Document Detail


Anchoring Barbs and Balloon Expandable Stents: What is the Risk of Perforation and Failed Stent Deployment?
MedLine Citation:
PMID:  22819740     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
PURPOSE: Balloon expandable stents may on occasion be deployed in close proximity to the anchoring barbs of endovascular grafts. The aim of this study was to determine the risk and effect of balloon perforation by anchoring barbs and to assess whether these risks are different if the balloon is protected by a covered stent mounted upon it. METHODS: A bench-top model was developed to mimic the penetration of anchoring barbs into the lumen of medium sized blood vessels. The model allowed variation of angle and depth of vessel penetration. Both bare balloons and those with covered stents mounted upon them were tested in the model to determine whether there was a risk of perforation and which factors increased or decreased this risk. RESULTS: All combinations of barb angle and depth caused balloon perforation but this was most marked when the barb was placed perpendicular to the long axis of the balloon. When the deployment of covered stents was attempted balloon perforation occurred in some cases but full stent deployment was achieved in all cases where the perforation was in the portion of the balloon covered by the stent. The only situation in which stent deployment failed was where the barb was intentionally placed in the uncovered portion of the balloon. This resulted in only partial deployment of the stent. CONCLUSIONS: Balloon rupture is a distinct possibility when deploying balloon-expandable stents in close proximity to anchoring barbs. Care should be taken in this circumstance to ensure that the barb is well away from the uncovered portion of the balloon.
Authors:
M J Bown; G J Harrison; T V How; J A Brennan; R K Fisher; S R Vallabhaneni; R G McWilliams
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-7-20
Journal Detail:
Title:  European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery     Volume:  -     ISSN:  1532-2165     ISO Abbreviation:  -     Publication Date:  2012 Jul 
Date Detail:
Created Date:  2012-7-23     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9512728     Medline TA:  Eur J Vasc Endovasc Surg     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2012 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.
Affiliation:
Department of Cardiovascular Sciences, University of Leicester, UK.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  No Benefit from Carotid Intervention in Fatal Stroke Prevention for >80-Year-old Patients.
Next Document:  High Compression Pressure over the Calf is More Effective than Graduated Compression in Enhancing Ve...