The Accuracy of Noninvasive Hemoglobin Measurement by Multiwavelength Pulse Oximetry After Cardiac Surgery.
MedLine Citation:
PMID:  21918163     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
Background:In March 2008, a new multiwavelength pulse oximeter, the Radical 7 (Rad7; Masimo Corp., Irvine, CA), was developed that offers noninvasive measurement of hemoglobin concentration. Accuracy has been established in healthy adults and some surgical patients, but not in cardiac surgery intensive care patients, a group at high risk of postoperative bleeding events and anemia in whom early diagnosis could improve management.Methods:In this prospective, observational study conducted in a cardiovascular intensive care unit, we compared hemoglobin concentrations shown by the Rad7 with arterial hemoglobin concentrations determined by an automated hematology analyzer, XE-2100 (Roche, Neuilly sur Seine, France). Two software versions of Rad7 (V 7.3.0.1 [42 points of comparison in 14 patients] and the updated V 7.3.1.1 [61 points of comparison in 27 patients]) were studied during two 1-week periods. Bias, defined as the difference between the 2 methods (Masimo SpHb - XE-2100 laboratory hemoglobin), was calculated. A negative bias indicated that the Masimo underestimated hemoglobin compared with the laboratory analyzer. Correlation between the perfusion index given by Rad7 and the hemoglobin bias was also studied.Results:Correlations between Rad7 and XE-2100 were weak for both software versions (R(2) = 0.11 for V 7.3.0.1 and R(2) = 0.27 for V 7.3.1.1). Mean bias was -1.3 g/dL for V 7.3.0.1 and -1.7 g/dL for V 7.3.1.1, with wide 95% prediction intervals for the bias (respectively, -4.6 to 2.1 g/dL and -5.7 to 2.3 g/dL). The absolute hemoglobin bias tended to increase when the perfusion index decreased. For the V 7.3.0.1 software, the average absolute bias was 1.9 g/dL for perfusion index <2 and 0.8 g/dL for perfusion index >2 (P = 0.03). For V 7.3.1.1, the mean absolute bias was 2.1 g/dL when the perfusion index was <2, and 1.6 g/dL when the perfusion index was >2 (P = 0.26).Conclusions:Our study demonstrates poor correlation between hemoglobin measured noninvasively by multiwavelength pulse oximetry and a laboratory hematology analyzer. The difference was greater when the pulse oximetry perfusion index was low, as may occur in shock, hypothermia, or vasoconstriction patients. The multiwavelength pulse oximetry is not sufficiently accurate for clinical use in a cardiovascular intensive care unit.
Authors:
Ba-Vinh Nguyen; Jean-Louis Vincent; Emmanuel Nowak; Michelle Coat; Nicolas Paleiron; Pierre Gouny; Mehdi Ould-Ahmed; Maité Guillouet; Charles Christian Arvieux; Gildas Gueret
Related Documents :
12641503 - Therapy of osteoporosis in patients with crohn's disease: a randomized study comparing ...
24371493 - The relationship between alanerv(®) consumption and erythrocytes' glyoxalases i and ii...
23166343 - Continuous versus bolus infusion of doxorubicin in children with all: long-term cardiac...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-9-14
Journal Detail:
Title:  Anesthesia and analgesia     Volume:  -     ISSN:  1526-7598     ISO Abbreviation:  -     Publication Date:  2011 Sep 
Date Detail:
Created Date:  2011-9-15     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1310650     Medline TA:  Anesth Analg     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
From the *Department of Anesthesiology and Critical Care Medicine, University Hospital, CHU Brest, France;
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Intraperitoneal Ropivacaine Nebulization for Pain Management After Laparoscopic Cholecystectomy: A C...
Next Document:  Hyperfibrinolysis Diagnosed by Rotational Thromboelastometry (ROTEM(R)) Is Associated with Higher Mo...