Document Detail


Accuracy of a single rigid conical cuff with standard-size bladder coupled to an automatic oscillometric device over a wide range of arm circumferences.
MedLine Citation:
PMID:  20686489     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
Although the upper arm has the shape of a truncated cone, cylindrical cuffs and bladders are currently used for blood pressure (BP) measurement. The aims of this study were to describe upper arm characteristics and to test the accuracy of a standard adult-size conical cuff coupled to an oscillometric device over a wide range of arm circumferences. Arm characteristics were studied in 142 subjects with arm circumferences ranging from 22 to 45 cm (study 1). In a subset of 33 subjects with the same range of arm circumferences, a rigid conical cuff with standard-size bladder (12.6 × 24.0 cm) and a rigid cylindrical cuff (13.3 × 24.0 cm), both coupled to a Microlife BP A100 device, were tested according to the requirements of the protocol of the European Society of Hypertension (ESH; study 2). Study 1. In all subjects, upper-arm shape was tronco-conical with slant angles ranging from 89.5° to 82.2°. In a multiple linear regression analysis, only arm circumference was an independent predictor of conicity (P<0.001). Study 2. The rigid conical cuff passed all three phases of the ESH protocol for systolic and diastolic BPs. Mean device-observer BP differences obtained with the conical cuff were unrelated to arm circumference. When the rigid cylindrical cuff was used, ESH criteria were not satisfied, and the cuff overestimated systolic BPs in subjects with large arms. BP can be measured accurately with the use of a standard-size rigid conical cuff coupled to a BP A100 device for a wide range of arm circumferences.
Authors:
Elisa Bonso; Francesca Saladini; Ada Zanier; Elisabetta Benetti; Francesca Dorigatti; Paolo Palatini
Related Documents :
23832349 - Regulation of cytochrome c- and quinol oxidases, and piezotolerance of their activities...
14959969 - Accuracy of aneroid sphygmomanometer blood pressure recording compared with digital and...
12608969 - Comparison of ankle-brachial pressure index measurements using an automated oscillometr...
10647359 - 'putting the pressure on': a study of compression sleeves used in breast cancer-related...
1604869 - A method for rapid intracompartmental pressure measurement.
2350639 - The use of automatic blood pressure monitors in the burned patient.
2603989 - Characterization of intact mesenteric lymphatic pump and its responsiveness to acute ed...
17489939 - Acute action of aminophylline in patients with amyotrophic lateral sclerosis.
1900259 - High density lipoprotein turnover in patients with hypertension.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2010-08-05
Journal Detail:
Title:  Hypertension research : official journal of the Japanese Society of Hypertension     Volume:  33     ISSN:  1348-4214     ISO Abbreviation:  Hypertens. Res.     Publication Date:  2010 Nov 
Date Detail:
Created Date:  2010-11-05     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9307690     Medline TA:  Hypertens Res     Country:  England    
Other Details:
Languages:  eng     Pagination:  1186-91     Citation Subset:  IM    
Affiliation:
Department of Clinical and Experimental Medicine, University of Padova, Padua, Italy.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Glomerular angiotensinogen is induced in mesangial cells in diabetic rats via reactive oxygen specie...
Next Document:  Comparison of strict- and mild-blood pressure control in elderly hypertensive patients: a per-protoc...