Accuracy of automated blood pressure monitors.
MedLine Citation:
PMID:  18755068     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: The purpose of this study is to determine if automated and aneroid manometers are as accurate a means of determining blood pressure as the mercury manometer. Obtaining vital signs for patients is considered standard of care, yet many dental offices do not routinely perform this health service because of technique inconsistencies and time constraints. The use of automatic blood pressure monitors addresses both concerns. The mercury column manometer, the control in this study, has long been considered the most accurate and preferred instrument for obtaining blood pressure measurements. METHODS: During this study, 94 participants (19 years of age and older) consented to having blood pressure taken by each of 4 different monitors. These included the mercury column manometer and stethoscope, the aneroid manometer and stethoscope, the automatic arm blood pressure monitor, and the automatic wrist blood pressure monitor. Each of 3 investigators was assigned to and calibrated for a specific monitoring device. All measurements were taken from the left arm with 5 minutes allowed between measurements. Identical stethoscopes were used with the manual monitors. Strict adherence to the manufacturers' directions and patient preparation was followed for all monitors. Investigators were not aware of readings obtained by other investigators during testing. Eighty-three subjects completed all tests. RESULTS: Review and analysis of data indicates little difference for pulse readings between the automated and digital methods. Systolic readings by automated wrist manometers were the most unreliable. Automated arm monitors tended to provide higher measures than the mercury standard on average, and demonstrated significantly different diastolic readings in one age group compared to the control. All monitors exhibited low reliability for participants over age 50 compared to the control. CONCLUSION: This study demonstrates there is inaccuracy in the use of automated blood pressure monitors and traditional aneroid manometers when compared to the gold standard mercury column manometer for subjects of all ages and blood pressure ranges.
Authors:
Debralee Nelson; Beverly Kennedy; Carissa Regnerus; Amy Schweinle
Related Documents :
3436078 - A finger volume-oscillometric device for monitoring ambulatory blood pressure: laborato...
17085808 - Animal models for hypertension/blood pressure recording.
9095818 - Caring for the intra-aortic balloon pump patient. most frequently asked nursing questions.
10829398 - Validation of automated oscillometric sphygmomanometer (hdbpm) for arterial pressure me...
19622688 - Asymmetric insular function predicts positional blood pressure in nondemented elderly.
7014658 - Effect of pindolol on blood pressure, plasma renin activity, and catecholamines in hype...
Publication Detail:
Type:  Comparative Study; Journal Article     Date:  2008-07-01
Journal Detail:
Title:  Journal of dental hygiene : JDH / American Dental Hygienists' Association     Volume:  82     ISSN:  1553-0205     ISO Abbreviation:  J Dent Hyg     Publication Date:  2008  
Date Detail:
Created Date:  2008-08-29     Completed Date:  2009-01-26     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8902616     Medline TA:  J Dent Hyg     Country:  United States    
Other Details:
Languages:  eng     Pagination:  35     Citation Subset:  D    
Affiliation:
Division of Health Sciences, School of Medicine, University of South Dakota, USA. Deb.Nelson@usd.edu
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Age Factors
Aged
Aged, 80 and over
Blood Pressure / physiology
Calibration
Dental Care
Diastole / physiology
Equipment Design
Heart Rate / physiology
Humans
Middle Aged
Oscillometry / instrumentation
Posture / physiology
Pulse
Reproducibility of Results
Respiration
Sphygmomanometers / classification,  standards*,  statistics & numerical data
Stethoscopes
Systole / physiology
Young Adult

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effects of 5 different finger rest positions on arm muscle activity during scaling by dental hygiene...
Next Document:  Cytosolic phospholipase A(2), lipoxygenase metabolites, and reactive oxygen species.