MD, PhD, orthopaedic surgeon at Medic Centrum Alkmaar, The Netherlands, for allowing us to perform measurements on his patients.

Open Access

This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.


References
1.. Amako M,Imai T,Okamura K. Recovery of shoulder rotational muscle strength after a combined Bankart and modified Bristow procedureJ Shoulder Elbow Surg.Year: 20081773874310.1016/j.jse.2008.01.14518558501
2.. Ambrosio F,Boninger ML,Souza AL,Fitzgerald SG,Koontz AM,Cooper RA. Biomechanics and strength of manual wheelchair usersJ Spinal Cord Med.Year: 20052840741416869087
3.. Bergmann JH,Leeuw M,Janssen TW,Veeger DH,Willems WJ. Contribution of the reverse endoprosthesis to glenohumeral kinematicsClin Orthop Relat Res.Year: 200846659459810.1007/s11999-007-0091-518264847
4.. Bigoni M,Gorla M,Guerrasio S,Brignoli A,Cossio A,Grillo P,Marinoni EC. Shoulder evaluation with isokinetic strength testing after arthroscopic rotator cuff repairsJ Shoulder Elbow Surg.Year: 20091817818310.1016/j.jse.2008.09.00219111477
5.. Boileau P,Watkinson DJ,Hatzidakis AM,Balg F. Grammont reverse prosthesis: design, rationale, and biomechanicsJ Shoulder Elbow Surg.Year: 2005141 suppl147S161S10.1016/j.jse.2004.10.00615726075
6.. Codine P,Bernard PL,Pocholle M,Benaim C,Brun V. Influence of sports discipline on shoulder rotator cuff balanceMed Sci Sports Exerc.Year: 1997291400140510.1097/00005768-199711000-000029372473
7.. Codine P,Bernard PL,Pocholle M,Herisson C. Isokinetic strength measurement and training of the shoulder: methodology and results] [in FrenchAnn Readapt Med Phys.Year: 200548809210.1016/j.annrmp.2004.07.00215748773
8.. Constant CR,Gerber C,Emery RJ,Sojbjerg JO,Gohlke F,Boileau P. A review of the Constant score: modifications and guidelines for its useJ Shoulder Elbow Surg.Year: 20081735536110.1016/j.jse.2007.06.02218218327
9.. Constant CR,Murley AH. A clinical method of functional assessment of the shoulderClin Orthop Relat Res.Year: 19872141601643791738
10.. Dauty M,Dominique H,Helena A,Charles D. Evolution of the isokinetic torque of shoulder rotators before and after 3 months of shoulder stabilization by the Latarjet technique] [in FrenchAnn Readapt Med Phys.Year: 20075020120810.1016/j.annrmp.2007.01.00917320998
11.. Demirors H,Circi E,Akgun RC,Tarhan NC,Cetin N,Akpinar S,Tuncay IC. Correlations of isokinetic measurements with tendon healing following open repair of rotator cuff tearsInt Orthop.Year: 20103453153610.1007/s00264-009-0827-919533125
12.. Drake GN,O’Connor DP,Edwards TB. Indications for reverse total shoulder arthroplasty in rotator cuff diseaseClin Orthop Relat Res.Year: 20104681526153310.1007/s11999-009-1188-920049573
13.. Drouin JM,Valovich-mcLeod TC,Shultz SJ,Gansneder BM,Perrin DH. Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurementsEur J Appl Physiol.Year: 200491222910.1007/s00421-003-0933-014508689
14.. Ellenbecker TS,Elmore E,Bailie DS. Descriptive report of shoulder range of motion and rotational strength 6 and 12 weeks following rotator cuff repair using a mini-open deltoid splitting techniqueJ Orthop Sports Phys Ther.Year: 20063632633516715833
15.. Ellenbecker TS,Mattalino AJ. Glenohumeral joint range of motion and rotator cuff strength following arthroscopic anterior stabilization with thermal capsulorraphyJ Orthop Sports Phys Ther.Year: 19992916016710322589
