Agreement between an ingestible telemetric sensor system and a mercury thermometer before and after linear regression correction.
MedLine Citation:
PMID:  20051735     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVES: To evaluate the use of the CorTemp ingestible sensor system to monitor athletes in the prevention of thermal injuries, 3 objectives were established: (1) to determine the agreement between the system and a mercury thermometer and quantify the effect of exceeding the recommended yearly manufacturer calibration, (2) to establish the effect of individual sensor correction on agreement, and (3) to determine the quantity of data required for effective correction.
DESIGN: Validation study.
PARTICIPANTS: Ninety-four ingestible sensors.
INTERVENTIONS: (1) Five comparisons were made between each sensor and a mercury thermometer across the range of 33 to 41 degrees C. This was performed immediately, and 14 to 18 months, after factory calibration. (2) Linear regression equations were created and used to correct sensor readings at approximately 37 degrees C; the corrected value was compared with the mercury thermometer. (3) Linear regression equations were created for each sensor using 2 to 5 data points.
MAIN OUTCOME MEASURES: Systematic bias and random error (95%).
RESULTS: (1) Systematic bias + or - random error (95%) was 0.73% + or - 0.23% ( approximately 0.27 + or - 0.09 degrees C) and 0.54% + or - 0.28% ( approximately 0.22 + or - 0.11 degrees C) immediately and 14 to 18 months post factory calibration, respectively. (2) Regression correction improved agreement through reductions in systematic bias. (3) Individualized equations using a minimum of 3 comparisons were required to reduce agreement to + or - 0.10 degrees C; the use of 5 comparisons minimized the number of readings exceeding + or - 0.10 degrees C.
CONCLUSIONS: (1) The CorTemp system inflates temperature measurements compared with a mercury thermometer. (2) Individual sensor calibration is warranted. (3) Correction equations should use a minimum of 3, preferably 5, comparisons. After regression correction, the system displays satisfactory accuracy for preventative monitoring.
Authors:
Graeme G Challis; Jon C Kolb
Related Documents :
16329245 - Tempo of hybrid inviability in centrarchid fishes (teleostei: centrarchidae).
17671625 - Instrumentation and automated photometric titration procedure for total acidity determi...
20851045 - An ensemble method based on uninformative variable elimination and mutual information f...
22350435 - Point process time-frequency analysis of dynamic respiratory patterns during meditation...
17886235 - Sample size evaluation for a multiply matched case-control study using the score test f...
7583625 - Neurodevelopmental animal models of schizophrenia.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't; Validation Studies    
Journal Detail:
Title:  Clinical journal of sport medicine : official journal of the Canadian Academy of Sport Medicine     Volume:  20     ISSN:  1536-3724     ISO Abbreviation:  Clin J Sport Med     Publication Date:  2010 Jan 
Date Detail:
Created Date:  2010-01-06     Completed Date:  2011-01-06     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9103300     Medline TA:  Clin J Sport Med     Country:  United States    
Other Details:
Languages:  eng     Pagination:  53-7     Citation Subset:  IM    
Affiliation:
Human Performance Lab, Faculty of Kinesiology, University of Calgary, Calgary, Alberta, Canada.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Calibration
Humans
Linear Models
Mercury Compounds*
Reproducibility of Results*
Telemetry / instrumentation,  methods*
Thermometers*
Time Factors
Chemical
Reg. No./Substance:
0/Mercury Compounds

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Bike Transalp 2008: liquid intake and its effect on the body's fluid homeostasis in the course of a ...
Next Document:  Alternations of heart rate variability at lower altitude in the predication of trekkers with acute m...