Document Detail


Accuracy of three pulse oximeters during exercise and hypoxemia in patients with cystic fibrosis.
MedLine Citation:
PMID:  8404189     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
To evaluate the reliability of noninvasive oximeters during exercise in patients with cystic fibrosis (CF), we studied 9 patients during exercise tests, using 3 devices (Hewlett-Packard 47201A, Biox-Ohmeda 3700, Nellcor 100), comparing oximetry readings with arterial blood co-oximetry. Fifty samples from preexercise, exercise, and recovery were collected; and paired measurements were analyzed for bias (arterial blood gas SaO2--noninvasive SaO2) and precision (1 SD of the bias). Values were available for the Hewlett-Packard and the Biox-Ohmeda at all 50 collection times. The Nellcor failed to read on four occasions, all during exercise. The accuracy of the Biox-Ohmeda readings was significantly worse during exercise than rest, while the Nellcor and the Hewlett-Packard performance did not differ significantly between exercise and rest. During hypoxemia (SaO2 < 90 percent as determined by co-oximetry on arterial blood), the Biox-Ohmeda and Nellcor consistently overestimated saturation, with mean bias of -6.1 percent and -3.0 percent, respectively. The Hewlett-Packard was more reliable and valid during hypoxemia, with a mean bias of -1.0 percent. The Nellcor read significantly more accurately in the six patients with severe digital clubbing than in the three patients with mild digital clubbing. It is important to be aware of possible limitations of noninvasive oximeters before accepting them as accurate.
Authors:
D M Orenstein; S E Curtis; P A Nixon; E R Hartigan
Related Documents :
8964719 - Role of hematocrit in the recruitment of pulmonary diffusing capacity: comparison of hu...
20889989 - Inappropriate exercise-induced increase in pulmonary artery pressure in patients with s...
1179039 - Idiopathic interstitial pulmonary fibrosis with hypercapnia.
8865409 - Exercise and resting pulmonary function in sarcoidosis.
6476589 - Hemodynamics of patients with severe chronic obstructive pulmonary disease during progr...
20671059 - Right ventricular load at exercise is a cause of persistent exercise limitation in pati...
9246389 - Brachial arterial blood flow during static handgrip exercise of short duration at varyi...
19024159 - A finite element model on effects of impact load and cavitation on fatigue crack propag...
19919979 - Six-minute walk test and cardiopulmonary exercise testing in patients with chronic hear...
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't; Research Support, U.S. Gov't, P.H.S.    
Journal Detail:
Title:  Chest     Volume:  104     ISSN:  0012-3692     ISO Abbreviation:  Chest     Publication Date:  1993 Oct 
Date Detail:
Created Date:  1993-11-18     Completed Date:  1993-11-18     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  0231335     Medline TA:  Chest     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  1187-90     Citation Subset:  AIM; IM    
Affiliation:
Department of Pediatrics, School of Medicine, University of Pittsburgh.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Anoxia / diagnosis*
Cystic Fibrosis / blood*,  physiopathology
Exercise / physiology
Exercise Test
Female
Heart Rate / physiology
Humans
Male
Oximetry / instrumentation,  standards*
Oxyhemoglobins / metabolism*
Reproducibility of Results
Grant Support
ID/Acronym/Agency:
MO1 RR00084/RR/NCRR NIH HHS; R01 HL 35334/HL/NHLBI NIH HHS; SO7 RR0550724/RR/NCRR NIH HHS
Chemical
Reg. No./Substance:
0/Oxyhemoglobins

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Follicular bronchitis in the pediatric population.
Next Document:  Anti-inflammatory action of erythromycin. Its inhibitory effect on neutrophil NADPH oxidase activity...