16.. Erol O,Ozcakar L,Celiker R. Shoulder rotator strength in patients with stage I-II subacromial impingement: relationship to pain, disability, and quality of lifeJ Shoulder Elbow Surg.Year: 20081789389710.1016/j.jse.2008.05.04318774737
17.. Fleury AM,Silva AC,Pochini A,Ejnisman B,Lira CA,Andrade Mdos S. Isokinetic muscle assessment after treatment of pectoralis major muscle rupture using surgical or non-surgical proceduresClinics Sao PauloYear: 20116631332010.1590/S1807-5932201100020002221484052
18.. Gerber C,Pennington SD,Nyffeler RW. Reverse total shoulder arthroplastyJ Am Acad Orthop Surg.Year: 20091728429519411640
19.. Greenfield BH,Donatelli R,Wooden MJ,Wilkes J. Isokinetic evaluation of shoulder rotational strength between the plane of scapula and the frontal planeAm J Sports Med.Year: 19901812412810.1177/0363546590018002022343977
20.. Harbo T,Brincks J,Andersen H. Maximal isokinetic and isometric muscle strength of major muscle groups related to age, body mass, height, and sex in 178 healthy subjectsEur J Appl Physiol.Year: 201111226727510.1007/s00421-011-1975-321537927
21.. Hiemstra LA,Sasyniuk TM,Mohtadi NG,Fick GH. Shoulder strength after open versus arthroscopic stabilizationAm J Sports Med.Year: 20083686186710.1177/036354650831442918319347
22.. Ivey FM Jr,Calhoun JH,Rusche K,Bierschenk J. Isokinetic testing of shoulder strength: normal valuesArch Phys Med Rehabil.Year: 1985663843864004537
23.. Kirkley A,Griffin S,Dainty K. Scoring systems for the functional assessment of the shoulderArthroscopy.Year: 2003191109112010.1016/j.arthro.2003.10.03014673454
24.. Leroux JL,Codine P,Thomas E,Pocholle M,Mailhe D,Blotman F. Isokinetic evaluation of rotational strength in normal shoulders and shoulders with impingement syndromeClin Orthop Relat Res.Year: 19943041081158020202
25.. Levy JC,Virani N,Pupello D,Frankle M. Use of the reverse shoulder prosthesis for the treatment of failed hemiarthroplasty in patients with glenohumeral arthritis and rotator cuff deficiencyJ Bone Joint Surg Br.Year: 20078918919510.2106/JBJS.E.0131017322433
26.. Liem D,Meier F,Thorwesten L,Marquardt B,Steinbeck J,Poetzl W. The influence of arthroscopic subscapularis tendon and capsule release on internal rotation strength in treatment of frozen shoulderAm J Sports Med.Year: 20083692192610.1177/036354650731309018272795
27.. Lin HC,Li JS,Lo SF,Shih YF,Lo CY,Chen SY. Isokinetic characteristics of shoulder rotators in patients with adhesive capsulitisJ Rehabil Med.Year: 20094156356810.2340/16501977-037819543668
28.. McMaster WC,Long SC,Caiozzo VJ. Isokinetic torque imbalances in the rotator cuff of the elite water polo playerAm J Sports Med.Year: 199119727510.1177/0363546591019001122008934
29.. McMaster WC,Long SC,Caiozzo VJ. Shoulder torque changes in the swimming athleteAm J Sports Med.Year: 19922032332710.1177/0363546592020003151636864
30.. Moraes GF,Faria CD,Teixeira-Salmela LF. Scapular muscle recruitment patterns and isokinetic strength ratios of the shoulder rotator muscles in individuals with and without impingement syndromeJ Shoulder Elbow Surg.Year: 2008171 suppl48S53S10.1016/j.jse.2007.08.00718201657
31.. Naveed MA,Kitson J,Bunker TD. The Delta III reverse shoulder replacement for cuff tear arthropathy: a single-centre study of 50 consecutive proceduresJ Bone Joint Surg Br.Year: 201193576121196544
32.. Otis JC,Warren RF,Backus SI,Santner TJ,Mabrey JD. Torque production in the shoulder of the normal young adult male: the interaction of function, dominance, joint angle, and angular velocityAm J Sports Med.Year: 19901811912310.1177/0363546590018002012343976
33.. Raiss P,Baumann F,Akbar M,Rickert M,Loew M. Open screw fixation of large anterior glenoid rim fractures: mid- and long-term results in 29 patientsKnee Surg Sports Traumatol Arthrosc.Year: 20091719520310.1007/s00167-008-0677-919039575
34.. Rokito AS,Zuckerman JD,Gallagher MA,Cuomo F. Strength after surgical repair of the rotator cuffJ Shoulder Elbow Surg.Year: 19965121710.1016/S1058-2746(96)80025-58919437
35.. Sirota SC,Malanga GA,Eischen JJ,Laskowski ER. An eccentric- and concentric-strength profile of shoulder external and internal rotator muscles in professional baseball pitchersAm J Sports Med.Year: 199725596410.1177/0363546597025001119006693
36.. Sirveaux F,Favard L,Oudet D,Huquet D,Walch G,Mole D. Grammont inverted total shoulder arthroplasty in the treatment of glenohumeral osteoarthritis with massive rupture of the cuff: results of a multicentre study of 80 shouldersJ Bone Joint Surg Br.Year: 20048638839510.1302/0301-620X.86B3.1402415125127
37.. Stickley CD,Hetzler RK,Freemyer BG,Kimura IF. Isokinetic peak torque ratios and shoulder injury history in adolescent female volleyball athletesJ Athl Train.Year: 20084357157710.4085/1062-6050-43.6.57119030134
38.. Tempelhof S,Rupp S,Seil R. Age-related prevalence of rotator cuff tears in asymptomatic shouldersJ Shoulder Elbow Surg.Year: 1999829629910.1016/S1058-2746(99)90148-910471998
39.. Tis LL,Maxwell T. The effect of positioning on shoulder isokinetic measures in femalesMed Sci Sports Exerc.Year: 1996281188119210.1097/00005768-199609000-000168883008
40.. Laarhoven HA,Slaa RL,Boer LM,Willems WJ. The Dutch Simple Shoulder Test, an efficient way of assessing shoulder functionDutch J Orthop.Year: 2001858
41.. Meeteren J,Roebroeck ME,Selles RW,Stam HJ. Responsiveness of isokinetic dynamometry parameters, pain and activity level scores to evaluate changes in patients with capsulitis of the shoulderClin Rehabil.Year: 20062049650110.1191/0269215506cr983oa16892932
42.. Verney J,Kadi F,Saafi MA,Piehl-Aulin K,Denis C. Combined lower body endurance and upper body resistance training improves performance and health parameters in healthy active elderlyEur J Appl Physiol.Year: 20069728829710.1007/s00421-006-0175-z16770464
43.. Walker SW,Couch WH,Boester GA,Sprowl DW. Isokinetic strength of the shoulder after repair of a torn rotator cuffJ Bone Joint Surg Am.Year: 198769104110443654695
44.. Wall B,Nove-Josserand L,O’Connor DP,Edwards TB,Walch G. Reverse total shoulder arthroplasty: a review of results according to etiologyJ Bone Joint Surg Am.Year: 2007891476148510.2106/JBJS.F.0066617606786
45.. Yamaguchi K,Ditsios K,Middleton WD,Hildebolt CF,Galatz LM,Teefey SA. The demographic and morphological features of rotator cuff disease: a comparison of asymptomatic and symptomatic shouldersJ Bone Joint Surg Am.Year: 2006881699170410.2106/JBJS.E.0083516882890
46.. Yian EH,Ramappa AJ,Arneberg O,Gerber C. The Constant score in normal shouldersJ Shoulder Elbow Surg.Year: 20051412813310.1016/j.jse.2004.07.00315789004

