Accuracy and 'range of uncertainty' of oscillometric blood pressure monitors around the upper arm and the wrist.
MedLine Citation:
PMID:  10212375     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Although they have been marketed widely, few data about the diagnostic accuracy of blood pressure monitors are available. METHODS: Repeated measurements of blood pressures in 85 patients were performed in random sequence with two oscillometric blood pressure monitors around the upper arm (Visomat OZ2) and the wrist (Omron R3( and with a standard sphygmomanometer. The oscillometric blood pressure monitors were validated according to protocols of the British Hypertension Society (BHS) and the American Association for the Advancement of Medical Instrumentation (AAMI). Subsequently, sensitivity and specificity of these monitors for the diagnosis of hypertension or exclusion of the possibility of its presence in a general medical outpatient population were calculated. RESULTS: Sphygmomanometric readings exceeded oscillometric blood pressure measurements by 3.7+/-7.5/4.8+/-5.6 mmHg (systolic/diastolic) for the upper arm and 5.7+/-6.2/6.8+/-6.8 mmHg for the wrist. Deviations occurred in both directions and were higher for blood pressures in the hypertensive range. Oscillometric blood pressure measurements at the upper arm, but not at the wrist, satisfied validation criteria of BHS and AAMI protocols. Optimal sensitivity and specificity for the diagnosis of hypertension, defined as blood pressure > 140/90 mmHg with a standard sphygmomanometer, was achieved with blood pressure limits of 133/82 mmHg for the Visomat OZ and 131/80 mmHg for the Omron R3. CONCLUSIONS: Average sphygmomanometer values exceed oscillometrically measured blood pressure values but individual disagreements cannot be predicted. Measurements at the upper arm are more accurate than are those at the wrist according to the validation protocols of the BHS and AAMI. Additional appraisal of sensitivities and specificities and of a 'range of uncertainty' for the diagnosis of hypertension may allow better judgement of accuracy of individual oscillometric blood pressure measurements.
Authors:
T Dieterle; E Battegay; B Bucheli; B Martina
Related Documents :
7215085 - Elevated leg systolic pressures and arterial calcification in diabetic occlusive vascul...
9821035 - Dobutamine effect on ankle-brachial pressure index in patients with peripheral arterial...
6869115 - Posture and the arterial pressure in the ischaemic foot.
3740925 - Dinamap fails to detect hypotension in very low birthweight infants.
7215085 - Elevated leg systolic pressures and arterial calcification in diabetic occlusive vascul...
22844615 - The use of nurse checklists in a bedside computer-based information system to focus on ...
Publication Detail:
Type:  Clinical Trial; Journal Article; Randomized Controlled Trial; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  3     ISSN:  1359-5237     ISO Abbreviation:  Blood Press Monit     Publication Date:  1998  
Date Detail:
Created Date:  1999-06-10     Completed Date:  1999-06-10     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  ENGLAND    
Other Details:
Languages:  eng     Pagination:  339-46     Citation Subset:  IM    
Affiliation:
Medical Outpatient Clinic and Hypertension Clinic, Department of Internal Medicine, University Hospital, CH-4031, Basel, Switzerland.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Aged
Aged, 80 and over
Arm / physiopathology*
Blood Pressure / physiology*
Blood Pressure Determination / methods*
Blood Pressure Monitoring, Ambulatory
Humans
Hypertension / physiopathology*
Middle Aged
Wrist / physiopathology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Target-organ damage and cardiovascular risk profile in resistant hypertension. Influence of the whit...
Next Document:  Validation of the Omron F3 blood pressure measuring device for use in the finger through simultaneou...