Abstract/OtherAbstract:
Less than 1 ml of air/24 h was entrapped in the transducer dome attached to each of 4 commercially available pressurized continuous-infusion flush devices. If entrained into the circulation, this amount is less than the volume of gas associated with significant morbidity in most experimental animal studies. The amount was not increased by periodic fast flushing but was further reduced with the addition of an air-eliminating filter between the flush reservoir and the dome. Other methods to further reduce bubble formation in this flush-dome-patient (FDP) system are discussed.
Authors:
D T Soule; D J Powner
Related Documents :
15584307 - Noise in pressure transducer readings produced by variations in solar radiation.
8896297 - Ambulatory blood pressure measuring devices.
21510847 - High blood pressure, antihypertensive medication and lung function in a general adult p...
654057 - A simple technique for measurement of pressure in the tympanitic rumen of cattle.
3653357 - Socioeconomic status and hypertension in children of two state schools in naples, italy...
7661077 - Clinicians and patients with hypertension: unsettled issues about compliance.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Critical care medicine     Volume:  12     ISSN:  0090-3493     ISO Abbreviation:  Crit. Care Med.     Publication Date:  1984 Jun 
Date Detail:
Created Date:  1984-07-13     Completed Date:  1984-07-13     Revised Date:  2000-12-18    
Medline Journal Info:
Nlm Unique ID:  0355501     Medline TA:  Crit Care Med     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  520-2     Citation Subset:  AIM; IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Air*
Equipment Design
Infusions, Parenteral / instrumentation*
Pressure

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Insertion of pulmonary artery catheters--a comparison of fiberoptic and nonfiberoptic catheters.
Next Document:  Continuous S-vO2 measurement and oxygen transport patterns in cardiac surgery patients.