Document Detail


Accuracy and reproducibility of 30 devices for self-measurement of arterial blood pressure.
MedLine Citation:
PMID:  8267944     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Thirty commercially available noninvasive devices for self-measurement of blood pressure were tested for accuracy and reproducibility against intraarterial measurement in 15 normotensive and 15 hypertensive subjects. For reference to outpatient measurements, a standard sphygmomanometer was included. Analysis of the data for any device was performed according to a statistical model in which two types of errors, representing accuracy and reproducibility, were distinguished: offset (mean difference between intraarterial measurement and device) and standard deviation (comprising intraindividual and interindividual variability). Linearity of the devices was analyzed by linear regression with intraarterial measurement as independent variable. The average offset of all tested devices amounted to 11.7 mm Hg (ranging from -0.2 to 21.4 between instruments) for systolic blood pressure and -1.6 mm Hg (range -11.4 to 8.1) for diastolic blood pressure. The standard deviation between instruments ranged from 7.0 to 15.4 mm Hg for systolic and from 3.3 to 16.0 mm Hg for diastolic blood pressure. Except for the devices based on the oscillometric principle, most (semi-) automated devices demonstrated a variability similar to the mercury sphygmomanometer (offset = 14.3 for systolic and 0.1 mm Hg for diastolic pressure; standard deviation = 7.5 for systolic and 5.6 mm Hg for diastolic blood pressure). Devices based on the oscillometric principle had a significantly larger variability. It is concluded that new devices should be tested for accuracy and reproducibility before application for clinical use. Blood pressure measurements from instruments with larger offset are only comparable to values from other instruments after rescaling in accordance with their offset.
Authors:
J van Egmond; J W Lenders; E Weernink; T Thien
Related Documents :
6979544 - A regional ultrasonic technique to measure left ventricular performance during coronary...
17346624 - Changes in sarcomeric and non-sarcomeric cytoskeletal proteins and focal adhesion molec...
17618034 - Cardiac arrest with continuous mechanical chest compression during percutaneous coronar...
12145614 - A new device for measuring esophageal variceal pressure.
10215714 - Theatre monitor alarm settings: a pilot survey in scotland and belgium.
24092334 - Definition and classification of pulmonary hypertension.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  American journal of hypertension     Volume:  6     ISSN:  0895-7061     ISO Abbreviation:  Am. J. Hypertens.     Publication Date:  1993 Oct 
Date Detail:
Created Date:  1994-01-31     Completed Date:  1994-01-31     Revised Date:  2009-02-24    
Medline Journal Info:
Nlm Unique ID:  8803676     Medline TA:  Am J Hypertens     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  873-9     Citation Subset:  IM    
Affiliation:
Institute for Anesthesiology, St. Radboud Hospital, Nijmegen, The Netherlands.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Blood Pressure
Blood Pressure Determination / instrumentation*
Female
Humans
Male
Middle Aged
Regression Analysis
Reproducibility of Results
Sensitivity and Specificity

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Rare renal transcription of the atrial natriuretic factor gene in rats. Demonstration by polymerase ...
Next Document:  Ambulatory monitoring of blood pressure in normal pregnancy.