Document Detail


Accuracy of volumetric flow rate measurements: an in vitro study using modern ultrasound scanners.
MedLine Citation:
PMID:  19854966     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: Volumetric flow measurement with Doppler ultrasound is useful in assessing blood flow as part of an evaluation of arteriovenous fistula maturity in patients undergoing hemodialysis. In this study, we assessed both accuracy and variability in volumetric flow measurements obtained using modern and commercially available ultrasound systems and an in vitro experimental setup. METHODS: Volumetric flow measurements using duplex ultrasound were obtained by 3 users operating 5 different systems for randomized flow in the range of 100 to 1000 mL/min. Users performed 3 consecutive measurements at a given flow rate. Data were analyzed using statistical techniques to assess measurement accuracy and variability. RESULTS: Over the span of flow rates studied, the root mean square error (RMSE) for the 5 ultrasound systems ranged from 38.8 to 79.7, 36.8 to 52.0, 73.0 to 85.3, 26.7 to 44.6, and 43.9 to 93.5 mL/min. Corresponding average RMSE values were 60.3, 42.7, 81.1, 37.2, and 64.4 mL/min, respectively. A linear regression analysis of mean interobserver measurements revealed an excellent correlation for all ultrasound systems (r(2) > 99.1%). Assessment of intraobserver measurements revealed no statistically significant differences for any ultrasound system evaluated (P > .94). Comparison of interobserver measurements indicates no statistically significant differences between any of the 5 systems (P > .14). CONCLUSIONS: Modern ultrasound systems are reasonably accurate in blood flow measurement in an experimental setup mimicking clinically relevant blood flow ranges in a hemodialysis fistula. Users need adequate training and experience to perform multiple measurements and use appropriate techniques to minimize errors in flow measurement.
Authors:
Kenneth Hoyt; Felix A Hester; Randall L Bell; Mark E Lockhart; Michelle L Robbin
Related Documents :
20800546 - Can doppler ultrasound-guided oocyte retrieval improve ivf safety?
9124066 - Cerebral blood flow of the exposed brain surface measured by laser doppler perfusion im...
1138476 - A directional quantifying doppler system for measurement of transport velocity of blood.
8083946 - Time-domain ultrasonography during pregnancy.
8894746 - Sneddon's syndrome (livedo racemosa and cerebral infarction) presenting psychiatric dis...
8594066 - Increased blood-brain permeability with hyperosmolar mannitol increases cerebral o2 con...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine     Volume:  28     ISSN:  1550-9613     ISO Abbreviation:  J Ultrasound Med     Publication Date:  2009 Nov 
Date Detail:
Created Date:  2009-10-26     Completed Date:  2010-01-06     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8211547     Medline TA:  J Ultrasound Med     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1511-8     Citation Subset:  IM    
Affiliation:
Department of Radiology, University of Alabama at Birmingham, Volker Hall G082, 1670 University Boulevard, Birmingham, AL 35294, USA. hoyt@uab.edu
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Arteries / physiology*,  ultrastructure*
Blood Flow Velocity / physiology*
Equipment Design
Equipment Failure Analysis
Humans
Phantoms, Imaging
Reproducibility of Results
Rheology / methods*
Sensitivity and Specificity
Ultrasonography, Doppler / instrumentation*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine