Accuracy of cornea and lens biometry using anterior segment optical coherence tomography.
MedLine Citation:
PMID:  18163839     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
We assess the accuracy of the Visante anterior segment optical coherence tomographer (AS-OCT) and present improved formulas for measurement of surface curvature and axial separation. Measurements are made in physical model eyes. Accuracy is compared for measurements of corneal thickness (d(1)) and anterior chamber depth (d(2)) using-built-in AS-OCT software versus the improved scheme. The improved scheme enables measurements of lens thickness (d(3)) and surface curvature, in the form of conic sections specified by vertex radii and conic constants. These parameters are converted to surface coordinates for error analysis. The built-in AS-OCT software typically overestimates [mean+/-standard deviation(SD)]d(1) by +62+/-4 mum and d(2) by + 4 +/- 88 microm. The improved scheme reduces d(1) (-0.4 +/- 4 microm) and d(2) (0 +/- 49 microm) errors while also reducing d(3) errors from +218 +/- 90 (uncorrected) to +14 +/- 123 microm (corrected). Surface x coordinate errors gradually increase toward the periphery. Considering the central 6-mm zone of each surface, the x coordinate errors for anterior and posterior corneal surfaces reached +3 +/- 10 and 0 +/- 23 microm, respectively, with the improved scheme. Those of the anterior and posterior lens surfaces reached +2 +/- 22 and +11 +/- 71 microm, respectively. Our improved scheme reduced AS-OCT errors and could, therefore, enhance pre- and postoperative assessments of keratorefractive or cataract surgery, including measurement of accommodating intraocular lenses.
Authors:
Mark C M Dunne; Leon N Davies; James S Wolffsohn
Related Documents :
20940809 - A technique for calculating the amplitude distribution of propagated fields by gaussian...
20212629 - Aberration corrections for a pos hologram scanner.
20802629 - Midfocus matching: a method for calculating lens positions for a gaussian beam in a two...
19351379 - Evaluation of optical quality in white light from wavefront aberrations for a myopic po...
19404349 - A low cost adaptive optics system using a membrane mirror.
18267679 - A lamb wave lens for acoustic microscopy.
8927719 - Microdosimetry of monoenergetic neutrons.
19493109 - Real-time data acquisition incorporating high-speed software correlator for single-mole...
25173259 - Conception and realization of a parallel-plate free-air ionization chamber for the abso...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of biomedical optics     Volume:  12     ISSN:  1083-3668     ISO Abbreviation:  J Biomed Opt     Publication Date:    2007 Nov-Dec
Date Detail:
Created Date:  2007-12-31     Completed Date:  2008-03-14     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9605853     Medline TA:  J Biomed Opt     Country:  United States    
Other Details:
Languages:  eng     Pagination:  064023     Citation Subset:  IM    
Affiliation:
Aston University, Ophthalmic Research Group, Birmingham, United Kingdom. m.c.m.dunne@aston.ac.uk
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Anterior Eye Segment / anatomy & histology*
Biometry
Cornea / anatomy & histology*
Humans
Lens, Crystalline / anatomy & histology*
Models, Anatomic
Software
Tomography, Optical Coherence / methods*,  statistics & numerical data

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Tissue birefringence of hypercholesterolemic rat liver measured with polarization-sensitive optical ...
Next Document:  In vivo video-rate cellular-level full-field optical coherence tomography.