Acute hemodynamic effects of different atrioventricular intervals in dual chamber pacemakers: is there an optimum atrioventricular delay.
MedLine Citation:
PMID:  1427936     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Conflicting data have been reported regarding adjustment of atrioventricular (AV) interval to maximise hemodynamic performance of dual chamber pacemakers. Eleven consecutive patients with complete heart block and dual chamber pacemakers were paced at three AV intervals (150, 200, 250 msec) and free running rates (60-93 bpm, mean 73 +/- 12 bpm) with simultaneous measurements of cardiac output, atrial contribution to left ventricular filling, left ventricular ejection fraction, and peak aortic velocity and acceleration by echo-Doppler techniques to define the optimum AV delay. At all the three AV intervals tested there was no difference in cardiac output (4.7 +/- 0.96, 4.83 +/- 1.12, 4.77 +/- 1.19 litres/min respectively, p = NS), left ventricular ejection fraction (60.2 +/- 10.6%, 61.2 +/- 9.9% and 64 +/- 8.3%, p = NS), atrial contribution to left ventricular filling (0.37 +/- 0.10, 0.38 +/- 0.09, 0.36 +/- 0.16, n = 8, p = NS), peak aortic velocity (104 +/- 8, 105 +/- 12, 104 +/- 13 cm/sec, p = NS) and aortic acceleration (19.68 +/- 4.26, 20.4 +/- 5.58 and 19.0 +/- 4.54 m/sec2, p = NS). Compared to AV delay of 150 msec an increase in cardiac output of 0.5 L/minute was observed in three patients at an AV delay of 200 msec and in one patient at the AV delay of 250 msec. These data suggest that it is difficult to generalize an optimum AV delay in patients with dual chamber pacemakers. With the AV interval in the range of 150-250 msec, only a minority of patients could improve their haemodynamics at rest by adjusting this interval if the baseline cardiac function was normal.
Authors:
J C Mohan; K K Sethi; N Pandit; M Bhargava; R Arora; M Khalilullah
Related Documents :
17184706 - Critical role of physiologist john a. johnson in the origins of minnesota's billion dol...
11341096 - Bradycardia and permanent pacing after bilateral thoracoscopic t2-sympathectomy for pri...
14516336 - Long-term complication rates in ventricular, single lead vdd, and dual chamber pacing.
11712296 - Endocardial electrograms from pacemaker catheters.
19025946 - Head-to-head comparison between delayed enhancement and percent infarct mapping for ass...
22143126 - Warfarin resistance associated with parenteral nutrition.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Indian heart journal     Volume:  44     ISSN:  0019-4832     ISO Abbreviation:  Indian Heart J     Publication Date:    1992 Mar-Apr
Date Detail:
Created Date:  1992-12-17     Completed Date:  1992-12-17     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  0374675     Medline TA:  Indian Heart J     Country:  INDIA    
Other Details:
Languages:  eng     Pagination:  79-85     Citation Subset:  IM    
Affiliation:
Department of Cardiology, G.B. Pant Hospital, New Delhi.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Echocardiography
Female
Heart Block / physiopathology,  therapy*
Hemodynamics*
Humans
Male
Middle Aged
Pacemaker, Artificial*
Ventricular Function, Left / physiology

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Percutaneous transluminal coronary angioplasty in patients over the age of 65 years.
Next Document:  Experience with recycled cardiac pacemakers.