Abstract/OtherAbstract:
BACKGROUND: Previously, we validated determination of ankle-brachial index using an automatic blood pressure device.
AIM: To test the feasibility and accuracy of the automatic method in assessing pathological ankle-brachial indexes in routine preventive examinations.
METHODS: Two physicians enrolled 354 subjects (74% men) randomly for automatic ankle-brachial index measurements using an OMRON HM 722 device. Ankle-brachial index was calculated by dividing the highest value obtained at each ankle by the highest arm value. Each subject with an abnormal (less than 0.90) automatic index and the six subsequent subjects underwent Doppler index determination.
RESULTS: Automatic ankle-brachial index determination was possible in both ankles in 350 subjects (99%; mean time 8.1+/-2.1 minutes). The incidence of abnormal automatic ankle-brachial index was 8% (n=28). Correlations between the automatic and Doppler methods were good in left and right legs (r=0.84 and 0.78, respectively; p<0.001). In subjects with an abnormal automatic index, correlations with Doppler indexes were good in both legs (r=0.67, p<0.001). In terms of detecting an abnormal index in a routine preventive examination, the automatic method had good sensitivity (92%), specificity (98%), positive predictive value (86%), negative predictive value (99%) and accuracy (97% compared with the Doppler method). Good results were obtained in subjects with an abnormal index in terms of agreement and concordance with the Doppler method (kappa=0.87).
CONCLUSIONS: The use of a commercially available automatic blood pressure device to detect peripheral artery disease appears feasible and quick in routine medical examinations.
Authors:
Daniel Benchimol; Xavier Pillois; Alain Benchimol; André Houitte; Pierre Sagardiluz; Luc Tortelier; Jacques Bonnet
Publication Detail:
Type:  Comparative Study; Evaluation Studies; Journal Article     Date:  2009-06-21
Journal Detail:
Title:  Archives of cardiovascular diseases     Volume:  102     ISSN:  1875-2136     ISO Abbreviation:  Arch Cardiovasc Dis     Publication Date:    2009 Jun-Jul
Date Detail:
Created Date:  2009-08-11     Completed Date:  2009-12-11     Revised Date:  2011-04-25    
Medline Journal Info:
Nlm Unique ID:  101465655     Medline TA:  Arch Cardiovasc Dis     Country:  Netherlands    
Other Details:
Languages:  eng     Pagination:  519-24     Citation Subset:  IM    
Affiliation:
Inserm U828, université Victor-Segalen de Bordeaux 2, hôpital Cardiologique, avenue du Haut-Lévêque, 33604 Pessac, France.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Ankle / blood supply*
Automation
Blood Pressure*
Blood Pressure Determination / instrumentation*
Brachial Artery / physiopathology*
Equipment Design
Feasibility Studies
Female
Humans
Male
Middle Aged
Peripheral Vascular Diseases / diagnosis*,  physiopathology,  ultrasonography
Predictive Value of Tests
Preventive Medicine / instrumentation*
Reproducibility of Results
Sensitivity and Specificity
Time Factors
Ultrasonography, Doppler

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Single-centre experience with the Thoratec paracorporeal ventricular assist device for patients with...
Next Document:  Incidence of atrial fibrillation during very long-term follow-up after radiofrequency ablation of ty...