Document Detail


Accuracy of the pulse delay time technique with triple cuff for objective indirect blood pressure measurement.
MedLine Citation:
PMID:  23235357     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
OBJECTIVE:: We developed a new technique to measure blood pressure (BP) - the triple cuff and pulse delay time technique (TC-PDT) - which measures delay time with a single sensor at a single site unlike the existing delay time methods. This study aimed to evaluate the adequacy and accuracy of TC-PDT. METHODS:: TC-PDT measures delay time, which represents the difference in the arrival time of the arterial pulse wave between the proximal and distal areas of the cuff and provides clear criteria for BP determination based on the mechanism. Adequacy of delay time in TC-PDT was confirmed in comparison with established delay time of the pulse transit time technique (PTT). TC-PDT accuracy was evaluated according to the standard guidelines. RESULTS:: In TC-PDT, delay time showed changes in parallel with those observed in PTT. The correlation coefficient between BP by TC-PDT and that by PTT was 0.98 (P < 0.01) for SBP and 0.93 (P < 0.01) for DBP. According to the International Protocol guidelines (2002), the BP measurements of 33 patients met the recommended values for phase 1, 2.1 and 2.2 measurements. As per the Association for the Advancement of Medical Instrumentation (AAMI) SP10 guidelines (2002), the mean difference in BP measured by TC-PDT and that measured by the auscultatory technique in 85 patients was 1.0 ± 5.1 mmHg for SBP and -0.4 ± 4.5 mmHg for DBP, which conformed to the requirements. CONCLUSION:: TC-PDT is adequate to measure delay time and provides adequate and accurate measurements of BP. This device has the potential to be a standard tool for indirect BP measurement.
Authors:
Tetsuya Fujikawa; Osamu Tochikubo; Teruyasu Sugano; Satoshi Umemura
Related Documents :
12911367 - The effect of variable-dose diazepam on dreaming and emergence phenomena in 400 cases o...
9285007 - Cardiopulmonary and anesthetic effects of propofol in wild turkeys.
3828177 - Cardiovascular and catecholamine responses to laryngoscopy with and without tracheal in...
3259337 - Comparison of propofol and thiopental for the induction of anesthesia.
23105927 - Decreases in casz1 mrna by an sirna complex do not alter blood pressure in mice.
2084277 - Comparison of the acute effects of captopril and of nifedipine on left ventricular dias...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-12-11
Journal Detail:
Title:  Journal of hypertension     Volume:  -     ISSN:  1473-5598     ISO Abbreviation:  J. Hypertens.     Publication Date:  2012 Dec 
Date Detail:
Created Date:  2012-12-13     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8306882     Medline TA:  J Hypertens     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
aDepartment of Medical Science and Cardiorenal Medicine, Yokohama City University Graduate School of Medicine, Yokohama bDepartment of Preventive Medicine, Yokohama City University School of Medicine, Yokohama, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effects of continuous positive airway pressure on blood pressure in hypertensive patients with obstr...
Next Document:  A functional variant in the serine-threonine kinase coding gene is associated with hypertension: a c...