Figures

[Figure ID: Fig1]
Fig. 1 

Pearson’s correlation between the maximal abduction and external rotation torque at 60° per second and the postoperative Constant-Murley score show a correlation between the force-generating capacity of patients with a reverse shoulder prosthesis and their postoperative Constant-Murley score for abduction and external rotation.



Tables
[TableWrap ID: Tab1] Table 1 

Maximum generated force in Nm for our data compared with the literature at similar or lower velocities


Study Subject Mean age (years) Dominance or side Velocity Abduction Adduction External rotation Internal rotation
Ambrosio et al. [2] M and F wheelchair users 43 60°/second 50.0 42.5 28.2 32.3
Codine et al. [6] M volunteers 26 D 60°/second 39.8 56.4
ND 42.7 56.2
M runners 23 D 60°/second 44.6 57.2
ND 45.8 54.7
M tennis players 26 D 60°/second 39.0 57.3
ND 40.1 54.0
M baseball players 20 D 60°/second 39.9 65.3
ND 39.8 55.5
Greenfield et al. [19] M and F 25 60°/second 14.1 15.4
Harbo et al. [20] M volunteers 53 60°/second 59.5 83.1
F volunteers 52 60°/second 37.4 45.7
Ivey et al. [22] M 36 60°/second 37.5 61.0 21.8 33.2
F 26 60°/second 19.5 34.2 13.0 17.9
Sirota et al. [35] M baseball players 24 D 60°/second 48.8 51.6
ND 44.2 52.3
Stickley et al. [37] F volleyball athletes 13 60°/second 16.8 22.3
Tis and Maxwell [39] F 25 D 60°/second 24.7 23.3
ND 24.7 21.2
Verney et al. [42] M volunteers 73 60°/second 46.0
30°/second 50.0
McMaster et al. [28, 29] M water polo players 26 Right 30°/second 51.8 99.1 38.2 65.9
Left 49.4 92.7 34.8 57.8
M swimming athletes 20 Right 30°/second 48.1 99.1 33.7 66.8
Left 48.6 102.2 31.8 55.9
M volunteers 22 Right 30°/second 35.3 54.0 29.1 39.9
Left 38.2 52.7 28.1 36.8
Otis et al. [32] M 26 D 48°/second 49.6 26.6 42.2
ND 46.4 26.6 38.0
Current study M and F reverse shoulder prosthesis 72 60°/second 15.2 16.1 9.3 8.2

M = male; F = female; D = dominant; ND = nondominant.


[TableWrap ID: Tab2] Table 2 

Mean maximum generated force (Nm) and SD at 60° per second for the whole group


Maximum torque at 60° per second All shoulders (N = 24) Primary (N = 13) Revision (N = 11) p value primary versus revision Cohen’s d primary versus revision
Abduction 15.2 ± 6.6 16.3 ± 5.6 13.4 ± 7.6 0.30 0.43
Adduction 16.1 ± 10.0 20.4 ± 11.8 11.8 ± 6.0 0.11 0.92
(N = 14) (N = 9) (N = 5)
External rotation 9.3 ± 4.4 9.3 ± 4.7 7.9 ± 4.0 0.58 0.32
Internal rotation 8.2 ± 2.6 9.2 ± 2.1 6.0 ± 2.5 0.07 1.38

[TableWrap ID: Tab3] Table 3 

Pearson’s correlation between maximum torque at 60° per second and the postoperative Constant-Murley score, DASH, and (D)SST


Maximum torque at 60° per second Constant-Murley score DASH (D)SST
Abduction 0.507
p = 0.014
−0.572
p = 0.004
0.519
p = 0.011
Adduction 0.393
p = 0.183
−0.319
p = 0.29
0.408
p = 0.166
External rotation 0.614
p = 0.026
−0.531
p = 0.062
0.600
p = 0.03
Internal rotation 0.441
p = 0.216
−0.498
p = 0.205
0.455
p = 0.206

(D)SST = Dutch translation of the Simple Shoulder Test.


[TableWrap ID: Tab4] Table 4 

Constant-Murley scores, relative Constant-Murley scores, DASH scores, and (D)SST


Scores All shoulders ± SD (range) Primary ± SD (range) Revision ± SD (range)
Constant-Murley preoperative 24 ± 11 (5–47) 28 ± 9 (13–47) 20 ± 12 (5–47)
Constant-Murley postoperative 50 ± 21 (8–87) 59 ± 20 (8–87) 38 ± 18 (11–73)
Relative Constant-Murley preoperative 33% ± 17% (7–71) 38% ± 14% (19–68) 27% ± 18% (7–71)
Relative Constant-Murley postoperative 70% ± 31% (9–124) 83% ± 30% (9–124) 53% ± 22% (14–92)
DASH postoperative 43.9 ± 25.6 (1.7–84.2) 31.5 ± 24.4 (1.7–77.5) 60.3 ± 17.1 (31.2–84.2)
(D)SST postoperative 7 ± 4 (0–13) 8 ± 4 (0–13) 4 ± 3 (1–10)

(D)SST = Dutch translation of the Simple Shoulder Test.



Article Categories:
  • Basic Research


Previous Document:  ?-Turn tendency in N-methylated peptides with dehydrophenylalanine residue. DFT study.
Next Document:  Liposome-encapsulated polyethylenimine/oligonucleotide polyplexes prepared by reverse-phase evaporat...